Marco Piccirelli1, Christopher Bockisch2, Marc Bovet3, and Roger Luechinger4
1Department of Neuroradiology, University Hospital Zurich, Zurich, Switzerland, 2Department of Ophthalmology, University Hospital Zurich, Zurich, Switzerland, 3ICT, University Hospital Zurich, Zurich, Switzerland, 4Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland
Synopsis
To improve neuro-ophthalmologic surgery,
biomechanical inside about the pathophysiological dynamic of eye movements is
needed. High spatiotemporal resolution dynamic MR imaging of the orbit during
eye movement shall provide such information. To enable clinical use, the
acquisition design needs to accommodate limited patient capabilities to perform
repetitive eye movement accurately.PURPOSE
Functional imaging of the orbital tissue might
improve strabismus surgery by providing new pathophysiological inside1
about the interlinked viscoelastic properties of the active and passive orbital
tissues; which are impossible to predict from static images or ophthalmological
tests.
To enable a clinical use of high spatiotemporal
resolution dynamic MR imaging of the orbit during eye movement, the patient
ability to maintain a reproducible movement in regard to the position and speed
timing is crucial.
The relatively long scantime can be decreased
by using higher B0 field strength, multichannel receive array for simultaneous
scanning of both orbits, and kt-SENSE. All three are challenging for the orbit,
due to the neighboring oral cavities, and therefore requiring good shimming.
Further, a minimum water-fat shift is aimed.
Despite
these numerous challenges, we proceeded to several methodological improvements
aiming clinical usability. Simultaneous eye tracking to control the
reproducibility of the eye motion, and may be used for rejection and
re-measurement of k-space profiles with wrong gaze direction.
METHODS
To help the patient to perform repetitive smooth-pursuit
eye movements a gaze target in form of a horizontally sinusoidally moving red target
(red dot, 2s period, peak velocity 64°/s, amplitude ±20°) is projected on a
dark grey background using a laser and galvanically driven moving mirrors with
500Hz position update, which gives the impression of a continuously moving
target, which strongly improved the presentation quality compared to a 50 or
60Hz video projection.
The eye position tracking was recorded with
an infrared camera at 500Hz (eyelink2000 system, SR Research, Mississauga,
Ontario, Canada).
Further, the segmented acquisition was
synchronized with the gaze target motion2.
B0 map based shimming was performed using a
rectangular shim volume including both orbits.
The subject was able to see through a two
channel microscopy coil, 47 mm in diameter (Philips Research; Hamburg, Germany)
one placed on each eye. On a 3T Philips Achieva scanner, kt SENSE with
acceleration factor from 2 to 5 were used to decrease acquisition time. The
flip angle (7.5°) of the segmented 3D T1 TFE sequence was signal optimized.
Other scan parameters are: 14 time frames of 170 ms duration, 38
over-continuous axial slices, TE 4.8 ms, no signal averaging. A field of view
of 120x120x30 mm3 was used to avoid foldover artifacts. Readout direction was
AP and a scan percentage of 80% was used to shorten scantime: scan matrix 300x240x19.
RESULTS
The smooth-pursuit reliability decreased
sensibly with increasing scantime. The eye movements could successfully be
tracked and the inconsistent data can be identified, see Figure1.
The multichannel microscopy coil enables
the simultaneous scanning of both orbit with a good SNR. The phase encoding in
the RL direction was efficient in preventing motion artifacts from spreading
over the orbit.
Saturation bands occurring within the orbit
need to be taken care with a slight f0-shift.
The
muscle path during eye movement can be followed accurately due to the high
contrast with the (fatty) orbital connective tissue.
DISCUSSION
We acquired high spatiotemporal resolution
dynamic MRI of the orbit in a scantime comparable to a single static high
resolution images, a reasonable smooth-pursuit duration for patient
capabilities. Increasing the kt-SENSE acceleration factor does decrease the scantime
and the SNR, but as the accuracy of eye movements is higher during the
acquisition duration, the overall image quality is equivalent.
To
enable a clinical use of dynamic orbital MRI, the patient’s motivation and training
to perform reproducible eye movement needs to be considered. Tracking eye
position allows cycles with bad motion and speed accuracy to be discarded.
CONCLUSION
Increasing the scantime of dynamic orbital MRI
might not improve image quality due patient limited cooperation. Therefore, eye
tracking during MRI acquisition is necessary for improving image quality.
Acknowledgements
No acknowledgement found.References
1.
Piccirelli M et al. Local
deformation of extraocular muscles during eye movement. Invest Ophthalmol Vis
Sci 50 (11), 5189-5196, 2009.
2.
Piccirelli
M et al. Proc. ESMRMB 2005
http://dx.doi.org/10.1594/ESMRMB05/418