Peng Cao1, Xiaoliang Zhang1, Ilwoo Park1, Chloe Najac1, Sarah J. Nelson1, Sabrina Ronen1, and Peder E. Z. Larson1
1Department of Radiology, University of California at San Francisco, San Francisco, CA, United States
Synopsis
This study aimed to
develop a lump-element double-tuned common-mode-differential-mode (CMDM)
radiofrequency (RF) surface coil with independent frequency tuning
capacity for MRS and MRI applications. This CMDM coil maintained intrinsically
decoupled magnetic fields, which provided sufficient isolation between the two
resonators. The results from in vivo experiments demonstrated high sensitivity
of both the 1H and 13C resonators.Purpose
This study aimed to
develop a lump-element double-tuned common-mode-differential-mode (CMDM)
radiofrequency (RF) surface coil
1-3 with independent frequency
tuning capacity for MRS and MRI applications. The presented double-tuned RF
surface coil design can be extended to a broad range of heteronuclear MRI
applications, including
13C,
23Na,
15N and
129Xe
and other hyperpolarized and non-hyperpolarized heteronuclear MR applications,
especially for those that require high local sensitivity.
Methods
Circuit analysis: The presented design
has two modes that operate with different current paths, allowing independent
frequency adjustment. According to Kirchhoff’s voltage law for two inner coil loops
and one auxiliary capacitor loop (in Figure 1a), the presented coil circuit can
be described as
$$\begin{cases}-j\omega L_{1}I_{1}+\frac{j}{\omega C_{S}}(I_{1}-I_{3})+\frac{j}{\omega C_{T1}}(I_{1}+I_{2})-j\omega L_{2}(I_{1}+I_{2})=0 \\-j\omega L_{1}I_{2}+\frac{j}{\omega C_{S}}(I_{1}+I_{3})+\frac{j}{\omega C_{T1}}(I_{1}+I_{2})-j\omega L_{2}(I_{1}+I_{2})=0 \\\frac{j}{\omega C_{T2}}I_{3}+\frac{j}{\omega C_{S}}(I_{3}+I_{2})+j\omega C_{S}(I_{3}-I_{1})=0\end{cases}$$Eq. 1
where
CS is the splitting capacitors, CT1 is the tuning capacitor
for differential mode and L1 and L2 are the equivalent
inductors on the outer loops and the central conductor, as defined in Figure
2a. Assuming the distribution of common mode (CM) current follows the
illustration of i1 in
Figure 1d (left), i.e., let I1 = I2 in Eq. 1,
the resonant frequency is given by
$$\omega_{CM}=\sqrt{\frac{\frac{1}{C_{S}}+\frac{2}{C_{T1}}}{L_{1}+2L_{2}}}$$Eq. 2
Utilizing
Kirchhoff’s voltage law and assuming the distribution of differential mode (DM)
current follow the illustration of i2
in Figure 1d (right), i.e., let I1 = -I2 in Eq. 1,
resonant frequency is given by
$$\omega_{DM}=\frac{1}{\sqrt{(\frac{C_{S}}{2}+C_{T2})2L_{1}}}$$Eq.
3
where
CT2 is the tuning capacitor for differential mode. The coil
prototype was tested on the bench and then examined in phantom and in vivo
experiments. Eqs. 2 and 3 clearly show that resonant frequencies of common mode
and differential mode can be adjusted independently via CT1 and CT2.
Bench test: For the tuning
dependency measurement, a network analyzer with full two ports (Agilent,
E5071C) was used. The frequency span was set as 100 kHz and the number of
points for frequency sweeping was 20000, allowing a nominal resolution of 5
Hz/point. The resonant frequency was detected automatically by searching the
minimum value of the magnitude S11 (which identifies the resonant mode) within
the span (i.e., 100 kHz), while the
impedance at the resonant frequency was measured on the Smith chart.
MRI protocol: All
MR experiments were performed on a 7T whole-body MRI scanner (GE
Healthcare, Waukesha, WI). The 1H-13C phantom was made
using a 21.7-mm-diameter cylinder syringe that was filled with ethylene glycol
(HOCH2CH2OH, anhydrous, 99.8%, Sigma-Aldrich, St. Louis,
Missouri). For the in vivo experiment, a three-plane and multi-slice GRE 1H
imaging was preformed, lasting 32 s. Non-selective dynamic spectroscopy data
(readout bandwidth = 5000 Hz and sampling points = 2048) were acquired
following a bolus injection of approximately 3 mL hyperpolarized [1-13C]-pyruvate
solution (80 mM), produced using SpinLab (GE Healthcare, Niskayuna, New York).
Results
Standard deviations of
frequency and impedance fluctuations measured in one resonator, while changing
the tuning capacitor of another resonator, were less than 13 kHz and 0.55 Ω.
The unloaded S21 was -36 dB and -41 dB, while the unloaded Q factor was 260 and
287, for
13C and
1H, respectively. In vivo hyperpolarized
13C MR spectroscopy
data demonstrated the feasibility of using the CMDM coil to measure the
3-second/spectrum-dynamics of lactate, alanine, pyruvate and bicarbonate signal
in a normal rat head along with fast acquiring
1H anatomical reference images.
Discussion
In the current study, we have
demonstrated the independent frequency tuning capacity of the developed
lump-element double-tuned CMDM coil. Using a circuit analysis, we demonstrated
that one tuning capacitor could be used to adjust the resonant frequency of one
mode without having current flowing in another tuning capacitor for the other
mode. The bench test result clearly showed that the two resonant frequencies
could be tuned independent of each other, verifying the circuit analysis. Compared with double-tuned quadrature
volume coils, the presented double-tuned surface coil provided relatively high
local sensitivity for heteronuclear MRI, but involved inhomogeneous B
1
fields.
Conclusion
Independent frequency
tuning capacity was demonstrated in the presented lump-element double-tuned
CMDM coil. This CMDM coil maintained intrinsically decoupled magnetic fields,
which provided sufficient isolation between the two resonators. The results
from in vivo experiments demonstrated high sensitivity of both the
1H
and
13C resonators.
Acknowledgements
The authors want to thank Lucas Carvajal for helping
on the MRI experiment. This work is supported by NIH grants
R00EB012064, R01EB016741, P01CA118816, P41EB013598, R01CA154915, R01CA172845
and R01EB008699.References
[1] Zhang X et al. Magn
Reson Med 2001;46(3):443-450;
[2] Zhang X et al. J Magn Reson
2003;161(2):242-251;
[3] Pang Y et al. Ieee T Med Imaging 2011;30(11):1965-1973.