A simple digital MRI system using a digital oscilloscope
Makoto Tsuda1 and Katsumi Kose1

1University of Tsukuba, Tsukuba, Japan

Synopsis

A digital MRI system was developed using a digital oscilloscope that had a waveform generator function. An RF pulse with 43.85 MHz Larmor frequency was successfully output using this oscilloscope and the NMR signal was captured using the undersampling technique (144 ns dwell time, 6.94 MHz sampling frequency). The acquired a set of 3D MR image using a 1.0 T permanent magnet MRI system demonstrated the validity of our approach.

Introduction

Various approaches to digital MRI systems have been reported [1-4]. Most of them used a Field Programmable Gate Array (FPGA) to generate radio frequency pulses and detect NMR signals. However, it is very time consuming and difficult for many people to develop such a sophisticated digital system. In 2014, we developed a simple digital MRI system that used a digital oscilloscope, an arbitrary waveform generator, and three 32-bit small board computers [5]. In this study, we have successfully further simplified the system using a digital oscilloscope that can generate the Larmor frequency RF pulse.

Materials and Methods

Our digital MRI system consists of a digital oscilloscope (PicoScope 5242B, Pico Technology, St Neots, UK), three 32-bit board computers (Arduino Due, Smart Project, Torino, Italy), and a laptop PC (CPU Intel Core i5, 2.60GHz) (Fig.1). The digital oscilloscope has the following specifications: 125 MS(mega sample)/s for two-channel 15-bit mode input and 200 MS/s for one-channel 14-bit output. This oscilloscope can be used for both RF pulse generation and RF signal detection (Fig.2). Figure 3 shows the block diagram of our digital MRI system. One of the board computers (master system) controls the total timing of the MRI pulse sequence (20 ms time resolution) and outputs trigger signals to other board computers and the oscilloscope. The board computers output waveforms of Gx, Gy, and Gz field gradients synchronously via each DA output to the gradient driver. We made the Larmor frequency (43.85 MHz) RF signal by writing the 43.85 MHz sinusoidal signal to the wave memory of the oscilloscope with 14-bit resolution and 5 ns dwell time in 1000 word length. The RF pulse was output to the RF transmitter for excitation. The MRI signal detected by the RF coil is amplified with a low noise preamplifier and supplied to the oscilloscope through the low-pass filter (50MHz). To reduce the amount of the sampling data and speed up the data transfer and processing time, the under sampling technique (dwell time = 144 ns, sampling frequency ~ 6.94 MHz) is used [4]. Imaging experiments were performed using a 1.0 T MRI system using a yokeless permanent magnet with a 90 mm gap [6]. A 3D gradient echo sequence (TR = 200 ms, TE = 11 ms, image matrix = 256×256×16) was used for image acquisition of a water phantom. The phase of the receive signals were corrected using that of the excitation RF pulses because the external trigger timing generated by the board computer was not synchronized with the RF oscillator time base.

Results and Discussion

Figure 4 shows a 2D cross section of a water phantom selected from a 3D image dataset acquired with a 3D gradient echo sequence (TR = 200 ms, TE = 11 ms, Image matrix = 256×256×16). Because the phase correction was successfully performed, no ghosting artifact was observed. In conclusion, because the total cost of the digital instruments is about $2,000 and the time required for the system development is about one year, our system can be a promising approach to development of a digital MRI system.

Acknowledgements

No acknowledgement found.

References

1. S. Jie, et al, Rev. Sci. Instrum. 76, 105101 (2005). 2. P. Stang, et al., IEEE Trans Med Imaging. 31(2): 370 (2012). 3. S. Hashimoto, et al, Rev. Sci. Instrum., 83, 053702 (2012). 4. P. Pe’rez, et al, Medical Engineering & Physics 26 523 (2004). 5. M. Tsuda, et al., 23th ISMRM, Toronto, Canada. p1843 (2015). 6. T. Shirai, et al, Magn Reson Med Sci 4,137 (2005).

Figures

Fig.1 Overview of the digital MRI system. A: Digital oscilloscope with an arbitrary waveform generator, B: 32-bit board computer (Arduino Due), C: Laptop PC.

Fig.2 The digital oscilloscope used for RF pulse generation and NMR signal detection. The trigger signals are supplied by the board computer used for MRI pulse sequence generation

Fig.3 The block diagram of the digital MRI system. Three board computers are used for timing generation of the MRI pulse sequence and gradient waveform. The digital oscilloscope is used for RF pulse generation and NMR signal detection.

Fig.4 2D cross section of a water phantom selected from a 3D image dataset acquired with a 3D gradient echo sequence (TR = 200 ms, TE = 11 ms, Image matrix = 256×256×16).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3634