Samuel Fielden1, Matthew Geeslin2, Xue Feng1, Wilson Miller2, Kim Butts Pauly3, and Craig Meyer1,2
1Biomedical Engineering, University of Virginia, Charlottesville, VA, United States, 2Radiology & Medical Imaging, University of Virginia, Charlottesville, VA, United States, 3Radiology, Stanford University, Palo Alto, CA, United States
Synopsis
A
rapid volumetric MR thermometry sequence, using a stack-of-spirals acquisition,
was implemented on a real-time platform in order to support animal model
experiments. The performance of the sequence, as measured by the maximum and
mean recorded temperature as well as the sequence accuracy, was assessed
against a clinically used 2DFT sequence in vivo. The use of efficient spiral
trajectories supports rapid generation of volumetric thermometry maps and
allows the visualization of the entire insonication as well as correlation with
post-ablation imaging of lesion size and location.Introduction
Real time MR thermometry, usually based on the
proton-resonance frequency shift, is a key aspect of MR-guided focused
ultrasound (MRgFUS) procedures [1]. The desire to monitor the entire insonicated
volume has led the field towards the development of rapid, 3D methods [2];
however, acquiring fully sampled 3D volumetric data to monitor heating is time
consuming, and so fast methods must be developed in order to meet the spatial
and temporal requirements for adequate monitoring of thermal therapy.
The data acquisition
efficiency of spiral trajectories is higher than that of Cartesian scanning.
Therefore, spiral trajectories are an attractive way to improve temporal
resolution while maintaining spatial resolution in MR thermometry [3-6]. Previously,
we have implemented spiral trajectories for 2D and 3D thermometry and compared trajectory
performance in terms of the focal spot size and position shift versus a
Cartesian acquisition [5,6]. The purpose of this study was to
implement the sequence on a real-time platform and demonstrate its preliminary
effectiveness in a more clinically-relevant set-up using an in vivo porcine
model for MRgFUS thalamotomy.
Methods
A 3D retraced
spiral-in/out (RIO) thermometry sequence was implemented on the RTHawk platform
(HeartVista, Inc.) to enable real-time sequence control and monitoring of a FUS
insonication. RTHawk interfaced with a GE Discovery MR750T 3T scanner at the
UVA Focused Ultrasound Center, where an Insightec ExAblate 650 focused
ultrasound transducer was used to induce focal heating.
Three
Yorkshire pigs were used for initial sequence evaluation. Non-ablative power
settings were used to compare the temperatures measured by the 3D sequence to
those measured by the standard 2D GRE thermometry sequence provided by Insightec.
Briefly, a target within the thalamus was identified, and then insonicated at
low power with monitoring performed by either the 2D or 3D method. After an
appropriate cool-down period, the same spot was insonicated again with the same
power settings and monitored by the alternate thermometry technique. Finally,
ablative power settings were used to create a lesion at each spot, again
monitored by either the 2D or 3D method. All insonications were 20 seconds
long. Maximum temperature, measured by the single hottest pixel, and mean
temperature, measured as the average of the 9 pixels centered at the hottest
pixel in a single plane, were recorded. Temperature standard deviation of each
thermometry image set was measured as the mean standard deviation over time of
a group of 9 pixels located within the brain away from the focal spot. After
FUS, the animals were moved into a head coil array for high quality
post-ablation assessment, which included T2-, Diffusion-,
and T2*-weighted
imaging using standard sequences.
Sequence
parameters for 3D RIO thermometry were: FA = 10°, TR/TE = 20.1/12.8 ms, readout
length = 8 ms, interleaves = 24 over a FOV of 280 mm2 for an
in-plane resolution of 1.5 mm2. 3D phase encodes = 16 with
through-plane resolution of 2 mm, so that through-plane FOV was 32 mm. Total
acquisition time per volume was 7.7 seconds. In one animal, a spectral-spatial
pulse was used for water-only excitation in order to reduce artifacts stemming
from fat, which necessitated an increase of TR to 22.7 ms and temporal update
interval of 8.7 seconds. Sequence parameters for the 2D thermometry method
were: TR/TE = 27.6/12.8 ms, pixel bandwidth = 44 Hz/px, matrix size 256 x 128
over a FOV of 280 mm2 for a resolution of 1.1 x 2.2 mm2, slice width of 3 mm, and update
interval 3.5 seconds.
Results
Real-time in vivo thermometry images are shown
in Fig. 1. The hot spot is resolved in volumetric space, albeit with lower SNR
with the spiral 3D sequence, due in part to reduced voxel size. In Fig. 2,
post-ablation 3D T2-weighted
images are shown along with an overlay of one of the 3D temperature maps
showing good correspondence between the two. Scatter plots of the maximum and
mean measured temperatures by each sequence is presented in Fig. 3. The maximum
temperatures measured by the 3D sequence correlate well with those measured by
the 2D sequence. The mean temporal standard deviations of the 2D and 3D
sequences were 0.8°C and 1.3°C, respectively.
Conclusions
The
efficiency of spiral readouts supports rapid generation of 3D temperature maps.
In vivo, we have successfully monitored the entire ablation with adequate
spatial resolution to qualitatively compare the ablation area with temperature.
We have achieved 1.3°C temperature accuracy with 8.7s temporal resolution. Future
work will focus on improving these specifications through the use of surface
coils and temporal acceleration methods [5].
Acknowledgements
UVa-Coulter Translational Research PartnershipReferences
[1]
Rieke V, Butts Pauly K. JMRI 2008;27:376390. [2] Todd N, et al. MRM 2012;67:724-730. [3]
Stafford, et al. MRM 2000; 43:909-912. [4] Josan, et al. ISMRM 20;1802. [5]
Fielden, et al. ISMRM 22;2346. [6] Fielden, et al. ISMRM 23;1631.