Volkan Acikel1 and Daniel B Ennis1,2
1Department of Radiological Sciences, University of California Los Angeles, Los Angeles, CA, United States, 2Department of Bioengineering, University of California Los Angeles, Los Angeles, CA, United States
Synopsis
MRI presents several potential risks for
patients with implanted devices and one of the main concerns is RF induced heating
of devices and nearby tissues. The aim
of this study was to compare the RF induced heating characteristic of wires at 1.5T
and 3T using simulations and direct measurements. Greater heating (SAR
amplification) can be observed at lower field strengths for longer wires.Introduction
MRI presents several potential risks for patients with implanted
devices and one of the main concerns is RF induced heating of devices and
nearby tissues. The aim of this study was
to compare 1.5T and 3T RF induced heating at the tip of a wire as a surrogate
for an implanted medical device.
Methods
The temperature increase at the wire-tip is related
to the tangential component of the incident E-field along the wire path. For a uniform
incident E-field, wire-tip heating has previously been formulated using the Modified
Transmission Line Method (MoTLiM)[1]:
$$\Delta T \approx C |\frac{E_0}{k_t} tan(k_t l_w /2 )|^2 \quad \quad \ \ \ \ \ [1]$$
where kt is the wavenumber, lw is the
lead length, E0 is the incident E-field, and C is the constant for EM
and thermal calculations needed to find the E-field from hypothetical voltage
(VHyp) and temperature from the SAR. Note that Equation 1 comes from
VHyp, which is a scaled version of the charge distribution along the
lead, defined by MoTLiM and assumes scattered fields decay fast and establish a
quadratic relationship between VHyp at the wire-tip and the
temperature rise (∆T). For straight
wires the wavenumber is a linear function of frequency. Consequently, higher
field strengths may cause lower temperature increases when the E-field is kept
constant. To better understand the
potential for temperature increases (∆T) the constant C (Eqn. 1) was set to
unity and the frequency range was evaluated between 64MHz and 128MHz (equivalent
to exposure at 1.5T to 3T) square of the VHyp is plotted.
Bare copper wires (1cm to 40cm and 0.5mm radius) were used to
evaluate the resonance behavior of the wire and compared with computational electromagnetics
simulations using the Method of Moments (MoM) (FEKO, Altair Engineering) and a 1V/m
incident E-field. For the simulations
and MoTLiM comparisons the relative permittivity and conductivity of the medium
was 70 and 0.42S/m, respectively, for all frequencies.
The temperature rise at the wire-tip was also tested with
MRI experiments at both 1.5T (Avanto, Siemens) and 3T (Prisma, Siemens). A bare wire with radius 0.4mm was used to achieve resonance behavior at a
shorter length, and immersed in an aqueous solution (14g/L HEC). In order to achieve a relatively uniform
E-field distribution along the wire, the wire was placed on a circular path
with radius 12.5cm inside a circular phantom with radius 16cm. The length of the wire length was changed
from 14cm-40cm (2cm increments) at 1.5T and 7cm-22cm (1cm increments) at 3T. Each wire was exposed to 4 W/kg whole-body SAR
for a duration of 10-minutes. The temperature
rise at the wire-tip was measured directly with a fiber optic temperature
probe. SAR at the wire-tip (SARw) was calculated from the initial
slope of the temperature data. After measuring temperature rise for each wire
length, a reference temperature measurement was acquired without the wire and the
incident SAR (SARi) was calculated from the slope of the reference
temperature rise. The calculated SAR-gain
due to the wire was found as the ratio of SARw/SARi.
Results
Figure 1 shows a comparison of SAR-gain at the wire-tip
obtained from MoM (red circles) compared to calculations made with MoTLiM (black
circles). Results are plotted by
normalizing to their maximum values. The results show excellent agreement
between MoM and MoTLiM over a wide range of wire lengths and field
strengths. The results also highlight
that greater heating (SAR-gain) can be observed at lower field strengths for
longer wires.
In Figure 2 the SAR-gain measured during the MRI experiments
is shown for both 1.5T (blue stars) and 3T (red squares). The results show the
maximum SAR-gain is higher (~1300) at 1.5T for longer wires (26cm) compared to
a maximum SAR-gain of 750 at 14cm at 3T. Note that the SAR-gain at 1.5T is
higher than 3T for all wires >16cm. Note the excellent agreement between
simulation and experimental results.
Discussion
RF induced wire-tip heating at 1.5T and 3T field
strengths were compared. To understand
the wire-tip heating (SAR-gain) results the resonance length of the wires must
be considered. While keeping the wire length constant as the frequency increases,
the SAR-gain can increase or decrease depending on the resonance behavior of
the wire. However, when the temperature
rise for different field strengths is compared at the resonance peaks
(different wire lengths), it can be seen that SAR-gain decreases with the
frequency.
Conclusion
Greater heating (SAR-gain) can be observed at lower field
strengths for longer wires as evident in both simulations and direct
measurements at 1.5T and 3T.
Acknowledgements
This work was support by NIH/NHLBI R21-HL127433 to DBE and
the Department of Radiological Sciences at UCLA.References
[1] Volkan Acikel, Ali Uslubas, and Ergin Atalar.
"Modeling of electrodes and implantable pulse generator cases for the
analysis of implant tip heating under MR imaging." Medical physics 42.7
(2015): 3922-3931.