Improved Thermometry Based on a Fast Spin Echo Sequence
Yuval Zur1

1GE Healthcare, Haifa, Israel

Synopsis

A time shifted Fast Spin Echo (FSE) thermometry sequence which is insensitive to B0 inhomogeneity is used in conjunction with restricted FOV module and parallel imaging to reduce scan time and increase the number of slices. Up to 5 slices are acquired in 3 sec. This sequence works well on a clinical Focused Ultrasound brain system. The temperature signal to noise ratio (TSNR) of this FSE sequence is approximately 2 times higher than the conventional Gradient Echo (GRE) sequence used today. 3 to 5 slices are acquired in 3 sec, rather than 1 slice with GRE.

Introduction

MR‑guided Focused Ultrasound (MRgFUS) is used to treat non-invasively many brain disorders. Currently gradient echo (GRE) is used for thermometry imaging using the PRF shift. Recently we presented a Fast Spin Echo (FSE) thermometry method (1) which is insensitive to B0 inhomogeneity. In this abstract we present improvements to this method such that it fulfils/exceeds imaging requirements and demonstrate an advantage over GRE.

Background

The MRgFUS system consists of a large semi-spherical helmet filled with circulating cooling water. The receive coil is a two water-proof loops immersed in the water bath. The GRE sequence is constrained to TE of 12.4 msec with minimum scan time and spatial resolution of 3.3 sec and 2 mm respectively with a single slice. The FSE sequence (Fig. 1) is a FSE with the 90° pulse shifted by T sec. Two shots are acquired with phase cycling (1). A PRF frequency shift Δf translates to a phase shift ΔΦ according to Eq. [1]:

$$\Delta\Phi (FSE) = 4\pi \cdot \Delta f \cdot T; \space\space\space\space \Delta\Phi(GRE)=2\pi\cdot \Delta f\cdot TE \space\space\space\space\space\space\space\space [1]$$

Temperature SNR (TSNR) of FSE is higher because 1) T can be increased without affecting scan time and 2) FSE is twice more sensitive than GRE for a given Δf, TE and T.

Method

To enable a long echo train we modulate the RF flip angles along the echo train (2). ky lines are acquired in a centric order, i.e. for echo 1, 2, 3, 4, 5 … etc. ky values are 0, 1, -1, 2, -2, 3, -3 … etc. To avoid image blurring the echo train length (ETL) is limited to 50 echoes. During reconstruction the even and odd echoes are separated (1) and the phase ΔΦ between them (Eq. [1]) is determined. The simulated even/odd echo amplitude vs. echo number is shown in Fig. 2a. The signal oscillations generate artifacts over many echoes. We use an algorithm (beyond abstract scope) to remove the oscillation, leaving a smooth decay (Fig. 2b). Data acquisition starts at echo 2 with minimal decay during the echo train. Optimal TSNR is obtained at T ≈ T2*. We use T = 25 msec which is much longer than the 12.4 msec TE in GRE.

Reduced Field of View: To reduce scan time and shorten echo train length (reduce blurring), we use a small FOV along the phase direction because the treated area requires an FOV of about 10 cm. To prevent aliasing the signal beyond the reduced FOV is saturated. The signal saturation module is insensitive to B1 and B0 inhomogeneity. Details are described in another abstract. By reducing the number of ky lines a full image is acquired with a single echo train of 25 to 50 echoes and scan time of 2*TR. 3 to 5 slices are acquired in 3 sec with a TR of 1200 to 1500 msec.

Parallel Imaging (PI): Further speedup is possible with PI, but because we have only 2 coils acceleration R is limited to 1 < R < 2. For parallel imaging we use GRAPPA recon (TGRAPPA (3)). Fig. 3 demonstrates k-space ordering for R = 2, where the ky lines are divided into blocks of 2 lines arranged in a centric order. In time frame 1 line 1 in each block is acquired, in time-frame 2, line 2. Similarly, any acceleration R = p/q can be obtained by using blocks of p lines and acquiring q lines in a block. Best results were obtained for p = 3 and q = 2, i.e. R = 3/2.

Results

We have acquired successfully many thermal images on a brain MRgFUS system at 1.5T. Fig. 4 shows temperature data before, during and after heating from 45 time-frames. Parameters: 128x192 matrix, FOV = 24 cm, reduced FOV factor = 0.38, ETL = 50 echoes, with 3.0 sec/frame and 3 slices. T = 25 msec. To compare TSNR between GRE and FSE accurately we acquired GRE and FSE images on a conventional 1.5T scanner with a 4 channel cardiac coil. A phase shift ΔΦ was created by shifting the center frequency by Δf = 15 Hz. TSNR is the ratio between the mean and standard deviation of the phase images. The results are listed in Table 1.

Conclusion

The results in Table 1 indicate a 1.9: 1 TSNR advantage of FSE over GRE with the same voxel size. FSE scans are faster with 3 to 5 slices in each time-frame. Reduced FOV provides any desired spatial resolution with significant speedup. FSE temperature imaging works well on a real MRgFUS system.

Acknowledgements

Goes to Alex Volovick for help with the focused ultrasound experiments.

References

(1) Y. Zur, Proceedings, ISMRM 2015, p. 4054.

(2) R. F. Busse et al. Effects of refocusing flip angle modulation and view ordering in 3D Fast Spin Echo, Mag. Res. Med. 2008; 60: 640 - 649.

(3) F. A. Breuer et al., Dynamic autocalibrated parallel imaging using temporal GRAPPA (TGRAPPA), Mag. Res. Med. 2005; 53: 981 - 985.

Figures

Fig. 1: Time shifted Thermal FSE.

Fig. 2: Even/odd echo amplitudes in Thermal FSE: the oscillations in (a) are corrected in (b).

Fig. 3: ky layout vs. echo number for PI with R = 2: even lines are acquired in frame 1 and odd lines in frame 2.

Fig. 4: Temperature (°C) rise and cool down vs. time at the focal spot of the MRgFUS brain system.

Table 1: Results from scans on a conventional machine with 4 channel cardiac coil. FSE/GRE TSNR ratio for the same voxel size is ~1.9. In case of PI, the number of slices increases because ETL decreases. Scan time is 3 sec and T = 25 msec.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3598