Matthew Bucknor1, Rutwik Shah1, Eugene Ozhinsky1, and Viola Rieke1
1Radiology and Biomedical Imaging, University of California, San Francisco, San Francisco, CA, United States
Synopsis
One
challenge in magnetic resonance guided focused ultrasound ablation of bone
lesions is extending the ablation zone deep to the cortical surface in order to
cover the full intramedullary extent of the lesion. Extension of the ablation
zone is important for complete treatment of for example, osteoid osteomas, and
improved local control of metastatic bone disease. This study used a swine
model to compare differences in the depth of the ablation zone between two
protocols using short versus longer sonications, respectively, while
maintaining the same overall sonication energy (by adjusting acoustic power).Purpose
Magnetic resonance guided
focused ultrasound (MRgFUS) has shown significant promise as a non-invasive
therapy for treatment of focal bone lesions such as bone metastases and osteoid
osteomas (1-3). One of the challenges in clinical treatments of bone lesions is
extending the depth of ablation sufficiently deep to the cortical surface to
permit ablation of the entire intramedullary extent of a lesion. When sound
waves reach a bone-soft tissue interface, there is rapid attenuation secondary
to reflection, scattering, and mode conversion, in addition to absorption with
approximately 60-80% loss of acoustic energy (4). We have previously
demonstrated that repetition of sonications within a given volume can
significantly increase the depth of an ablation beyond the cortical surface
(5). Anecdotally, in clinical practice, many MRgFUS interventionalists
additionally use strategies such as decreasing the sonication frequency or
increasing the duration of the sonication (while lowering the acoustic power)
to achieve a deeper ablation. The purpose of the current study was to
investigate whether or not for a specific energy dose there was a difference in
ablation depth beyond the cortical surface with long versus short sonications.
Methods
All experimental procedures
were done in accordance with National Institutes of Health guidelines for
humane handling of animals and received prior approval from the local
Institutional Animal Care and Use Committee. We performed MRgHIFU using an
ExAblate® 2000 system (InSightec, Haifa, Israel) integrated with a 3.0 Tesla MR
scanner (GE Healthcare, Waukesha, WI, USA) in a swine model. Multiplanar
T2-weighted fast spine echo (FSE) and LAVA 3D spoiled gradient echo images were
acquired for treatment planning. Two discrete ovoid treatment targets were
prescribed at each proximal diaphysis and distal diaphysis of the left
femur. Six sonications were used to create each lesion at 1.0 MHz with a phased
array transducer of 208 elements embedded within the scanner table. During the
course of the treatment, the energy level was maintained at 400 J for all
sonications. For the proximal sonications, the acoustic power was set at 10 W
with sonication duration of 40 seconds. For the distal sonications, the
acoustic power was set at 20 W with sonication duration of 20 seconds. Average
increased temperature at the bone-soft tissue interface was approximately 76 °C
for the shorter sonications and 66 °C for the longer sonications.
Post-treatment MR imaging was performed immediately following the treatment
with delayed contrast enhanced segmented inversion recovery fast gradient echo
(TR/TE/flip angle=1.5ms/15.2ms/15), LAVA 3D spoiled gradient echo sequences. A
dose of 0.1mmol/kg Gd-DTPA was delivered 10min before imaging to assess the
ablation size.
Results
MRgFUS was used to create
two focal ablation lesions within the left femur. Delayed contrast enhanced MR
images demonstrated ovoid regions of hypoenhancement, which correlated with the
bone targets. The depth of ablation was greater for the shorter high power
sonications (7 mm), compared to the longer lower power sonications (5 mm)
(Figure 1). The other two dimensions of the ablation were also greater for the
shorter sonications (26 x 16 mm, craniocaudal by anteroposterior), compared to
the longer sonications (11 x 11 mm).
Discussion/Conclusion
Interestingly, this basic
science study suggests that during MRgFUS of bone, increasing the duration of
the sonication while maintaining the same total acoustic energy did not result
in deeper ablation volumes. In fact, the shorter high power sonications
demonstrated relatively increased depth and overall ablation size. Increased
depth of bone ablation during clinical treatments which is anecdotally attributed
to the use of longer sonications is possibly instead related to simultaneous
modifications of additional technical parameters (e.g. an uncompensated
decrease in acoustic power, for overall higher energy sonication; or decreased
sonication frequency). This study importantly helps us to better understand the
factors influencing the size of the ablation zone during MRgFUS bone
treatments.
Acknowledgements
The authors would like to thank Mr. Vahid Ravanfar RT (MR) and Ms. Carol Stillson, BA, VT for their technical support in
imaging and animal preparation. References
1. Napoli A, Mastantuono M, Cavallo
Marincola B, et al.: Osteoid Osteoma: MR-guided Focused Ultrasound for Entirely
Noninvasive Treatment. Radiology 2013; 267:514–521.
2. Gianfelice D, Gupta C, Kucharczyk W, Bret P, Havill D,
Clemons M: Palliative treatment of painful bone metastases with MR
imaging--guided focused ultrasound. Radiology 2008; 249:355–363.
3. Hurwitz MD, Ghanouni P, Kanaev SV, et al.: Magnetic
resonance-guided focused ultrasound for patients with painful bone metastases:
phase III trial results. J Natl Cancer Inst 2014; 106.
4. Sikov MR: Effect of ultrasound on development. Part 1:
Introduction and studies in inframammalian species. Report of the bioeffects
committee of the American Institute of Ultrasound in Medicine. J Ultrasound
Med 1986; 5:577–583.
5. Bucknor MD, Rieke V, Do L, Majumdar S, Link TM, Saeed M:
MRI-guided high-intensity focused ultrasound ablation of bone: evaluation of
acute findings with MR and CT imaging in a swine model. J Magn Reson Imaging
2014; 40:1174–1180.