Wietske van der Zwaag1, David G. Norris2, and José P. Marques2
1Spinoza Centre for Neuroimaging, Amsterdam, Netherlands, 2Donders Institute for Brain, Cognition and Behaviour, Nijmegen, Netherlands
Synopsis
In
this simulation study we evaluate the TE that offers the optimum BOLD contrast
per unit time for a given field strength (taking into account the
field-specific T1, T2* and ΔT2* values) as a function of the number
of slice excitations to acquire the volume (minimum TR achievable), using the
Ernst angle for excitation. Generally, optimum BOLD sensitivity is found at TEs
closer to T2* when shorter TRs (higher multiband acceleration factors) are
employed. At longer TRs efficiency constraints tend to make the optimum TE
shorter with this effect being more pronounced at lower static fields.Target audience
Anyone working with fMRI
Purpose
To find
the optimal echo time (TEopt) for highly accelerated fMRI acquisitions.
A
simple calculation shows that the largest BOLD signal is obtained at TE≈T2* of gray matter(1). Many fMRI
experiments, however, employ a shorter TE to improve temporal sampling
density, the BOLD contrast-to-noise ratio (CNRBOLD), or to reduce through-slice dephasing. Here, we
simulated CNRBOLD as a function of TR, TE, field strength and number
of slice excitations Nslices, taking into account the BOLD signal amplitude (which peaks at TE≈T2*GM ), the
increase in CNRBOLD with increased temporal sampling density as well
as the signal loss due to T1-weighting, always using the Ernst angle
for excitation. TR
was assumed to be >>T2. BOLD changes in a venous compartment were also simulated to
estimate BOLD specificity.
Methods
BOLD detection
power was modelled taking into account TR, sampling time and expected T2*BOLD
contrast: $$CNR_{BOLD} = \Delta R_2^* \times TE\times exp^{\frac{-TE}{T_2^*}} \frac{S_{\alpha_{Ernst}} \sqrt{ETL}}{\sqrt{TR}}$$
whereby $$$S_{\alpha_{Ernst}} = sin(\alpha_{Ernst})\frac{1-exp^{\frac{-TR}{T_{1}}}}{1-cos(\alpha_{Ernst})exp^{\frac{-TR}{T_{1}}}}$$$.
The minimum TR was assumed to be $$$TR = (TE+\frac{ETL}{2})N_{slices}$$$,
with the centre of k-space sampled in the middle of the echo train, where ETL
is the echo train length and Nslices the number of slice excitations.
A 0.65mm3 7T field map (in rad/s,
scaled appropriately for lower field strengths) was used to calculate signal
loss from through-slice dephasing assuming an axial slice orientation and $$$CNR_{BOLD} =CNR_{BOLD}sinc(\gamma \frac{\delta \Delta B_{0}}{\delta z}\Delta z TE)$$$.
Simulations were performed for a range of ETLs, B0,
number slice excitations and thicknesses. For brevity only results for 100/60/30/15
slices, ETL=30ms, B0 =1.5/3/7T
and a slice thickness of 2 mm are presented. Table 1 lists the relevant
relaxation parameters used in the simulations.
Results and discussion
Table 2 shows TEopt values for protocols with different NSlices. It can be seen that, even in the absence of through-slice
dephasing, TEopt was generally found to be < T2* and closer to the typically used TE values at lower field strength (Table 2). However, TEopt tends to increase with decreasing NSlices. The variation is larger at lower field strengths where the shorter T1 values make large NSlices (and increased TRs) suboptimal. Figure 1 compares BOLD sensitivity at the different fields for NSlices=100 and NSlices=15 (corresponding to whole brain coverage with a through-slice
acceleration factor of ~6 for 2mm slices).
There,
the increase of TEopt can be seen, but also the fact that lower
fields benefit more from multiband technology (Figure 1) with a CNRBOLD
improvement of 90, 40 and 18% at 1.5, 3 and 7T respectively (ignoring g-noise
penalties).
Relative venous
signal contributions (Figure 1) increase with B0 and reduce with TE in
all cases.
When taking
into account through-slice dephasing due to B0 field inhomogeneities, shorter
optimal TE values are found above the frontal sinuses at 3T (Figure 2). At the
optimal echo time, this indeed results in lower CNRBOLD in the affected
regions (Figure 3, top row). However, if a shorter TE (for illustration: 35ms@1.5T,
30ms@3T, 25ms@7T, Figure 3 bottom row) is used to prevent BOLD signal losses
over the sinuses, the BOLD contrast in the rest of the brain is significantly
lowered, while only small increases in CNRBOLD are found in the affected
regions.
For example, TE=35ms at 1.5T leads to
homogeneous BOLD CNR levels throughout the example slice, but this tSNR level
is 10-20% lower than the slower, TE=74ms acquisition (Figure 3, top left). For
TE=30ms at 3T versus the 44ms TE protocol (Figure 3, middle column), there is a
4% increase in tSNR in the frontal region, accompanied by a larger, 6%, loss of
CNRBOLD in the cortex. For 7T the choice for a shorter TE might be more
beneficial, as here the through-slice dephasing in a 2-mm slab is more
significant.
For simplicity, we have kept the ETL constant throughout the simulations shown here, its increase implies an increase of SNR at the cost of increased distortion and TR.
Conclusion
The use of TE
values shorter than T2* has become commonplace to allow for the acquisition of a
large number of slices in a limited TR and to reduce dropout. However, the application of parallel
imaging acceleration in the slice-encoding direction means that optimal BOLD
contrast is achieved at TE values close(r) to T2*, anywhere in the brain at
1.5T and 3T, and also in well-shimmed tissue at 7T.
Considerations on the importance of removing cardiac noise, or studying temporal characteristics of the hrf, have not been taken into account here, but could justify sacrificing BOLD sensitivity for temporal resolution.
Acknowledgements
No acknowledgement found.References
(1) Menon et al, MRM, 1993
(2) van der
Zwaag et al, NMR in Biomedicine, 2015
(3)
Triantafyllou et al, Plos One, 2011
(4) van der
Zwaag et al, Neuroimage, 2009