Steffen Weiss1, Jochen Keupp1, and Edwin Heijman2
1Tomographic Imaging, Philips Research, Hamburg, Germany, 2Oncology Solutions, Philips Research, Eindhoven, Netherlands
Synopsis
MR-HIFU therapies require stable
PRFS-based temperature monitoring over the treatment duration (>30min).
Drift effects may be corrected using independent temperature information e.g.
based on T2 or T1 maps and simultaneously obtained using fast interleaved
sequences. As interleaves may impact PRFS-accuracy, a dynamic on-off-interleaved
sequence pattern is introduced to selectively measure their influence on the
PRFS phase. The mechanism of the phase variations is explored using multiple
variants of the interleaved sequence. It is shown that eddy currents induced
during the interleaved sequences represent the main cause of the phase
variation. The spatio-temporal characteristics of the effect are explored as a
first step towards compensation.Objective
MR guided high-intensity focused
ultrasound (MR-HIFU) has been established as a treatment option that elegantly
combines two non-invasive technologies. It is performed for ablation,
hyperthermia and hyperthermia-mediated local drug delivery and commonly
monitored by proton resonance frequency shift (PRFS)-based sequences.
Especially hyperthermia requires a temperature accuracy of 1K for at least
30min. Baseline drift and other phase errors (e.g. by motion) may be corrected
using an additional sequence to measure absolute temperature information.
Recently, ultra-fast interleaving of arbitrary sequences was demonstrated on
clinical scanners [1] and applied to interleaved PRFS- and T2-based temperature
mapping [2]. Fast interleaved sequences may compromise the accuracy of the PRFS
method, e.g. via spin effects or residual eddy currents. Here, various
methodological effects of such fast interleaved sequences on the PRFS phase are
examined that could limit temperature accuracy.
Methods
The influence of interleaving on
the PRFS phase was examined by MRI of a spherical phantom (270mm diameter) using
the body-coil of a 1.5T scanner (Philips Achieva). Individual dynamics of a
multi-slice T1w-FFE-EPI PRFS sequence (three slices acquired in 5.4s,
TE/TR=19.5ms/55ms, 288² reconstruction matrix, (1.4mm)² x 4mm voxels) were interleaved
with TFE-shots of a saturation-prepared Look-Locker sequence (single slice,
TFE-shot duration of 1.5s, 10 phases per shot). In order to identify the effect
of interleaving, a dedicated on-off pattern was introduced (Fig.1, top). For
each dynamic, the average phase in ROIs of 8x8 pixels was calculated per slice.
To quantify the contribution of the interleaved sequence to the phase variation,
a square wave model was fitted to the average phase values as a function of the
dynamic scan number (Fig. 1, bottom). The amplitude p0 of the square wave was
evaluated for different experimental conditions to explore the mechanism of the
observed phase variation. These included the variation of the slice position of
the interleaved sequence relative to the PRFS slices, the time order of the
acquisition of the PRFS slices, an extra delay before the PFRS sequence,
comparison of the effect in gel and water to test for any magnetization
transfer effects, and omitting all RF and/or all gradients for the interleaved
sequence.
Results
A high value for p0 was only
found in the PRFS slice acquired first in time after the interleave (Fig.2).
However, a variation of the position of the interleaved slice relative to the
PRFS slices did not change the p0 value markedly. A comparison of measurements
in gel and water samples of about equal T1 did not result in any marked
differences either. The phase amplitude p0 vanished when all gradients of the
interleaved sequence were switched off, but p0 remained unchanged when omitting
the RF (Fig.3). Initial delay experiments performed for a selected ROI revealed
an exponential decay of p0 (τ ≈
500ms). An evaluation of p0 for ROIs across the entire phantom revealed a
pronounced position dependence (Fig. 4a), which required careful re-evaluation
of various earlier experiments. The repetition of this experiment with delays
showed that the position dependence is also time dependent (Fig 4 b,c,d). Deliberate
detuning of the eddy current compensation of the MR system resulted in a largely
amplified effect (Fig 5).
Discussion
Initial experiments had been designed
to test the hypothesis that residual magnetization excited by the interleaved
sequence contributes to the PRFS phase variations, which however proofed wrong.
The gradient on/off experiment clearly hinted to an effect caused by eddy
currents. Clinically relevant temperature changes may be mimicked if this
effect is not considered (a phase error of 100mrad corresponds to a temperature error of 1.3K). The observed phase errors depend on gradient design and parameters
that influence the time between the interleave and the acquisition of the
k-space center of a PRFS slice, which provides options for optimization. All
measurements so far have been performed on the same 1.5T system, and further
experiments are required to explore whether the effect requires calibration per
sequence, per system or per system type. Potentially, a further refined eddy
current compensation technique can be used to minimize the effect.
Conclusion
Fast interleaving of additional
sequences with PRFS temperature mapping can result in phase errors due to eddy
currents induced by the interleaved sequence that mimic clinically relevant
temperature changes. Further work is required to find adequate techniques to
either minimize the effect or to calibrate for it.
Acknowledgements
No acknowledgement found.References
1. Henningsson
M, et al. Magn Reson Med 2015;73:692–696
2. Keupp J,
et al. Proc. ISMRM 2015, p4061.