The effect of sonication duration on ablation depth during MR-guided focused ultrasound of bone: acute findings with MR in a swine model
Matthew Bucknor1, Rutwik Shah1, Eugene Ozhinsky1, and Viola Rieke1

1Radiology and Biomedical Imaging, University of California, San Francisco, San Francisco, CA, United States

Synopsis

One challenge in magnetic resonance guided focused ultrasound ablation of bone lesions is extending the ablation zone deep to the cortical surface in order to cover the full intramedullary extent of the lesion. Extension of the ablation zone is important for complete treatment of for example, osteoid osteomas, and improved local control of metastatic bone disease. This study used a swine model to compare differences in the depth of the ablation zone between two protocols using short versus longer sonications, respectively, while maintaining the same overall sonication energy (by adjusting acoustic power).

Purpose

Magnetic resonance guided focused ultrasound (MRgFUS) has shown significant promise as a non-invasive therapy for treatment of focal bone lesions such as bone metastases and osteoid osteomas (1-3). One of the challenges in clinical treatments of bone lesions is extending the depth of ablation sufficiently deep to the cortical surface to permit ablation of the entire intramedullary extent of a lesion. When sound waves reach a bone-soft tissue interface, there is rapid attenuation secondary to reflection, scattering, and mode conversion, in addition to absorption with approximately 60-80% loss of acoustic energy (4). We have previously demonstrated that repetition of sonications within a given volume can significantly increase the depth of an ablation beyond the cortical surface (5). Anecdotally, in clinical practice, many MRgFUS interventionalists additionally use strategies such as decreasing the sonication frequency or increasing the duration of the sonication (while lowering the acoustic power) to achieve a deeper ablation. The purpose of the current study was to investigate whether or not for a specific energy dose there was a difference in ablation depth beyond the cortical surface with long versus short sonications.

Methods

All experimental procedures were done in accordance with National Institutes of Health guidelines for humane handling of animals and received prior approval from the local Institutional Animal Care and Use Committee. We performed MRgHIFU using an ExAblate® 2000 system (InSightec, Haifa, Israel) integrated with a 3.0 Tesla MR scanner (GE Healthcare, Waukesha, WI, USA) in a swine model. Multiplanar T2-weighted fast spine echo (FSE) and LAVA 3D spoiled gradient echo images were acquired for treatment planning. Two discrete ovoid treatment targets were prescribed at each proximal diaphysis and distal diaphysis of the left femur. Six sonications were used to create each lesion at 1.0 MHz with a phased array transducer of 208 elements embedded within the scanner table. During the course of the treatment, the energy level was maintained at 400 J for all sonications. For the proximal sonications, the acoustic power was set at 10 W with sonication duration of 40 seconds. For the distal sonications, the acoustic power was set at 20 W with sonication duration of 20 seconds. Average increased temperature at the bone-soft tissue interface was approximately 76 °C for the shorter sonications and 66 °C for the longer sonications. Post-treatment MR imaging was performed immediately following the treatment with delayed contrast enhanced segmented inversion recovery fast gradient echo (TR/TE/flip angle=1.5ms/15.2ms/15), LAVA 3D spoiled gradient echo sequences. A dose of 0.1mmol/kg Gd-DTPA was delivered 10min before imaging to assess the ablation size.

Results

MRgFUS was used to create two focal ablation lesions within the left femur. Delayed contrast enhanced MR images demonstrated ovoid regions of hypoenhancement, which correlated with the bone targets. The depth of ablation was greater for the shorter high power sonications (7 mm), compared to the longer lower power sonications (5 mm) (Figure 1). The other two dimensions of the ablation were also greater for the shorter sonications (26 x 16 mm, craniocaudal by anteroposterior), compared to the longer sonications (11 x 11 mm).

Discussion/Conclusion

Interestingly, this basic science study suggests that during MRgFUS of bone, increasing the duration of the sonication while maintaining the same total acoustic energy did not result in deeper ablation volumes. In fact, the shorter high power sonications demonstrated relatively increased depth and overall ablation size. Increased depth of bone ablation during clinical treatments which is anecdotally attributed to the use of longer sonications is possibly instead related to simultaneous modifications of additional technical parameters (e.g. an uncompensated decrease in acoustic power, for overall higher energy sonication; or decreased sonication frequency). This study importantly helps us to better understand the factors influencing the size of the ablation zone during MRgFUS bone treatments.

Acknowledgements

The authors would like to thank Mr. Vahid Ravanfar RT (MR) and Ms. Carol Stillson, BA, VT for their technical support in imaging and animal preparation.

References

1. Napoli A, Mastantuono M, Cavallo Marincola B, et al.: Osteoid Osteoma: MR-guided Focused Ultrasound for Entirely Noninvasive Treatment. Radiology 2013; 267:514–521.

2. Gianfelice D, Gupta C, Kucharczyk W, Bret P, Havill D, Clemons M: Palliative treatment of painful bone metastases with MR imaging--guided focused ultrasound. Radiology 2008; 249:355–363.

3. Hurwitz MD, Ghanouni P, Kanaev SV, et al.: Magnetic resonance-guided focused ultrasound for patients with painful bone metastases: phase III trial results. J Natl Cancer Inst 2014; 106.

4. Sikov MR: Effect of ultrasound on development. Part 1: Introduction and studies in inframammalian species. Report of the bioeffects committee of the American Institute of Ultrasound in Medicine. J Ultrasound Med 1986; 5:577–583.

5. Bucknor MD, Rieke V, Do L, Majumdar S, Link TM, Saeed M: MRI-guided high-intensity focused ultrasound ablation of bone: evaluation of acute findings with MR and CT imaging in a swine model. J Magn Reson Imaging 2014; 40:1174–1180.

Figures

Coronal (with respect to the femur) MR post-treatment images showing the ablation sites. (a) The longer duration lower power ablation along the proximal aspect of the left femur (arrow) demonstrates decreased intramedullary extent relative to (b) the shorter duration higher power ablation along the distal aspect of the left femur (arrow).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3608