Stephen J. Kadlecek1, Mehrdad Pourfathi1,2, Harrilla Profka1, and Rahim R. Rizi1
1Radiology, University of Pennsylvania, Philadelphia, PA, United States, 2Electrical and Systems Engineering, University of Pennsylvania, Philadelphia, PA, United States
Synopsis
Hyperpolarized imaging sequences are subject to potential bias and spread of apparent signal to neighboring voxels due to the time-dependence of magnetization as RF pulses are applied and as polarization is lost through spin-lattice relaxation. In this abstract, we discuss a method to detect and correct for artifacts and demonstrate it using chemical shift imaging. The method utilizes periodic resampling of nonselective spectra to correct for signal dynamics. We show that this technique results in a higher fidelity, "de-blurred" image.Introduction
All methods to image hyperpolarized agents
in vivo are subject to potential bias,
artifact and blurring due to the time-dependence of magnetization as RF pulses
are applied and as polarization is lost through spin-lattice relaxation. These problems are magnified when the longer
pulses or acquisition periods, characteristic of spectroscopic imaging, are
used. In this abstract, we discuss a
method to detect and correct for artifacts using chemical shift imaging.
Materials and Methods
To compare excitation and analysis methods, a healthy mouse
was imaged using four injections of 80mM 300 μL hyperpolarized 1-
13C pyruvate. Imaging was performed using a Bruker Avance
400 vertical bore imaging spectrometer and a 25mm ID
1H/
13C
saddle coil (Bruker). The neutral,
isotonic agent was prepared using a Hypersense (Oxford Instruments). A cylindrical sodium lactate phantom (OD=3mm,
4M) was placed at one side of the animal bed.
The sedated animal (isoflurane) was monitored for respiration and controlled
for temperature (37°)
using a rectal probe.
Acquisitions 1 and 3 utilized chemical shift imaging with a
progressive flip angle as per [1], beginning with α=3.58° and increasing to 90° over 256 acquisitions in an effort to maintain constant
signal intensity. Acquisitions 2 and 4 used a constant flip angle of α=8° for each excitation.
Other parameters were common to all acquisitions:
SW=5000 Hz, AT=51.3ms, TR/TE=53/0.47ms, FOV=25x25mm
2,
ST=8mm, axial, covering both kidneys, MS=16x16 for an in-plan resolution of
1.56x1.56mm
2. Images were
separated by approximately 1 hour to allow for sample polarization. k-space was sampled using a custom outward
spiral trajectory which was modified such an additional k=0 (in-plane
nonselective) spectrum was acquired after every 10 k≠0 encodings. This
allowed agent and metabolite time-dependences to be observed independently. Multislice reference proton images with
ST=1mm and the same FOV were acquired between
13C images. Metabolite maps were reconstructed by
integrating individual peaks in each voxel, and interpolated to the 256x256
grid of the proton image.
Results and Discussion
Fig. 1 shows the pyruvate and lactate peak integrals for the
24 k=0 acquisitions spaced throughout the 13.6s of imaging time. The progressive flip-angle spectra (top
traces) maintained more consistent peak integrals, although due to in/outflow
and ongoing label exchange, significant trends were observed that differed by
chemical species. This highlights the
fact that no acquisition scheme can account for signal dynamics when multiple
species are being studied.
Figs. 2 and
3
show the reconstructed CSI spectra reconstructed from constant and progressive
flip-angle acquisitions overlaid on a 1mm proton image positioned at the center
of the
13C slice.
Signal-to-noise of the agent in an individual voxel is at most ~30, as
compared to ~12 and 8 for the metabolites lactate and alanine. The most robust metabolism was observed in the
small intestine, with lesser activity seen in the kidneys and skeletal
muscle. Because of the larger signal
near k=0, the constant flip-angle images display superior signal-to-noise, but
visible blurring due to loss of signal for large k acquisitions.
This effect is more clearly observed in
figs. 4 and
5, which
show pyruvate and lactate maps, respectively.
The top row is reconstructed directly from the localized spectra of
figs. 2 and
3. In the bottom row, the
time-dependence of signal amplitude is eliminated by dividing each acquisition
by the metabolite-specific curves of
fig. 1 (“de-blurring”). In this way, the underlying structure is
clarified (see, e.g., the ring-like metabolite structure of the left kidney,
and the highly heterogeneous metabolism in the intestine; both are observed in independent
images, and signal-to-noise is sufficient to ensure that neither is artifact.
Conclusions
We have introduced a method to observe and correct for
signal variability during an extended imaging protocol using hyperpolarized
agents. Notably, the observed signal
dynamics requires an independent correction for each chemical species, and this
correction minimizes blurring due to motion or relaxation-induced dynamics. This technique may be generalized to other
sequences; e.g., multi-echo sequences in which the dynamics of T
2
relaxation must be accounted for or understood.
It may be that investigators will choose to tolerate some blurring to
increase signal near k=0, but this method of periodic resampling will
nonetheless allow an understanding of blurring artifacts and aid in image
interpretation.
Acknowledgements
No acknowledgement found.References
[1]
Yen Y-F, et al., Magn Reson Med 62:1-10 (2009)