Antonio Javier Gonzalez1, Luis Fernando San Sebastian1, Sebastian Stanculovic2, Reineiry Emilio Garcia3, Ralph Wissmann2, Sven Junge2, and Jose Maria Benlloch1
1Institute for Instrumentation in Molecular Imaging, i3M, Valencia, Spain, 2Preclinical Imaging, Bruker BioSpin, Ettlingen, Germany, 3Institute of Design and Manufacture, Valencia, Spain
Synopsis
A novel approach for radio-frequency shielding using hybrid carbon fiber materials and slotted copper layers. A proof of concept has been built and tested in high frecuency MR scanners. The purpose is to use it to shield the MR RF field on PET electronics, allowing simultaneous PET-MR acquisitions.Purpose
Simultaneous information of MR and molecular Positron Emission Tomography
(PET) imaging, especially in the pre-clinical small animal field, is strongly
being currently investigated. This implies the working compatibility of both
modalities, MR and PET. On one hand, the homogeneity of the B0 and B1 MR fields
should not be compromised by either ferromagnetic components in the PET
electronics such as Nickel or eddy currents generated in the possible RF
screen. On the other hand, PET electronics have to properly be RF shielded. In
this work we present a novel RF shielding design for a new development of a PET
insert for pre-clinical MR systems. We propose a hybrid approach merging Carbon
Fiber (CF) structures to shield the B1 field and slotted copper layers (SCL) to
reduce eddy currents produced by the switching gradient coils.
Methods
A single PET ring made out of 8 detectors blocks using monolithic
crystals and arrays of 16x16 SiPMs has been designed. The inner and outer
diameters are 116 mm and 196 mm, respectively. An inner tube to the PET
together with lateral rings has been made out of CF using aluminum molds. Packs
of two or three unidirectional CF layers of 200 um each have been tested. The unidirectional
CF layers are always at 90o one to each other. In one case one CF layer
was aligned to the axial axis of the PET and MR, and in another case at 45o
(see Figure 1). The CF structures are prepared in
out-of-autoclave curing process (16h at 70oC), using an exhaustive
control of compaction levels through LVDT (Linear Variable Differential
Transformer), as a quality control to ensure the best possible mechanical (Tensile
Strength: 54 MPa Flexural Strength: 80 MPa Compressive Strength: 142 Mpa),
electrical and toughness properties. The SCL screens are only used on the
outer to the PET cylinder. The copper layers are of 35 um thickness. Slits of 2
mm wide are traced along the axial axis of the cylinder at 30 mm gaps. A second
identical copper layer separated by 100 um FR4 from the first and 16 mm
shifted, avoids the penetration of the RF field, minimizing eddy currents.
These structures have been tested in the Bruker BioSpec 70/30 system
equipped with a BGA 12S gradient system. The standard 112/86 RF coil was also
used for reference measurements.
EPI ghosting acquisitions using 300 kHz bandwidth,
TR = 2000ms, TE = 37 ms, SpinEcho mode, automatic ghost correction, 16 slices
with 1 mm thickness and 2 mm gap were carried out on both the CF and SCL
structures.
Results
Concerning RF shielding, tests were carried out on the two CF layers
directions (0 and 45o) using a network analyzer. They are comparable
with standard SCL shields and there is almost no difference between the two CF
tubes types. Although the
RF shielding of CF tubes is sufficient for the use in PET/MR systems, a direct
comparison of the RF shielding
properties of CF tubes with the SCL structure has shown that there is small
difference due to the lower electrical conductivity of CF layers (compared to
the copper).
The EPIs on the SCL have been acquired using the standard 112/86
resonator with a ghosting uniform along all slices. On the CF screens there was
a variation visible along the slices.
The appearance of eddy currents while using the different RF shields
surrounding the coil has also been inspected by PRESS tests. The lowest
influence is visible for the reference coil. A slight influence could be
detected for both CF shields, which was larger for the 45o
configuration.
Discussion and conclusions
This work shows a novel RF shielding for PET inserts developments based
on a hybrid approach. The optimum RF shielding cavity resulted of CF laminates
and slotted copper layers. The inner tube of the cavity and the lateral rings
are made out of CF, which shielded the RF as good as simply copper. However,
because this material showed some eddy currents originated by the gradient
switching coils, the outer cylinder of this structure uses a double SCL
separated by 100 um FR4. Lateral CF rings and exterior SCL tube are in contact
using conductive rubbers and foams.
In addition to these results, the PET performance
has already been tested outside the MR with unprecedented results.1
Tests inside the MR using the proven hybrid RF shielding are scheduled for late
November 2015.
Acknowledgements
No acknowledgement found.References
1 A.J.
González, et al., “A PET detector ring with homogenous spatial resolution in
the presence of a magnetic field”, IEEE NSS-MIC Conf. Record 2015, San Diego (USA).