Nicholas P Blockley1, Naomi C Holland1, and Alan J Stone1
1FMRIB Centre, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom
Synopsis
The results of a recently described
streamlined quantitative BOLD (qBOLD) technique suggest that this method may
overestimate the deoxygenated blood volume (DBV), leading to an underestimate
of the oxygen extraction fraction (OEF). We hypothesise that this is due to the
effect of diffusion, which is assumed to be zero in the analytical qBOLD model.
In this study we performed Monte Carlo simulations to investigate this
hypothesis and found that DBV and OEF measurements were vessel size dependent.
However, the R2’ measurements which underlie qBOLD were found to be a
reliable measure of deoxyhaemoglobin content above a vessel radius of 10μm.Purpose
Recently a streamlined quantitative BOLD
(qBOLD) technique was described to improve the robustness of measurements of
the oxygen extraction fraction (OEF) and deoxygenated blood volume (DBV) by
simplifying the analysis model1. This was achieved by removing
confounding effects during data acquisition; enabling the number of model
compartments to be reduced. However, measurements of DBV using this technique
were larger than typically observed; potentially explaining why the measured OEF in
these experiments was lower than expected. We hypothesise that this is
due to the effect of diffusion, which is not accounted for in the qBOLD model
where the static dephasing regime (SDR) is assumed2. The SDR assumption
has been challenged by the results of numerical simulations of the Gradient
Echo Sampling of Spin Echo (GESSE) pulse sequence that demonstrate a diffusion
dependent shift in the R2′-weighted signal curve3 (R2′ - reversible transverse relaxation rate). However, the streamlined qBOLD approach uses the Asymmetric Spin Echo (ASE) pulse sequence, for which the effect of diffusion has not been examined. In this work we compare the effect of
diffusion on the ASE technique versus the GESSE technique using Monte Carlo simulations.
We then examine the implications of this effect on measurements of R2′, DBV and OEF using ASE.
Methods
Synthetic R2′-weighted qBOLD signals were generated following previously described Monte Carlo simulation approaches3,4. In brief, blood vessels were modelled as infinitely long cylinders, which were uniformly and randomly distributed in a spherical universe up to a defined volume fraction. The blood oxygenation of these vessels was modelled as an intravascular to extravascular susceptibility difference. A proton was allowed to randomly diffuse around this universe sampling the magnetic field perturbations due to the vessels. The magnetic field at the proton location manifests as a phase accrual, which was sampled every 200μs and saved every 2ms. This process of universe generation and proton phase accrual was repeated 10,000 times. Phases were added or subtracted to simulate the GESSE and ASE techniques and the sum of phases across all protons used to simulate the magnitude signal. The amount of R2′-weighting is determined by the spin echo displacement time τ. The range of τ values was matched for GESSE and ASE. Due to the low vessel volume fraction (~2-5%), protons spend the majority of their time in the extravascular space and as such these simulations model the extravascular signal (a first-order approximation of the experimentally observed signal).
This approach was used to investigate the effect of varying vessel radius in the range 1μm to 1mm, for a fixed diffusion coefficient4 D=1.3μm2ms-1. GESSE and ASE signals were compared for DBV=3%, OEF=40%, Hct=40%. The analytical qBOLD model1 (Fig. 1, Eq. 1) was used to estimate R2′, DBV and OEF from the simulated ASE data. Comparison was made with the SDR by setting D=0.
$$OEF=\frac{R_2^\prime}{DBV\;\gamma\frac{4}{3}\pi\;\Delta\chi_0\;Hct\;B_0}\tag{1}$$
Results
Fig. 2 compares the signal decay for the
GESSE and ASE methods. As in previous work
3 a shift in the GESSE
signal curve is observed with decreasing vessel radius (Fig. 2a). This effect
is not observed for the ASE signal, although the short τ signal shows
increasing attenuation with decreasing vessel radius (Fig. 2b). The full analytical
solution for the SDR is plotted as a black line with 1mm vessels approaching
this regime. Fig. 3 displays the fitted qBOLD parameters as a function of
vessel radius. When D=0 (SDR) the estimated parameter values are independent of
vessel radius, whereas with diffusion all parameters show vessel radius
dependence.
Discussion
Quantitative BOLD is a promising technique
for the measurement of OEF using endogenous contrast.
The simulations presented here reveal that the
GESSE and ASE pulse sequences are both affected by diffusion, but in different
ways. As hypothesised for ASE, diffusion is shown to result in an overestimation of
DBV for typical parenchymal vessel radii (5-25μm), consistent with previous ASE
based qBOLD measurements
1,5. In contrast, R
2′
measurements appear to be insensitive to vessel radius above a critical value (>10μm),
providing a more robust parameter albeit sensitive to the product of DBV and
OEF. With this in mind, it seems likely that the underestimation of OEF is due
to the overestimation of DBV. Further work is required to understand the
contribution of intravascular signal and vessels with multiple radii and blood
oxygenation levels.
Conclusions
The effect of diffusion manifests as a vessel
radius dependence of R
2′, DBV and OEF measured using qBOLD with an
ASE acquisition. However, by incorporating the results of these simulation in a
modified qBOLD model, more accurate estimates of DBV may be possible and hence
more accurate measurements of OEF.
Acknowledgements
This work was funded by EPSRC grant EP/K025716/1.References
1. Stone AJ, Blockley
NP. A streamlined approach to mapping the oxygen extraction fraction (OEF) and deoxygenated
blood volume (DBV) using the quantitative BOLD technique. In Proceedings of the
23rd Annual Meeting of the ISMRM, Toronto, 2015. Abstract 219.
2. He X, Yablonskiy
DA. Quantitative BOLD: Mapping of human cerebral deoxygenated blood volume and
oxygen extraction fraction: Default state. Magn. Reson. Med. 2007;57:115–126.
3. Dickson JD, Ash
TWJ, Williams GB, Harding SG, Carpenter TA, Menon DK, Ansorge RE. Quantitative
BOLD: the effect of diffusion. J. Magn. Reson. Imaging 2010;32:953–961.
4. Boxerman JL,
Hamberg LM, Rosen BR, Weisskoff RM. MR contrast due to intravascular magnetic
susceptibility perturbations. Magn. Reson. Med. 1995;34:555–566.
5. An H, Lin W. Impact
of intravascular signal on quantitative measures of cerebral oxygen extraction
and blood volume under normo- and hypercapnic conditions using an asymmetric
spin echo approach. Magn. Reson. Med. 2003;50:708–716.