Wietske van der Zwaag1,2, Mayur Narsude3, Olivier Reynaud2, Dan Gallichan2, and José P. Marques4
1Spinoza Centre for Neuroimaging, Amsterdam, Netherlands, 2EPFL, Lausanne, Switzerland, 3Lausanne, Switzerland, 4Donders Institute for Brain, Cognition and Behaviour, Nijmegen, Netherlands
Synopsis
3D-EPI-CAIPI was combined with a cylindrical excitation profile
to reduce the brain area from which signal is generated, and hence, the
parallel imaging undersampling artefacts and signal loss. This 3D-EPI-CAIPI
with cylindrical excitation can be used to acquire fMRI data with submillimetre
spatial resolution, 1-second temporal resolution and very high BOLD
sensitivity.Introduction
The maximum parallel imaging acceleration factor generally
depends on coil geometry, k-space trajectory and slice positioning. Unfolding
data becomes increasingly difficult with higher acceleration factors, resulting
in a g-factor penalty in (temporal) SNR. In functional MRI acquisitions with
3D-EPI, acceleration in both phase-encoding directions, but especially the
slab-selection direction, leads to shorter volume TRs (1,2) and can be highly
beneficial in increasing BOLD sensitivity.
In this study, we show that, using a 32-channel coil and
3D-EPI-CAIPI acquisition (3), a cylindrical excitation can be used to achieve
high undersampling factors (4x3), while maintaining high image SNR as well as tSNR
and hence, improved BOLD signal detection.
Methods
3D-EPI-CAIPI was combined with a 2D-RF
pulse (4) to selectively excite a cylinder. Placing the cylinder along the
read-out axis maximizes time benefits. Two protocols were compared in 4
volunteers at 7T, both cylindrical excitation and a standard slab selection:
-
a 2-mm isotropic acquisition (FOV=200*200*120mm,
TR/TRvol/TE/α=55ms /1.1s/27ms/18o, GRAPPA=3x2, ΔCAIPI=1, readout
gradient and cylinder along the left-right axis, cylinder radius 25mm, alias
22cm)
-
0.9*0.9*2.0 mm
acquisition (FOV=200*200*120, TR/TRvol/TE/α=55ms /1.1s/27ms/18o, GRAPPA = 4x3, ΔCAIPI=1;
same cylinder).
Cardiac and respiratory data were
collected for physiological noise removal using RETROICOR. An auditory stimulus
(5s natural sounds, 15s silence alternated for 5 minutes) was used to test BOLD
sensitivity in the primary auditory cortex.
Image SNR was estimated by dividing
the mean signal in a large mid-cylinder ROI (Figure 1, blue box) by the
standard deviation of the noise in an extra-cerebral ROI not affected by image
artefacts.
Results
Image SNR in the 2-mm isotropic data was 30±4 % (mean ±
stderr) higher in the cylindrical excitation data than in the equivalent slab
selection data. This improvement was
further emphasized on the 0.9mm data - 80±30 %. The difference in image SNR is
clearly visible in Figure 1, where both images are scaled to give equal
appearance to the noise in the background.
Higher image SNR translated also into
increases in tSNR values, of 35±4 % in the 2mm data and 23±2 % in the 0.9mm
data. Again, the tSNR is visibly improved in the tSNR maps in Figure 2.
The high tSNR led to highly significant BOLD responses in
the cylindrical EPI data, an example of which is shown in Figure 3. Note the
high temporal resolution, evident from the shown unfiltered time course of a
single voxel. Activation patterns were highly reproducible between subjects and
acquisition methods. No significant differences in number of significantly
(p<0.05, FWE) voxels were found, possibly due to the much changed number of
voxels included in the GLM mask.
Discussion
One of the main advantages of the 3D-EPI-CAIPI sequence is
its flexibility in optimising the acquisition for maximum spatial or temporal
resolution (1,2,3,5). The cylindrical excitation pattern could be used to
reduce the imaging FOV, thus accelerating the acquisition and allowing higher
spatial resolution, however, it is probably more advantageous to increase the
GRAPPA undersampling factor, which also leads to significantly shorter volume
acquisition times and can be done without incurring the usual undersampling artefacts
(g-factor noise amplification), as large parts of the imaging FOV do not
contribute with any signal that needs to be unwarped. The higher tSNR in the
0.9mm cylindrical acquisition data is a direct result from this improved
unfolding, as all parameters apart from the excitation profile of the pulse remained
the same between cylindrical and slab selection acquisitions.
Conclusion
Here, we successfully demonstrated that 3D-EPI-CAIPI with
cylindrical excitation can be used to acquire fMRI data with submillimetre
spatial resolution, 1-second temporal resolution and very high BOLD
sensitivity.
Acknowledgements
No acknowledgement found.References
1) Poser et
al 2009 MRM 2) Narsude et al 2014 MRM 3) Narsude et al 2015 MRM 4) Reynaud et
al 2014 MRM 5) Poser et al 2014 ESMRMB