The effect of diffusion on quantitative BOLD parameter estimates acquired with the Asymmetric Spin Echo technique
Nicholas P Blockley1, Naomi C Holland1, and Alan J Stone1

1FMRIB Centre, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom

Synopsis

The results of a recently described streamlined quantitative BOLD (qBOLD) technique suggest that this method may overestimate the deoxygenated blood volume (DBV), leading to an underestimate of the oxygen extraction fraction (OEF). We hypothesise that this is due to the effect of diffusion, which is assumed to be zero in the analytical qBOLD model. In this study we performed Monte Carlo simulations to investigate this hypothesis and found that DBV and OEF measurements were vessel size dependent. However, the R2’ measurements which underlie qBOLD were found to be a reliable measure of deoxyhaemoglobin content above a vessel radius of 10μm.

Purpose

Recently a streamlined quantitative BOLD (qBOLD) technique was described to improve the robustness of measurements of the oxygen extraction fraction (OEF) and deoxygenated blood volume (DBV) by simplifying the analysis model1. This was achieved by removing confounding effects during data acquisition; enabling the number of model compartments to be reduced. However, measurements of DBV using this technique were larger than typically observed; potentially explaining why the measured OEF in these experiments was lower than expected. We hypothesise that this is due to the effect of diffusion, which is not accounted for in the qBOLD model where the static dephasing regime (SDR) is assumed2. The SDR assumption has been challenged by the results of numerical simulations of the Gradient Echo Sampling of Spin Echo (GESSE) pulse sequence that demonstrate a diffusion dependent shift in the R2′-weighted signal curve3 (R2′ - reversible transverse relaxation rate). However, the streamlined qBOLD approach uses the Asymmetric Spin Echo (ASE) pulse sequence, for which the effect of diffusion has not been examined. In this work we compare the effect of diffusion on the ASE technique versus the GESSE technique using Monte Carlo simulations. We then examine the implications of this effect on measurements of R2′, DBV and OEF using ASE.

Methods

Synthetic R2′-weighted qBOLD signals were generated following previously described Monte Carlo simulation approaches3,4. In brief, blood vessels were modelled as infinitely long cylinders, which were uniformly and randomly distributed in a spherical universe up to a defined volume fraction. The blood oxygenation of these vessels was modelled as an intravascular to extravascular susceptibility difference. A proton was allowed to randomly diffuse around this universe sampling the magnetic field perturbations due to the vessels. The magnetic field at the proton location manifests as a phase accrual, which was sampled every 200μs and saved every 2ms. This process of universe generation and proton phase accrual was repeated 10,000 times. Phases were added or subtracted to simulate the GESSE and ASE techniques and the sum of phases across all protons used to simulate the magnitude signal. The amount of R2′-weighting is determined by the spin echo displacement time τ. The range of τ values was matched for GESSE and ASE. Due to the low vessel volume fraction (~2-5%), protons spend the majority of their time in the extravascular space and as such these simulations model the extravascular signal (a first-order approximation of the experimentally observed signal).

This approach was used to investigate the effect of varying vessel radius in the range 1μm to 1mm, for a fixed diffusion coefficient4 D=1.3μm2ms-1. GESSE and ASE signals were compared for DBV=3%, OEF=40%, Hct=40%. The analytical qBOLD model1 (Fig. 1, Eq. 1) was used to estimate R2′, DBV and OEF from the simulated ASE data. Comparison was made with the SDR by setting D=0.

$$OEF=\frac{R_2^\prime}{DBV\;\gamma\frac{4}{3}\pi\;\Delta\chi_0\;Hct\;B_0}\tag{1}$$

Results

Fig. 2 compares the signal decay for the GESSE and ASE methods. As in previous work3 a shift in the GESSE signal curve is observed with decreasing vessel radius (Fig. 2a). This effect is not observed for the ASE signal, although the short τ signal shows increasing attenuation with decreasing vessel radius (Fig. 2b). The full analytical solution for the SDR is plotted as a black line with 1mm vessels approaching this regime. Fig. 3 displays the fitted qBOLD parameters as a function of vessel radius. When D=0 (SDR) the estimated parameter values are independent of vessel radius, whereas with diffusion all parameters show vessel radius dependence.

Discussion

Quantitative BOLD is a promising technique for the measurement of OEF using endogenous contrast. The simulations presented here reveal that the GESSE and ASE pulse sequences are both affected by diffusion, but in different ways. As hypothesised for ASE, diffusion is shown to result in an overestimation of DBV for typical parenchymal vessel radii (5-25μm), consistent with previous ASE based qBOLD measurements1,5. In contrast, R2′ measurements appear to be insensitive to vessel radius above a critical value (>10μm), providing a more robust parameter albeit sensitive to the product of DBV and OEF. With this in mind, it seems likely that the underestimation of OEF is due to the overestimation of DBV. Further work is required to understand the contribution of intravascular signal and vessels with multiple radii and blood oxygenation levels.

Conclusions

The effect of diffusion manifests as a vessel radius dependence of R2′, DBV and OEF measured using qBOLD with an ASE acquisition. However, by incorporating the results of these simulation in a modified qBOLD model, more accurate estimates of DBV may be possible and hence more accurate measurements of OEF.

Acknowledgements

This work was funded by EPSRC grant EP/K025716/1.

References

1. Stone AJ, Blockley NP. A streamlined approach to mapping the oxygen extraction fraction (OEF) and deoxygenated blood volume (DBV) using the quantitative BOLD technique. In Proceedings of the 23rd Annual Meeting of the ISMRM, Toronto, 2015. Abstract 219.

2. He X, Yablonskiy DA. Quantitative BOLD: Mapping of human cerebral deoxygenated blood volume and oxygen extraction fraction: Default state. Magn. Reson. Med. 2007;57:115–126.

3. Dickson JD, Ash TWJ, Williams GB, Harding SG, Carpenter TA, Menon DK, Ansorge RE. Quantitative BOLD: the effect of diffusion. J. Magn. Reson. Imaging 2010;32:953–961.

4. Boxerman JL, Hamberg LM, Rosen BR, Weisskoff RM. MR contrast due to intravascular magnetic susceptibility perturbations. Magn. Reson. Med. 1995;34:555–566.

5. An H, Lin W. Impact of intravascular signal on quantitative measures of cerebral oxygen extraction and blood volume under normo- and hypercapnic conditions using an asymmetric spin echo approach. Magn. Reson. Med. 2003;50:708–716.

Figures

Fig. 1 - Application of the qBOLD model is a multi-step process. R2′ is estimated from long τ data points. By extrapolating back to τ=0 from these data points, and comparing with the measured signal, DBV can be measured. Finally, R2′ is divided by DBV to estimate the OEF (Eq. 1).

Fig. 2 - R2′-weighted data can be acquired in multiple ways. Here the differing effect of diffusion on (a) the Gradient Echo Sampling of Spin Echo (GESSE) method and (b) the Asymmetric Spin Echo (ASE) method is examined. A shift in the signal curve is observed for GESSE, but not ASE.

Fig. 3 - Parameters extracted from the Asymmetric Spin Echo (ASE) signal curve using the qBOLD model as a function of vessel radius; (a) reversible transverse relaxation rate, (b) deoxygenated blood volume and (c) oxygen extraction fraction. A physiological diffusion coefficient (D=1.3μm2ms-1) was also compared with the static dephasing regime (D=0).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3731