Stefan Glöggler1, Silvia Rizzitelli2, Noel Pinaud3, Gérard Raffard2, Véronique Bouchaud2, Stéphane Sanchez2, Alan Wong4, and Yannick Crémillieux2
1University of Southampton, Southampton, United Kingdom, 2CRMSB, Université Bordeaux, Bordeaux, France, 3Université Bordeaux, Bordeaux, France, 4CEA Saclay, Gif-sur-Yvette, France
Synopsis
We present two highly sensitive microcoil probes (1H and 13C) that are connected to microdialysis probes inside a MRI scanner. Under continuous flow conditions the metabolic response can be monitored in real-time in vitro and in vivo.Purpose
For the analysis of
metabolites it can be challenging to obtain enough sample volume for the
detection in regular high field NMR spectrometers. Microcoil detection can be
used to obtain a higher mass sensitivity than with regular high-field
spectrometers and to measure small quantities of metabolites with high
sensitivity.
1 However, handling of small sample volumes (≤ 1 µL) is difficult and typically, samples containing
metabolites of interest need to be collected and can only be analyzed at a
later point of time.
2 In order to monitor metabolic responses in
real-time inside an imaging scanner we have developed a method that combines a
sensitive microcoil for detection with a microdialysis for sample extraction.
In this way it is possible to measure metabolic responses in vitro and in vivo under
continuous flow conditions in real-time without the need of a dedicated
spectroscopy magnet.
Methods
Two microcoil probes were built, one for
13C (at B0
= 4.7 T, which corresponds to 50 MHz carbon frequency) and the second one for
1H
detection (at B0 = 7 T, which corresponds to 300 MHz proton
frequency). An example probe is depicted in figure 1. In order to optimize for
the signal-to-noise ratio, the NMR-coils were designed to have a filling factor
of 93%. This was achieved by winding double parallel coils (30 µm or 62.5 µm
diameter of the wire) with a length of 2.4 mm around a susceptibility matched
polyamide tube (O.D. 780 µm, ID 750 µm). In addition, the coils were placed
inside a tube containing fluorinated liquid which is matched to the
susceptibility of the copper wire.1 Both detectors were designed as
double resonant probes. For the
13C resonator the same NMR coil is
used for
13C and
1H detection and excitation. In case of
the 1H resonator two independent circuits are utilized one that
contains a microcoil for 1H excitation and detection and the second
one containing a surface coil matched to the
13C frequency in the
respective magnetic field. The inlet of the microcoil is connected to the
outlet of a microdialysis probe, which sits inside a buffer solution containing
U87 human glioma cells or inside the cortex of a rat brain. The inlet of the
microdialysis is connected to a syringe filled with additional buffer solution which
is pumped under continuous flow through the microdialysis probe to collect
extracellular metabolites. The cell suspension was infused with
13C
labeled 3-
13C-pyruvate (150 mM) and the metabolites measured under
continuous flow.
1H and
13C experiments were performed by
averaging over 3.5 minutes and 15 minutes respectively on a Bruker Biospec 7 T
or 4.7 T.
Results and Discussion
For the carbon probe, the
proton channel is only used to shim the field within the sample volume and
serves for the decoupling of protons. With the proton probe, shimming is
performed directly via the proton channel. The carbon channel is however used
to distinguish between resonances that originate from infused
13C
labeled material or a different source. The sensitivity of the two probes was
firstly tested in vitro. For
13C labeled lactate a limit of
detection (LOD) of 10mM (10 nmol) was achieved within 100 s with the 13C
probe. The 1H limit of detection of the methyl group (thee protons)
of unlabeled lactate amounts to 1.8 mM (1.8 nmol) in 60 seconds for the 1H
probe.
In vitro experiments under
continuous flow conditions within the timescale of the experiments (3.5 minutes
for protons) resulted in a proton LOD of 0.8 mM (800 pmol) of metabolites originating
from 3-
13C pyruvate. Carbon experiments yielded a LOD of 3.3 mM (3.3
nmol) on a 15 minutes timescale under continuous flow. Figure 2 summarizes the
results for both experiments. The conversion of pyruvate into alanine,
glutamine and lactate is observed under continuous flow in both cases and was confirmed
by high resolution NMR at 500 MHz. Furthermore,
the performance of the microcoil was tested in vivo on a healthy rat to detect
normal brain metabolism, using the same experimental setup. A preliminary essay
demonstrated the presence of 8 mM endogenous lactate in the dialysate.
Conclusion
We have demonstrated a new
approach that combines the use of microcoil detection and a microdialysis to
measure extracellular metabolites inside a MRI scanner in real-time. We expect
the applications of this approach numerous since it may be readily applied to
in vivo human measurements and opens pathways to measure immediate metabolic
responses.
Acknowledgements
The authors acknowledge funding from Labex TRAIL (Oncoflux) and ANR Gd_Lung.References
1.
Olson DL, Peck TL, Webb AG, Magin RL, Sweedler JV,
High-resolution microcoil 1H-NMR for mass-limited, nanoliter-volume samples.
Science 1995; 22:1967-1970.
2.
Nandi P, Lunte SM, Recent trends in microdialysis sampling
integrated with conventional and microanalytical systems for monitoring
biological events: A review. Anal. Chim. Acta. 2009; 651:1-14.