Shiloh Sison1, Xin Huang2, Shi Feng3, Ji Chen2, Richard Williamson3, and Gabriel Mouchawar3
1St. Jude Medical, Sunnyvale, CA, United States, 2University of Houston, Houston, TX, United States, 3St. Jude Medical, Sylmar, CA, United States
Synopsis
E-fields tangential to implantable cardiac pacemakers and defibrillators
were analyzed at Normal and First Level Controlled Operating Modes as well as
30uT peak B1+. 1.5T body transmit and head and extremity transmit RF coils were
simulated. A comparison of the Etans generated by the body transmit coils is
made to those generated by head and extremity transmit coils. If RF heating and
voltage can be shown to be safe with unrestricted body coils then no additional
analysis is necessary for head and extremity coils since there is a 9x margin
in RF heating and 3x margin in voltage.PURPOSE
When subjected to the RF field of an
MRI, a cardiac pacemaker or defibrillator may encounter heating of the lead
electrode and/or voltage at the device header causing unintended stimulation of
cardiac tissue. The RF field may be generated by a transmit body coil, head
coil or extremity coil. Due to the size, power and positioning restrictions of
head and extremity coils, the fields generated are much smaller than those
generated by body coils. The purpose is to demonstrate that evaluation of transmit
head and extremity coils is unnecessary for cardiac pacemakers or defibrillators
that are shown to be safe with body transmit coils.
METHODS
The tangential component of electrical
field (Etans) were extracted along cardiac pacemaker and defibrillator lead
pathways in five body models (FATS, Duke, Ella, Billie and Thelonious [2]) in 217
body coil simulations and 133 head and extremity simulations. For the body
coils the simulations were performed at landmarks from the top of the head to
the lower extremities in 10cm increments.
The body coil diameters ranged from 60cm
to 73cm, lengths from 35cm to 73cm with high pass (HP), low pass (LP) and band
pass (BP) configurations. LP and HP head coils with diameters from 27cm to 31cm
and lengths from 27cm to 38cm were simulated. HP and LP upper and lower
extremity coils with diameters from 14cm to 22cm and lengths 14cm to 22cm were
also simulated. The simulations were validated through 60 E-field measurements taken
in an ASTM phantom filled with 1.2S/m media centered in a 1.5T body coil.
The average value of Etan along the
length of the implant was calculated and scaled to Normal Operating Mode, First
Level Controlled Mode, and 30uT as defined in [3]. The 90th, 99th, 99.9th and
99.99th percentile average Etan was extracted from the distribution. This
analysis was performed for the body coil simulations and then repeated for the
combination of head and extremity coils simulations.
The Ratio2 was calculated
from (Etan_body/Etan_HE)2
since temperature rise and power deposition are proportional to Etan2.
The Ratio was calculated from (Etan_body/Etan_HE) since header voltage is
proportional to Etan. The average Ratio2 was calculated across the
percentiles for Normal Operating Mode and repeated for First Level Controlled
Mode. The average Ratio was calculated across the percentiles for 30uT values.
RESULTS
Validation of the Simulated E-fields is shown in Figure 1. The measurements and
simulations were normalized. For the measured E-field values ≥ 10 V/m, the
standard deviation of the errors between the normalized measured and normalized
simulated values was 2.59%.
Etan comparison between body coils and head/extremity
coils are shown in Table 1.
DISCUSSION
The Ratio2 averaged across the percentiles for normal
operating mode is 26. This means that, as a first order approximation, a
temperature rise or power deposition calculated using body coils will be need
to be divided by 26 to reach the levels generated by head and extremity coils
alone. This increases to 37 for first level controlled mode. With an additional
safety factor of three to account for second order effects there is still a 9x
margin for normal operating mode and 12x for first level controlled mode.
The Ratio averaged across the percentiles for 30uT is 10. This means
that, as a first order approximation, a header voltage calculated using body
coils will need to be divided by 10 to reach the levels generated by head and
extremity coils alone. With an additional safety factor of three to account for
second order effects there is still a 3x margin for the header voltage.
CONCLUSION
If RF heating and voltage can be shown to be safe with unrestricted body
coils for cardiac pacemaker and defibrillators pathways then no additional
analysis is necessary for head and extremity coils since there is a 9x margin
in RF heating and 3x margin in voltage.
Acknowledgements
No acknowledgement found.References
[1] ISO/TS 10974:2012 Assessment of the safety of magnetic resonance
imaging for patients with an active implantable medical device
[2] Christ A, Kainz W, Hahn
EG, Honegger K, Zefferer M, Neufeld E, Rascher W, Janka R, Bautz W, Chen J,
Kiefer B, Schmitt P, Hollenbach HP, Shen J, Oberle M, Szczerba D, Kam A, Guag
JW, Kuster N. The Virtual Family – Development of surface-based anatomical
models of two adults and two children for dosimetric simulations. Physics in
Medicine and Biology 2009:55:N23-N28.
[3] IEC 60601-2-33:2010+AMD1:2013+AMD2:2015 CSV
Medical electrical equipment -
Part 2-33: Particular requirements for the basic safety and essential
performance of magnetic resonance equipment for medical diagnosis