Lenz lenses: RF field amplification and increased sensitivity using passive inductive elements
Peter T. While1,2, Nils Spengler3,4, and Jan G. Korvink3

1Department of Radiology and Nuclear Medicine, St. Olav's University Hospital, Trondheim, Norway, 2IMTEK - Laboratory for Simulation, University of Freiburg, Freiburg, Germany, 3Institute of Microstructure Technology, Karlsruhe Institute of Technology, Karlsruhe, Germany, 4IMTEK - Laboratory for Microactuators, University of Freiburg, Freiburg, Germany

Synopsis

Lenz lenses are discrete elements that focus magnetic flux by virtue of their configuration. By the principle of reciprocity, Lenz lenses may be used to increase both B1-field strength and receiver sensitivity in an NMR experiment. We present the theory that describes Lenz lenses and their geometrical optimization, plus some preliminary experimental results that demonstrate a three-fold increase in local SNR.

Purpose

Signal-to-noise ratio (SNR) in an NMR experiment is linearly proportional to the sensitivity of the RF receiver1, defined as the B1-field strength per unit current. The SNR may be increased by reducing the physical size of the coil and increasing the filling factor. However, in many instances, the minimum achievable size of the coil is restricted by technical limitations, or it is not possible to place the active electrical element in close proximity to the sample region of interest (ROI). The purpose of Lenz lenses is to introduce a passive element into the RF transmit/receive chain that remotely focusses the magnetic flux of the transmitter, which, by the principle of reciprocity, also increases the sensitivity of the receiver2.

Theory

The basic design of a Lenz lens involves two concentric circular conductors of different sizes that are connected in series to form one or more closed loops comprising an annulus3, as depicted in Fig. 1. In the presence of an external field, a current will be induced in the Lenz lens according to Faraday’s law of induction and the current sense will be such that it inductively opposes the incident field, as dictated by Lenz’s law. Since the current sense of the inner loop is opposite to that of the outer loop, the total flux within the inner loop will actually be enhanced. Equivalently, the sensitivity of an NMR system (in receive mode) to fields emanating from within the inner loop will be increased2.

The current induced in the Lenz lens, IL, can be modelled analytically by equating the electromotive force induced by the external field to the electric potential across the equivalent RL-circuit of the lens. For example, for a single lens placed concentrically at the midpoint of a Helmholtz pair (Fig. 1), we obtain:

$$I_{L}=\frac{2i\omega(M_{\mathrm{BH}}-M_{\mathrm{SH}})I_{\mathrm{H}}}{[R_{\mathrm{L}}-i\omega(L_{\mathrm{S}}+L_{\mathrm{B}}-2M_{\mathrm{SB}})]}\qquad\qquad(1)$$

where ω is the frequency, IH is the current of the Helmholtz pair, RL is the resistance of the Lenz lens, LS and LB are the self-inductances of the inner (S – small) and outer (B – big) loops, and MBH, MSH and MSB are the mutual inductances between different combinations of the various loops (denoted by the subscripts). Note that for high-frequency applications it is important to include the skin-effect in the resistance and self-inductance calculations.

It is relatively straightforward to generalize Eq. (1) to the case of multiple Lenz lenses placed arbitrarily within any external field by introducing additional mutual inductance terms. This can be represented by a matrix equation $$$\mathcal{L}\boldsymbol{I}_{\mathrm{L}}=\boldsymbol{H}$$$, with elements given by:

$$\mathcal{L}_{kk}=R_{\mathrm{L}k}-i\omega(L_{\mathrm{S}k}+L_{\mathrm{B}k}-2M_{\mathrm{S}k\mathrm{B}k})\qquad\qquad$$

$$\mathcal{L}_{kj}=-i\omega(M_{\mathrm{B}k\mathrm{B}j}-M_{\mathrm{S}k\mathrm{B}j}-M_{\mathrm{B}k\mathrm{S}j}+M_{\mathrm{S}k\mathrm{S}j})\qquad\qquad(2)$$

$$\boldsymbol{H}_{k}=i\omega(M_{\mathrm{B}k\mathrm{H1}}+M_{\mathrm{B}k\mathrm{H2}}-M_{\mathrm{S}k\mathrm{H1}}-M_{\mathrm{S}k\mathrm{H2}})I_{\mathrm{H}}\qquad\qquad$$

where k = 1:K, j = 1:K, jk, and K is the number of lenses. Note, for example, that MBkSj is the mutual inductance between the outer loop of the kth lens and the inner loop of the jth lens (H1 and H2 refer to individual Helmholtz loops for demonstrative purposes only).

Method and Results

Eqs. (1)-(2) were used to optimize the geometry and placement of Lenz lenses for a simple theoretical NMR microscopy example, with the goal of maximizing sensitivity while maintaining a defined level of field homogeneity (i.e. flip angle) within a cylindrical ROI (using the Matlab® function fmincon). Fig. 2 demonstrates that gains of up to 1.5, 2.2 and 3.0 are possible using one, two and four lenses, respectively, over the use of a Helmholtz pair alone, while maintaining a 1% average field error within the ROI. These gains can be increased further by relaxing the homogeneity constraint.

To demonstrate the effect experimentally, a Lenz lens (not optimized) was patterned by means of gold electro-deposition onto a water-filled, rectangular-shaped capillary, which was embedded within a 10 mm NMR tube containing deuterated agarose gel (Fig. 3a). The tube was aligned carefully within a saddle-shaped coil of a 500 MHz spectrometer (Bruker BioSpin), and images were acquired using a spin echo sequence. Fig. 3 displays an approximately three-fold increase in sensitivity to the water nearby the inner loop of the Lenz lens compared to the water elsewhere in the capillary.

Conclusion

Lenz lenses are simple broadband discrete elements that permit the remote localization of the sensitive volume of an NMR detector, and afford substantial increases in field efficiency.

Acknowledgements

No acknowledgement found.

References

1. D.I. Hoult and R.E. Richards, 1976. "The signal-to-noise ratio of the nuclear magnetic resonance experiment". J. Magn. Reson. 24(1):71-85.

2. J. Funk, N. Spengler, P.T. While, J.G. Korvink, 2015 (patent pending). "RF resonator with a Lenz lens". EP 15174571.

3. J. Schoenmaker, K.R. Pirota, J.C. Teixeira, 2013. "Magnetic flux amplification by Lenz lenses". Rev. Sci. Instrum. 84:085120.

Figures

Fig. 1: Model diagram of a single Lenz lens placed symmetrically within a Helmholtz pair.

Fig. 2: Colour contour plots of the Bz-field induced by a Helmholtz pair and (a) zero, (b) one, (c) two or (d) four optimized Lenz lenses (1% average field error within the ROI – blue dashed rectangle).

Fig. 3: (a) Photograph of the Lenz lens, capillary and NMR tube used in the experimental demonstration; (b-c) MRI images at two magnifications that display the corresponding localized increase in field sensitivity.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3565