Sascha Krueger1 and Tim Nielsen1
1Philips GmbH, Innovative Technologies, Research Laboratories, Hamburg, Germany
Synopsis
4D image data are used in radiation therapy planning
to estimate motion of the tumor or regions at risk. Today 4D CT is commonly
used for this purpose. Due to better soft tissue contrast and quantitative and
functional imaging, a clinical demand for MRI in therapy planning exists. Consequently,
there is also a growing interest in 4D MRI techniques. A scan-time-efficient 4D
MRI method utilizing the MRI Navigator as motion sensor with high geometric fidelity is proposed.Introduction
4D image data are used in radiation therapy
planning to estimate motion of the tumor or regions at risk. While today 4D CT is
commonly used for this purpose, MRI offers better soft tissue contrast and
quantitative and functional imaging. Therefore MRI is increasingly used in therapy
planning and there is also a growing interest in 4D MRI
techniques. 4D MRI may be of additional relevance for repeated inter-fraction
re-planning. A prospective 4D MR imaging approach based on a
highly-efficient 4D MRI method was presented recently [1,2] where the standard
respiratory belt sensor is used as motion sensor to infer the tissue motion. However, correlation of sensor and actual tissue motion is a major concern in these techniques. Therefore, an implementation based on the
MRI pencil beam navigator approach is proposed here. A refined MR Navigator method
was developed as accurate, reliable and robust motion sensor in 4D MRI for therapy-planning independent of anatomy or scan
geometry.
Methods
As in [1,2] the volume of interest is covered by a
stack of 2D image slices with an image of each slice and respiratory
phase being acquired within a few hundred milliseconds (e.g. using single-shot
TSE or TFE) to obtain a respiratory-cycle-resolved 4D data set based on a short
initial respiratory signal calibration phase. As in [1] we utilized a highly
efficient scheme with dynamic choice of phase and slice and dynamic reduction of
sequence inherent idle time. For highest robustness manual interaction was
integrated into the scanning workflow: If navigator-measured tissue position drifts out of the
acceptance window of the inital calibration phase [1], a decision dialog will be presented whether to instruct the subject to
breathe normally again and proceed with scanning, to finalize the scan
prematurely in case all statistically significant motion
states have already been captured or to restart the acquisition. Even if incomplete, the obtained 4D data can
be imported into existing therapy planning systems.
The 4D MR navigator methos was refined to trigger robustly and
accurately independent of scan and navigator geometry, the most prominent issue being MR Navigator slice saturation artifacts, i.e. artifacts resulting immediately
after a slice was acquired in case the slice intersects the MR navigator:The algorithm estimates the
severity of the expected artifact based on the intersection angle and excludes a definable number of navigator
acquisitions for triggering to let the MR navigator regain its steady state and
to get rid of eventual saturation artifacts.
27 volunteer subjects were included in a feasibilty study. A navigator acquisition rate of
20Hz was used. Different geometries and target organs were tested to verify
robustness of the method.
MR navigator
triggered scans were acquired (FOV 375 (FH) x 260 (AP) mm2, 32 slices, slice thickness 5 mm, resolution 1.5mm x 1.5 mm, single-shot TSE, TR 6400 ms, TE 77 ms, echo spacing 5.2 ms, flip angle 90°, refocussing angle 120°, 10 phases, shot duration 330 ms) and compared to
respiratory-belt triggered scans with same parameters for scan efficiency,
image quality and overall robustness. All scans were performed on a 3T Ingenia
MR system (Philips Healthcare, Best, The Netherlands) using the digital anterior and
posterior receive coils.
Results
and Discussion
The navigator method allows robust and accurate
triggering in various tested scan and navigator geometries using the saturation prediction
algorithm (Figure 1). 4D
MR images with high geometric consistency could be generated in a volunteer
study with 27 subjects (Figure 2). The mean scan time
for a 4D scan with 10 respiratory phases and 32 slices was (334+-36)s. It was found that navigator-based triggering
results in more geometrically consistent 4D MRI images than possible with the
respiratory belt. The quality difference can be especially high in cases where the belt signal correlates poorly with the internal motion which was found reproducibly in certain volunteers with frequent switching between breathing modes or generally less differentiatable breathing pattern (Figure 3). Another advantage of the proposed technique is the faster workflow and the ability to place the MR navigator pencil beam close to the target anatomy. The latter results in 4D scans with highest possible geometric fidelity which is crucial for therapy planning.
Conclusion
An efficient and precise prospective 4D MRI method was presented
using a refined MR-based respiratory navigator as motion sensor to accurately track the tissue motion at the target anatomy. The main usage scenarios of this 4D MRI technique are EBRT planning. The technique may be also of interest for combined MR-EBRT or MR-HIFU systems.
Acknowledgements
No acknowledgement found.References
[1]
Nielsen and Krueger, Proc. ISMRM 2016, #3997
[2] Hu et al., Int. J. Radiat.
Oncol. Biol. Phys. 86, 198-203 (2013)