Graham Wiggins1,2 and August Frank3
1The Center for Advanced Imaging Innovation and Research (CAI2R), Department of Radiology, New York University School of Medicine, New York, NY, United States, 2The Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, NY, United States, 3University of Pennsylvania, Philadelphia, PA, United States
Synopsis
The sensitivity of surface coil arrays can be
maximized by conforming them closely to the human body, maximizing coil loading
and minimizing the contribution of coil noise. There
is, however, great diversity in body shape and size. Rigid coils must adopt a
compromise, allowing for the largest cases while not giving up too much
sensitivity when imaging smaller examples. Flexible arrays can be wrapped
tightly around the body, but if the overall array dimensions are fixed there
will always be gaps or undesired overlaps depending on body habitus. We present here a geometrical approach based on
a trellis-like lattice of interlinked slats which allows a surface coil array
to be re-shaped while maintaining coil tuning, coil loading and decoupling of
neighboring coils.Purpose
The sensitivity of surface coil arrays can be
maximized by conforming them closely to the human body, maximizing coil loading. There is, however, great
diversity in body shape and size. Rigid coils must adopt a compromise, allowing
for the largest cases while not giving up too much sensitivity when imaging
smaller examples. Flexible arrays can be wrapped tightly around the body, but
if array dimensions are fixed there will always be gaps or
undesired overlaps depending on body habitus.
A
surface coil array which could stretch would allow optimal conformation to the
human body. Other authors have demonstrated a flexible knee coil in
which the surface coil conductors were constructed of braided copper and
changed size as the array was elongated [1]. This caused shifts in resonant frequency and loading such that it was not possible to
present the optimal noise-matched impedance to the preamp in all
configurations. We present here a geometrical approach based on a trellis-like
lattice of interlinked slats which allows a surface coil array to be re-shaped
while maintaining coil tuning, coil loading and decoupling of neighboring
coils.
Methods
The Trellis Coil design principle is shown in
Fig. 1. Two sets of parallel lines at different angles to each other are
overlaid. The lines are linked at each crossing point and the distance d
between linking points along each line is a constant regardless of the angle
between the lines. Paths for surface coil conductors are defined which pass
through these crossing points. To create a 10% overlap between neighboring
coils to minimize inductive coupling, it is necessary to have 10 crossing
points per coil, determining the density of the underlying lattice. As the
angle between the two sets of parallel lines in the underlying lattice is
changed the shape of the surface coil changes while always maintaining a 10%
overlap between neighboring coils. The coil perimeter and surface area enclosed
by each coil change only slightly over this range.
As a practical example, consider a cylindrical
coil array for knee imaging with 8 overlapped elements. If the size of the coil
is to be adjustable from a diameter of 15 cm to 22 cm, the coil element must
stretch from 6.5 cm to 9.5 cm in width. To examine the behavior and practical
implementation of this concept, a two-coil example was constructed. The
underlying lattice was designed in Solidworks (Waltham, MA, USA). Rods from one
layer protruded through holes in the other layer of struts and were each capped
to produce hinge points linking the structure together (Figure 2). Holes for mounting pins
and wire guides were incorporated in the design. It was 3D printed in ABS
material on a Stratasys Fortus 360 printer (Eden Prairie, MN, USA). Coil elements were
constructed out of circuit board segments and Teflon-coated multistrand wire
(Fig. 3), with appropriate capacitors to bring the coils to resonance at 123.2
MHz. Circuit boards are secured to the trellis with pins which slide in
slots to maintain the orientation of the boards. The coils were tuned and
matched when placed 1 cm above a body-sized tissue equivalent phantom (εr=40.5, σ=0.58 S/m).
Results
Over the range of extensions studied here the
coil tuning, match and decoupling all remained very stable as the array was
stretched (Fig 3). S
11 and S
21 values were always less than -13
dB. The wires slide easily around the guides and have not succumbed
to metal fatigue after several hundred extension and contraction cycles. The
unloaded to loaded
Q ratio for the surface coils was 242 / 33 = 7.3 in all configurations.
This compares to 337 / 36 = 9.4 for a standard circuit board coil element of
the same size. The coil circumference defined by the trellis does not stay constant as the array is flexed (Figure 4), which causes the wires to fall slack for some extensions. This did not compromise decoupling over this range of extensions.
Discussion
The use of thin wires reduced the unloaded
Q of
the trellis coil elements, but as they are designed to always be in close proximity
a sufficiently high
Q ratio of 7.3 is
achieved. The unloaded
Q may be increased by using thicker multistrand wire,
finding the optimum trade-off between conductivity and flexibility. The concept
can be extended to cylindrical structures [2] (Fig. 5) and multi-row arrays. Possible applications include
knee and extremity coils, body arrays and pediatric coils. It can also be used
to maintain the optimal distribution of electric dipole antennas around the
torso for 7T body arrays [3].
Acknowledgements
This work was supported by the Center for
Advanced Imaging Innovation and Research (www.cai2r.net),
a NIBIB Biomedical Technology Resource Center (NIH P41 EB017183). References
[1] Nordmeyer-Massner
J, Magnetic Resonance in Medicine 67:872–879 (2012)
[2] Wiggins G, Proc. ISMRM 2016 (submitted)
[3] Zhang B, Proc. ISMRM 2016 (submitted)