Fangfang Tang1, Fabio Freschi1,2, Maurizio Repetto1,2, Feng Liu1, and Stuart Crozier1
1School of Information Technology and Electrical Engineering, The University of Queensland, Brisbane, Australia, 2Department of Energy, Politecnico di Torino, Torino, Italy
Synopsis
In
MRI, gradient coil switching generates eddy currents in surrounding conductors including
intra-coil eddy currents in the surrounding gradient coils (crosstalk). In
order to investigate the intra-coil eddy currents, a set of gradient coils with
differing track widths has been designed for use in a cylindrical MRI system. It
was found that the surrounding coils with wide tracks produced significantly
larger eddy currents than the cold shield. It is therefore necessary to take
into account the crosstalk between coils when evaluating coil performance. We
found that the optimal track width to use in gradient coil design is around 14 mm. Purpose
To
study inductive interactions between gradient coils and to find the optimal
track widths in the design of gradient coils for a conventional MRI
configuration.
Methods
An
MRI system comprises three sets of spatial encoding gradient coils to produce a
magnetic field along x, y and z directions [1].
Due to the switching of gradient coils, eddy currents are induced in the
surrounding coil tracks especially when wide tracks are used for the purpose of
heating reduction [2].
Here, we will study the eddy currents induced in the surrounding coils with
respect to different coil track widths in a conventional MRI system.
A
series of gradient coils with different track widths were designed [3].
The maximum track widths of transverse coils
varied from 10 to 40 mm with a step of 2 mm and the z-coil had a fixed track width of 6 mm. All the coils were designed
with a cold shield made of aluminum to produce a maximum gradient strength G0
= 30 mT/m in a 50×50×40 cm region of interest (ROI). The maximum field error
was ±5% for all the coils. A minimum gap between tracks of 1mm was used. Table
1 lists the properties of the coils and cryostat, and Fig.1a presents the
configuration of the gradient coils. The electromagnetic analysis was based on a
multilayer integral method [4].
A set of simulations were defined by changing the track width of transverse
coils (Fig.2b shows the mesh of a passive y-coil).
In this work, the x-coil was driven
by a current at a frequency 1 kHz, the y
and z coils were un-energized (passive
coils). Performance figures were compared between the active coil (energized
coil) with surrounding coils and the isolated coil. It was assumed that an
isolated coil was not surrounded by the neighboring coils. This allows us to
study the impacts the surrounding coils have on the active coil’s performance.
Results
Fig.2a
and b describe the induced current density distribution in the passive y-coil and cold shield. The transverse
coils have a maximum track width of 40 mm. The current density
distribution in the passive y-coil is
much higher than that of the cold shield, and mainly accumulates in the right
middle of the passive y-coil. Fig.2c
compares the real part of the magnetic field along the x-axis across the ROI induced by the eddy currents in the passive
coils and cold shield. We can see that the field produced by the passive coils
is higher and more nonlinear than that of cold shield.
Fig.3
presents a significant difference of the coil performance between the isolated x-coil and the active x-coil with surrounding y and z-coils. The difference increases rapidly with the track width in a
range of 10 to 20 mm and keeps relatively stable for track widths from 20 to 40
mm. The total power loss dissipated by x,
y and z coils reaches a local minimum value of 20.9 kW at track widths of
14 and 16 mm. The resistance reaches a minimum at a track width of 14 mm. The
inductance and coil efficiency decrease with the track width. The figure of
merit (FoM, defined as η2/L, where η is coil efficiency and L
is inductance) and η2/R have peak values at the track width of
14 mm. η2/R of the coil with a track width of 14
mm is 10.1% larger than the case of 10 mm and 4% larger than the case of 20 mm.
Discussion
and Conclusion
The magnetic field shown in Fig.2c demonstrates that the
surrounding passive coils dissipate more eddy currents than those of the cold wall.
The eddy currents mainly concentrate in the part overlapping with the “eye” of the
active x-coil because of the higher local
current density in this region [5]. Despite the fact inductance has been slightly reduced, the
eddy currents in the passive coils leads to more power loss, lower efficiency,
lower FoM and η
2/
R as well as a larger secondary field, effects
which should be mitigated as much as possible. From this study, we found that the
total power loss, coil resistance, FoM and η
2/
R reach a local
minimum value at a track width of 14 mm, meanwhile the coil with a track width of
14 mm has a relative high efficiency. Therefore, the track widths around 14 mm
are recommended to use in the design of a conventional transverse coil. It is
hoped that this intra-gradient coil interactions study will be useful for future
gradient coil design and analysis.
Acknowledgements
No acknowledgement found.References
[1] J. Jin, Electromagnetic analysis and design in magnetic resonance imaging
vol. 1: CRC press, 1998.
[2] M.
A. Richard, N. J. Mastandrea Jr, and D. A. Lampman, "Gradient coils with
reduced eddy currents," ed: Google Patents, 1998.
[3] H.
S. Lopez, F. Liu, M. Poole, and S. Crozier, "Equivalent magnetization
current method applied to the design of gradient coils for magnetic resonance
imaging," Magnetics, IEEE
Transactions on, vol. 45, pp. 767-775, 2009.
[4] H.
Sanchez Lopez, F. Freschi, A. Trakic, E. Smith, J. Herbert, M. Fuentes, et al., "Multilayer integral
method for simulation of eddy currents in thin volumes of arbitrary geometry
produced by MRI gradient coils," Magnetic
Resonance in Medicine, vol. 71, pp. 1912-1922, 2014.
[5] F. Tang, H. S. Lopez, F. Freschi, E.
Smith, Y. Li, M. Fuentes, et al.,
"Skin and proximity effects in the conductors of split gradient coils for
a hybrid Linac-MRI scanner," Journal
of Magnetic Resonance, vol. 242, pp. 86-94, 2014.