Maximilian N. Voelker1, Oliver Kraff2, Daniel Brenner3, Astrid Wollrab4, Oliver Weinberger5, Moritz C. Berger6, Simon Robinson7, Wolfgang Bogner7, Christopher Wiggins8, Robert Trampel9, Tony Stöcker3, Thoralf Niendorf5,10, Harald H. Quick2,11, David G. Norris2,12, Mark E. Ladd2,6, and Oliver Speck4,13
1Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Hospital Essen, University of Duisburg-Essen, Essen, Germany, 2Erwin L. Hahn Institute for Magnetic Resonance Imaging, University of Duisburg-Essen, Essen, Germany, 3German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany, 4Otto-von-Guericke-University, Magdeburg, Germany, 5Berlin Ultrahigh Field Facility (B.U.F.F.), Max-Delbrueck-Center for MolecularMedicine, Berlin-Buch, Germany, 6Medical Physics in Radiology, German Cancer Research Center (dkfz), Heidelberg, Germany, 7High Field MR Center, Department of Biomedical Imaging and Image-guided Therapy, Medial University of Vienna, Vienna, Austria, 8ScanNexus, Maastricht, Netherlands, 9Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany, 10Experimental and Clinical Research Center, a jointcooperation between the Charité Medical Faculty and the Max Delbrück Center for MolecularMedicine, Berlin, Germany, 11High Field and Hybrid MR Imaging, University Hospital Essen, University Duisburg-Essen, Essen, Germany, 12Donders Centre for Cognitive Neuroimaging, Nijmegen, Netherlands, 13Leibniz Institute for Neurobiology, Magdeburg, Germany
Synopsis
The “traveling heads” is an experiment started in 2014 to assess the
comparability and reproducibility of multicenter human brain imaging at 7T. This is of particular interest as 7T MRI is currently
being discussed to become a clinical system in the very near future. The number
of installations continues to increase, with currently approximately 60
research sites in operation worldwide. As an advantage, this new technology
provides higher SNR, yet the artifact-to-noise ratio
is also increased. This can influence the image quality severely and may be
different at individual UHF sites, where system hardware differences could
diminish reproducibility.Purpose
In this study we compared the image quality of state-of-the-art 7T brain
imaging sequences performed in the same two “traveling heads”
1 between eight
different 7T MRI sites.
Methods
Two male subjects (37 and 33 yrs.) were imaged at eight UHF sites all operating
a 7T whole-body MRI system from the same vendor (Siemens). The systems have
differences in basic imaging components, which might influence image quality.
Two sites (2, 6) have an actively shielded magnet installed, whereas the rest
of the sites use an older, passively shielded magnet. There are also two
different gradient coil versions installed at the different sites. The
high-performance version (70 mT/m, other version: 38 mT/m) comes with a reduced
gradient linearity and is installed at five sites (2, 5-8). At 6/8 sites a commercial
RF head coil (Nova Medical) with 1 TX and 32 RX channels was used, while at two
sites (1, 4) a similar 24 RX channel version from the same manufacturer was
available. To facilitate data analysis, the measurement setup was kept as
similar as possible between the sites and the subjects were positioned as standardized
as possible with the help of locally available cushions and pillows, as well as
auto-align localizer scans.
The imaging protocol consisted of a spin-echo based B1 mapping sequence
(TA: 0:27, 8x8x5mm³) for transmitter calibration verified by a DREAM2 B1-mapping sequence (TA: 0:05, 5x5x5mm³).
Subsequently, the following imaging sequences were performed: MP2RAGE (TA: 9:38min,
0.75×0.75×0.75mm³), TSE (TA: 7:05min, 0.3×0.3×2.0mm³), modified TOF3 (TA: 6:41, 0.2×0.2×0.4mm³), and SWI (TA: 9:05,
0.3×0.3×1.0mm³). An EPI sequence with sinusoidal readout (TA: 4:22,
1.3×1.3×1.5mm³) was used to acquire functional data during 4 minutes scan time
with no task.
All measurements were acquired within 1 hour per subject. Data were analyzed
qualitatively and quantitatively after co-registration between systems.
Registration and image processing was done with FSL (FMRIB Software Library v5.0).
MP2RAGE images were registered to the MNI template. The TSE data were
registered to MP2RAGE to use VOI masks gained from previous analysis of
MP2RAGE. SWI and TOF measurements were co-registered to a second reference scan
of each subject taken at Site 3. ROIs
were defined in multiple brain structures, and image contrasts were compared
between all sites. For anatomic sequences (MP2RAGE and TSE), the contrast, T1, and
volume of gray and white matter, CSF and subcortical nuclei were compared. Images
from TSE were analyzed in two subgroups for each RF coil, as sequence
parameters had to be changed for the 24ch coil. For the angiographic sequences
(TOF and SWI), vessel contrast against surrounding background was used for
quantification. Temporal SNR (tSNR) maps calculated from EPI as well as B1 and B0
field maps were analyzed for each brain. Additionally, both subjects were
rescanned at Sites 2, 3 and 4 to assess intra-system variability.
Results
MP2RAGE images showed very high agreement in contrast and measured T1 values
(Fig.1, Table 1). Maximum variation of calculated brain size between the sites
was 6% for Subject 1 and 3% for Subject 2. Rescans at 3 sites showed higher
precision (1% difference) for same-system data, indicating systematic
differences between the sites. T1 values were within 3% variation, except for
inferior regions like the brainstem, where the signal level typically drops off
with the RF coils used. On TSE images (Fig. 2) one can clearly distinguish the
different RF coils. The gray to white matter contrast differences between sites
were below 8%. However, in central subcortical regions larger contrast
deviations (5–20%) were found between the sites with different RF coils (Table
1). The TOF angiography (Fig.3) analysis showed high agreement except for 5/20
vessels, where measured contrast was more than 10% lower for the 24ch coil. SWI
image data (Fig.4) had higher vessel to background contrast deviations when
compared to a reference scan, inter-site comparison showed a difference between
5% and 18% depending on vessel position and size.
Discussion and
Conclusion
UHF MR systems are very sensitive measurement devices
4. Even when provided
from the same vendor, there are differences between individual devices and
sites that potentially might influence image quality. Effects of hardware
differences such as RF coils were revealed, but we did not control for physiological
differences in the individual subjects such as heart rate or breathing
frequency that might also influence image quality. Nevertheless, our results
show, that inter-site comparability of UHF measurements can be achievable in
the near future. This is a promising finding for the comparability of studies,
such as fMRI or relaxometric quantification of disease markers, and does ensure
the same diagnostic reliability on all MRI devices.
Acknowledgements
This
work was support by a grant of the German Research Foundation (DFG) / project
German Ultrahigh Field Imaging / Grant n. LA
1325/5-1.
UHF-adapted
imaging sequences were provided by
Siemens HealthcareReferences
1
Voelker et al., Proc. Intl. Soc. Mag. Reson. Med. 22, 3202 (2014)
2
Brenner et al., Proc. Intl. Soc. Mag. Reson. Med. 22,
1455 (2014)
3 Johst et al., Invest Radiol. 47(8):445-50 (2012).
4
Bernstein et al., JMRI 24:735–746 (2006)