Daniel J Martire1, Krzysztof Wawrzyn1, William Bradfield Handler1, and Blaine A Chronik1
1Physics and Astronomy, Western University, London, ON, Canada
Synopsis
In this project, a method to quantify the
gradient-induced vibration of a medical device in an MR system using a laser
Doppler vibrometer is presented. A copper annulus was suspended in a 3T
scanner, at a position chosen to maximize the x-component of the x-gradient.
The displacement of the device was measured at different gradient field
strengths. Typical observations were understood, which should lead towards
helping establish a test standard for gradient induced vibrations of implanted
medical devices.Purpose
In the complex electromagnetic environment of an MRI
scanner, any kind of conducting material can experience vibration due to the
forces on eddy-currents induced in the conductor by the rapid switching of
gradient fields. Such vibration could be dangerous or cause discomfort to the
patient.
1 Graf et al. describe a method to measure time-varying
torque of a stabilized device, and ISO/TS 10974 gives an overview of good lab
practice when conduction measurements of vibration.
2,3 Currently,
there does not exist a standardized method to measure device vibration
quantitatively. In this project, a method to quantify the gradient-induced
vibration of a medical device in an MR system using a Doppler laser vibrometer
is presented along with an investigation of its various systematic errors.
Methods
All vibration measurements were conducted in a 3T MRI system
(Siemens Prisma, Robarts Research Institute, Western University), with a
non-contact laser-Doppler vibrometer system (OFV-505/5000, Polytec GmbH,
Germany). A copper annulus with an outer diameter of 10 cm, inner diameter of 7
cm, and thickness of 3 mm was suspended with at various positions and
orientations in the scanner bore, chosen to maximise the x-component of the
magnetic field for the x-gradient. (See Fig 1.). The gradient waveform used was
a repeated trapezoidal with rise time of 0.5 ms, flat top time of 100 ms, fall
time of 0.5 ms, and off time of 500 ms.
Vibration was recorded with no gradients running, to measure
ambient noise of the environment. Furthermore, movement of the apparatus
(without the device attached, and with gradients running) was measured to
determine vibration caused mechanically or acoustically by the scanner. The
displacement of the device was then measured at gradient field strengths of 10,
20, 30 and 40 mT/m, driving only the x-gradient coil. A db/dt probe was placed
within several centimeters of the device in order to capture a reference signal
indicating the timing and slew rate. This step is critical for the proper
analysis of the signals.
Additionally, the device was struck while in the
bore to determine its behaviour (without driving gradients). Next, the device
was twisted about its vertical axis to observe the effect of the elastics
returning the device to equilibrium. Finally, in the lab, the device was struck
with a hammer.
Results
A typical measurement is shown in figure 2 along with the corresponding
gradient sequence. The device moves on the order of hundreds of microns when
the gradient ramps-up and down as the alternating magnetic moment of the device
attempts to align itself with the static field. There are two other forms of
vibration present as well. One is a low-frequency continuous oscillation
between 0 and 50 microns. The other is a fast underdamped harmonic oscillation
the device experiences immediately after the gradient coil finishes a ramp. Figure
3 shows a close-up of the device as the gradient field ramps-up to 30 mT/m in
0.5 ms. The annulus was displaced a mean distance of 223 microns and a maximum
distance of 363 microns.
Figure 4 shows the comparison of the annulus being
struck in the lab to inside the bore, while figure 5 shows how the magnitude of
the vibration changes with slew rate.
Discussion
The vibrations shown in figure 4 are very similar. Both
signals resemble that of an underdamped harmonic oscillator. In the bore, the
time constant is -553 and has a frequency of 730 Hz. In the lab, the time
constant is -25 and has a frequency of 84 Hz. The difference in frequency is
explained by an increased elastic restoring force, due to induced eddy
currents.
The lower frequency vibration is similar to the vibration
within the holding elastic observed in the labs, with no twisting.
The magnitude of vibration increases linearly with slew rate
as would be expected. The error was determined from a combination of background
noise present in the system, contribution from other sources of vibration, and from
the variance of the measurements.
Conclusions
To assess the safety of a device, it is necessary to
understand its displacement caused by the gradients ramping, resultant internal
vibrations, and the motion of the device with respect to the measurement
holder. With the method presented, it should be a straightforward next step to
develop a standardized test to measure gradient-induced vibration of any
medical device intended for use inside an MRI. A better understanding of this
phenomenon will aid in the design of medical devices so as to minimize the
vibrations induced.
Acknowledgements
I would like to acknowledge the members of the xMR lab and machine shop at Western University for their helpful contributions to this project. I also acknowledge NSERC for the funding required to complete this project.
References
1. Hartwell, R. C., & Shellock, F. G. (1997). MRI of
cervical fixation devices: sensation of heating caused by vibration of metallic
components. J Magn Reson Imaging, 7(4), 771–772. http://doi.org/10.1002/jmri.1880070427
2. Graf, H., Lauer, U. a, & Schick, F. (2006).
Eddy-current induction in extended metallic parts as a source of considerable
torsional moment. Journal of Magnetic Resonance Imaging, 23(4),
585–90. doi:10.1002/jmri.20539
3. ISO/TS
10974 Ed. 1 (2012), Assessment of the safety of magnetic resonance imaging for
patients with an active implantable medical device; International Standards
Organization Technical Specification, www.iso.org.