Measurement and Simulation of Susceptibility Artifacts in Variable-TE Radial MRI: Application in an MR-safe Guidewire
Katharina E. Schleicher1, Stefan Kroboth1, Klaus Düring2, Michael Bock1, and Axel Joachim Krafft1,3,4

1Dept. of Radiology - Medical Physics, University Medical Center Freiburg, Freiburg, Germany, 2MaRVis Medical GmbH, Hannover, Germany, 3German Cancer Consortium (DKTK), Heidelberg, Germany, 4German Cancer Research Center (DKFZ), Heidelberg, Germany

Synopsis

A simulation framework is presented to optimize radial acquisition schemes with variable echo times which are designed to minimize the directional anisotropy of the artifact of an MR-safe guidewire. The simulation results are compared to measurements. We could theoretically and experimentally verify that the homogeneity of the artifact can be improved via the variable-TE method.

Introduction

Guidewires are essential tools for intravascular interventions – unfortunately, conventional metallic guidewires can cause excessive heating during MRI procedures.1 Recently, MR-safe guidewires have been developed from a glass fiber reinforced structure with embedded iron particles as local passive markers. The susceptibility artifacts of these markers depend strongly on the echo time (TE) and on the orientation against B0.2 Recently, a radial acquisition scheme with variable TEs was proposed yielding a more homogeneous artifact.3 In this work a simulation framework was implemented to gain insight into the artifact behavior and, thus, to further optimize radial acquisition schemes.

Material and Methods

Simulation Framework

In the simulation framework the intra-voxel dephasing of a magnetic distortion field Bz (e.g. from the markers of the guidewire) is calculated. MR signals are computed from the magnetization on a 15-fold denser ‘spin’-grid in each spatial direction compared to the intended imaging resolution. For each grid location, the magnetization evolution during data acquisition is simulated, and all grid values are finally summed up to form the simulated k-space. Signal values of ‘spins’ which are spatially located within the guidewire material or which are spectrally located outside the range of the slice-selection process due to influence from Bz are set to 0 while the remaining ones are initialized as 1. The simulation is performed for conventional Cartesian sampling as well as for radial acquisitions with constant and variable TEs.

To simulate the artifacts of an MR-safe guidewire (MaRVis Medical GmbH, Hannover, Germany), the guidewire was modeled as a chain of spherical iron particles (size: 8±2µm) placed along its central strand (mean particle distance along guidewire direction: 50±30µm, radius of central strand: 60µm). Particle sizes and distances along the guidewire followed a Gaussian random distribution whereas the axial variations of the particle position (perpendicular to guidewire orientation) followed a uniform random distribution. Each iron particle was represented by a point-like dipole field so that the total Bz can be calculated from the summation of individual dipole fields:

$$B_z^{~'}=\sum_i\frac{\mu_0m_z}{4\pi}\cdot\frac{2(z-z_i)^2-(x-x_i)^2-(y-y_i)^2}{((x-x_i)^2+(y-y_i)^2+(z-z_i)^2)^{5/2}}$$

Here, (xi,yi,zi) reflects the position of the ith dipole. The magnetic moment mz of the particles is assumed to be aligned with B0 and cal­cu­lated from the particle volume and the sa­turation magnetization of iron.

Variable-TE Acquisition Scheme

In the radial acquisition with variable TEs, a different TEn is used for every spoke n so that TE is shortest (TEmin=1.8ms) for readout-direction parallel to B0 and longest (TEmax=10.8ms) perpendicular to B0 to minimize the directional anisotropy of the dipole artifact. For all other directions, two different variation schemes are implemented: (1) Sine-modulation 3: $$$\mathrm{TE}_n=\mathrm{TE}_\mathrm{min}+(\mathrm{TE}_\mathrm{max}-\mathrm{TE}_\mathrm{min})\cdot\vert\mathrm{sin}(\phi_n)\vert$$$, and (2) Sine-squared modulation: $$$\mathrm{TE}_n=\mathrm{TE}_\mathrm{min}+(\mathrm{TE}_\mathrm{max}-\mathrm{TE}_\mathrm{min})\cdot\mathrm{sin}^2(\phi_n)$$$. The latter scheme is chosen as the inverse of the cos2-modulated dipole-field.

For both schemes, simulations were carried out with the following parameters: FOV=270x270mm2, object-size=180x250mm2, BW=610Hz/px, slice-thickness=5mm, matrix=128x128. The acquisition was simulated for 201 radial spokes (angular range: 2π) with 256 readout points per spoke (2-fold oversampling). Images were reconstructed using the NUFFT algorithm.4 Simulations were based on a GPU implementation 5 and remaining computation was done in MATLAB.

For comparison, images of the guidewire were acquired at 1.5T (Siemens Symphony) using the same parameters as in the simulation (TR=14ms, α=15°). Both, simulations and measurements, were also carried out using a conventional Cartesian acquisition as reference. To measure the artifact width, profiles perpendicular to the wire were defined and the FWHM was computed for sections of the wire perpendicular and parallel to B0 (Fig. 2).

Results and Discussion

Measured and simulated Cartesian (Fig. 1a,e) and radial (Figs. 1b,f) images consistently show the expected anisotropic artifact for constant TE and a more uniform artifact pattern for the variable-TE imaging strategies (Figs. 1c,d,g,h). Quantitatively, simulated artifact widths were systematically smaller (15-25%) whereas the artifact ratios (rw) are in agreement which means that our simulations correctly describe the artifact behavior. Only minor differences in rw were found for the sin- and sin2-modulated acquisitions, although the artifact visually appears more defined in the sin2-case.

Even though the simulation systematically underestimates the absolute artifact width, simulations proved as a useful tool for sequence optimization as reflected in a correct reproduction of the artifact ratios. To further improve the simulation, iron particles could be simulated more realistically, e.g. as cluster-forming ellipsoids. Based on the simulations, further optimizations such as different TE modulation schemes, variation of TEmax and TEmin, and variable-TRs can be easily tested. This will facilitate the development of efficient data acquisition strategies for improved guidewire visualization.

Acknowledgements

Grant support from the Deutsche Forschungsgemeinschaft (DFG) under grant number BO 3025/2-2 is gratefully acknowledged.

References

1. Nitz et al. JMRI 2001;13:105-14.

2. Rubin et al. Invest Radiol 1990;25:1325-32.

3. Krafft et al. ISMRM 2015, 1662.

4. Fessler et al. IEEE T-SP, 51(2):560-74, Feb. 2003.

5. Kroboth, Master‘s Thesis, TU Graz, 2013.

Figures

Figure 1: Measured (a-d) and simulated (e-h) guidewire images. (a,e) Cartesian images with constant TE=5ms. Radial images (b,f) with constant TE=5ms, (c,g) with variable TE and |sin|-modulation, and (d,h) with sin2-modulation.

Figure 2: In (a) blue and red areas indicate the regions wherein artifact widths were measured. (b) Two exemplary, spline-interpolated profiles for the artifact perpendicular (blue) and parallel (red) to B0. Black circles indicate measured data points. The horizontal lines indicate the half maximum levels of the peaks, respectively.

Table 1: Values and uncertainties of the artifact widths for both measurement and simulation. The associated ratios rw (artifact ⊥ to B0 vs. artifact ∥ to B0) are given in the last row.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3582