Patrick Bluem1, Andrew Kiruluta2, Pierre-Francois Van de Moortele3, Gregor Adriany3, and Zoya Popovic1
1Department of Electrical, Computer, and Energy Engineering, University of Colorado at Boulder, Boulder, CO, United States, 2Massachusetts General Hospital, Harvard Medical School, Boston, MA, United States, 3University of Minnesota, Center for Magnetic Resonance Research, Minneapolis, MN, United States
Synopsis
Traditional
MRI reactive near-field probe design for B1 field uniformity assumes
quasi-static fields. However, for B0>4T, the quasi-static
approximation is no longer valid since the wavelength is smaller than the FOV
and field wave modes appear, affecting image quality. This work presents the use of a copper strip
waveguide structure combined with a traveling wave excitation at 7T, 10.5T human
wide-bore and 16.4T small animal scanners, while observing the effect on a
cylindrical distilled water phantom. A
simple flexible copper strip wearable wrap is shown to improve SNR and field
distribution in UHF-MRI.Purpose
To
improve field coupling and control using a hard surface waveguide that increases the SNR and B
1 field
homogeneity in a homogeneous phantom imaged in UHF-MR systems (7T and above).
Methods
Traditional
MRI reactive near-field probe design for B
1 field uniformity assumes
quasi-static fields. However, for B
0>4T,
the quasi-static approximation is no longer valid since the wavelength is smaller
than the FOV .
1 The
increase in frequency results in the formation of field modes with high field
non-uniformity, undesirable in MRI. One
solution to the problem is an electrically “hard surface” (EHS) flexible cylindrical
waveguide wrapped around the imaging volume. Such an EHS is formed of a
longitudinally corrugated surface, where the tangential electric and
longitudinal magnetic fields are equal to zero along the surface.
2 Fundamentally,
the EHS modifies the boundary conditions of the imaging
volume, thus changing the field profile. A simple EHS is implemented with
copper tape on a thin flexible plastic sheet for use at 7T, 10.5T human wide-bore
and 16.4T small animal scanners.
At 7T and 10.5T, 3cm-wide copper strip spacing is 3 cm and the entire
structure is wrapped around a distilled water cylindrical phantom (L=37.5cm,
D=16cm). At 16.4T, 3cm-wide copper strips
are spaced 1cm apart, due to both a smaller wavelength and phantom diameter (L=34 cm, D=9cm).
Results
A
series of experiments were performed with right-handed circular polarized patch
travelling wave (TW) probe excitations (Fig. 1) for all three systems (Fig. 2a). The circular patches, scaled to the
respective Larmor frequencies, were placed 15 cm from the phantom for the 7T
and 10.5T experiments. This was experimentally found to be the optimal probe distance
which did not produce an overpowering near-field contribution that degrades the
circular polarization, thus losing power to the cross-polarized field modes. The reference voltage of the excitation is kept constant across experiments on the same scanner. The resulting image without the EHS (Fig. 3a) has
relatively low SNR and poor field uniformity with only one probe
excitation. For the first EHS experiment, a 17-cm long EHS was placed
around the air-gap between the probe and the phantom in the 7T bore, placed 1cm away from the probe with 3cm
of the EHS wrapped around the phantom, greatly increasing the field
coupled to the proximal side, but with the field attenuating rapidly along the
phantom length (Fig. 3b). Next, only the full 37.5-cm length of the phantom is
covered with the EHS, creating two distinct high field regions, with greatly
reduced field in the center (Fig. 3c).
The higher field strength farthest from the probe is due to an end
reflection and would not be present in a material with higher loss or a large
imaging volume. However, the relative magnitude of the field is much lower than
in the previous case. The final 7-T experiment has both the air gap and the
phantom surrounded by one complete EHS wrap, resulting in a higher SNR and overall better
quality image (Fig. 3d). The overall benefit from the EHS observed in
simulations is the 5 times larger B
1 strength relative to the case
with no EHS. The overall change is also apparent
at 10.5T and 16.4T. At 10.5T, the probe
was placed 15cm away from the phantom and both the air gap and the complete
phantom were covered in the EHS, resulting in a drastic increase in the
relative field magnitude (Fig.4), with a focused field distribution at the end
closest to the exciter probe, while the center of the phantom is almost
invisible. At 16.4T, the setup is fairly different (Fig. 2b) with a large
discontinuity between the bore and the gradient coil region. For this system, the EHS not only increased
the SNR seven times (Fig. 5) compared to an image with no guiding structure, but
also yielded transmit B
1 magnitude large enough to apply a refocusing RF pulse and collect spin echo images (not shown)
with the patch probe excitation scaled to 698MHz.
3Conclusions
A
simple flexible wearable copper strip electrically-hard surface (EHS) structure
is shown to improve SNR and field distribution in UHF-MRI. With more complicated periodic flexible structures,
boundary conditions of the electromagnetic field on the phantom can be tailored
for a desired field distribution. This flexible EHS wrap can also be used with
classical MRI volume coils, as the boundary conditions on the phantom do not
depend on the excitation.
Acknowledgements
The authors acknowledge support by the National Science
Foundation under a collaborative research grant ECCS 1307614.References
1.
D. O. Brunner, et al. “Travelling-wave
nuclear magnetic resonance,” Nature
457(7232). 994-U2. 2009.
2.
G.
Ruvio, et al. "Modal propagation
in ideal soft and hard waveguides," IEEE APS Intern. Symp Dig.,
vol.4, no., pp.438-441 vol.4, 22-27 June 2003.
3.
P.
Bluem, et al. "Travelling-wave
excitation for 16.4T small-bore MRI," IEEE MTT-S Intern. Symp. Dig.
pp.1-4, 17-22 May 2015