Synopsis
We propose a flexible coil array design
consisting of circumferential shielding stacked on circular coils to achieve
robust decoupling between coil when the array is either bended or on a
flat plane. Two types of circumferential
shielding were tested through numerical simulation. The results demonstrated
that both types can have good decoupling on a curved surface with S21
< -17.1 dB. The maximum |B1| for the array coil with Type-II (two-parallel-ring)
shielding was 63% larger than that with Type-I (single-strip) shielding. Future
work will empirically construct these coils to compensate simulation inaccuracies
and validate their performance In Vivo.PURPOSE
MRI phased-array coil can achieve large FOV and high SNR simultaneously
if coils in an array can be sufficiently decoupled. Coil decoupling can be achieved by either overlapping coil elements
1 or adding inductors and capacitors between
coils
2. Note that the anatomical
structure can vary significantly among clinical applications, a flexible coil
array allowing for a higher filling factor for different subjects is desired
3-4. However, decoupling by overlapping coil
restricts the mechanical flexibility of the coil array. Coil decoupling by adding
inductors and capacitors has the disadvantage of ohmic contact loss.
Here we propose
to decouple between coils in an phased-array by using circumferential shielding
stacked on top of the coil (
Fig.1). We hypothesize
that such a stacked structure is particularly useful when the coil array may be
mechanically bended, because clear gaps between coils allows bending. Also the
stacked structure permits exchanged for extendable anatomical range. Here we studied
two types of circumferential shielding (
Fig.2)
and compared their performance using full-wave electromagnetic simulation. Simulation
inaccuracies and variations of components can be easily compensated by further
fine-tuning the practical coils in the future. Our results demonstrate that
both types achieved good isolation between coils (S
21 lower than
-17.1dB) and provided localized sensitivity B
1 regardless of the array
was flat or bended.
METHODS
The
simulation setup was illustrated in
Fig.25.
Two decoupled PEC loops were located above the center of the RF coils at right in
order to evaluate isolation between coils using S
12. A liquid-filled
cylindrical phantom was included to simulate the loading, which reduced the quality
factor (Q) such speeded up to meet convergence criteria. We simulated two side-by-side
circular coils (diameter = 50 mm, width = 5 mm, thickness = 0.02mm) built by copper
sheet adhering on FR4 substrate with a 2 mm gap between coils. Each circular coil
used three evenly distributed tuning capacitors. The capacitor value was tuned
to drive resonance frequency to approach the
1H Larmor frequency (125
MHz) at 3T. The electromagnetic simulation was done using HFSS (Ansys, Canonsburg,
USA). All analysis parameters used the following parameters: max ΔS = 0.02, minimal coverage pass = 2, mixed-order
basis functions, sweep-type = fast. Two types of circumferential shielding were
modeled: type-I (
Fig.2C) was a single strip
with width = 8 mm
6; Type-II (
Fig.2D) consisted of two parallel
rings (width = 1mm, gap = 8mm). The sizes of both
shielding were adjusted to achieve the best decoupling. To demonstrate the
flexibility as changing the relative position of two coils on decoupling, each
coil array was either placed on the flat top of the phantom or bended to fit to
the curvature of the phantom. Maps of |B
1| were evaluated by formulas
described in Ref.
7.
RESULTS
Fig.3 shows |B
1| maps when
coils were placed on the flat top of the phantom without and with decoupling shielding,
respectively. Coupled B
1 field was apparent beneath the RF coils at
left. The coupled |B
1| reduced to 3.1% and 6.8% with type-I and type-II shielding, respectively. The max |B
1|
beneath the RF coil at right and S
21 with type-I and type-II
shielding were {6.31x10
-6(a.u.), -20.4(dB)}
and {8.41x10
-6(a.u.), -19.6(dB)}, respectively.
Fig.4 shows |B
1| maps when coils were bended
to fit to the curvature of a 200mm-diameter cylinder. The coupled B
1
field without shielding was about 85% of flat-top results in Fig.3. With type-I and type-II shielding,
the coupled |B
1| reduced to 2.5% and 4.9% and the resonant frequency drifted ≤ 0.75 MHz. The max |B
1| beneath the RF coil at right and S
21 were {3.74x10
-6(a.u.), -18.2(dB)} and {5.66x10
-6(a.u.), -17.2(dB)}, respectively.
Figs.5
shows the |B
1| maps on the curvatures of 180mm and 220mm-diameter cylinders.
Since bending an array inevitably changed the relative positions of two coils,
the mutual inductance was no longer the same and the coupling between coils was
significantly different after such a geometrical deformation. However, both types
of circumferential shielding had similar B
1 distribution and penetration-depth, successfully maintained the image quality in these scenarios.
DISCUSSION
We used numerical simulation to demonstrate that a
flexible coil array consisting of circumferential shielding stacked on circular
RF coils can be used to acquire high quality MRI even when the array was
mechanically bended. While we only demonstrated an array of two coils, we
expect that more channels with circumferential shielding can be integrated into
an array in order to increase FOV while maintaining the same decoupling between
coils. Also our simulation results suggest that type-II (two-parallel-ring)
structure performs better than type-I (single-strip). Future work will empirically
construct these coils and validate their performance In Vivo.
Acknowledgements
This study was supported by Ministry of Science and Technology, Taiwan (MOST 104-2314-B-002-238, MOST 103-2628-B-002-002-MY3), National Health Research Institute, Taiwan (NHRI-EX104-10247EI), and Ministry of Economic Affairs, Taiwan (100-EC-17-A-19-S1-175).References
1. Romer PB, et al. Magn Reson
Med. 1990;16:192-225.
2. Lee RF, et al. Magn Reson Med. 2002;48:203-213.
3. Keil B, et al. Magn Reson Med. 2011;66:1777-1787.
4. Duan Y, et al. Proc ISMRM
2013. #4383.
5. Martius S, et al. Proc ISMRM 2015. #3099.
6. Gilbert
KM, et al. Magn Reson Med. 2011;67:1487-1496.
7. Collins CM, et al. Magn Reson
Med. 2001;45:684-691.