Dan Ma1, Shivani Pahwa1, Vikas Gulani1, and Mark Griswold1
1Radiology, Case Western Reserve University, Cleveland, OH, United States
Synopsis
The
goal of this study is to characterize and improve the accuracy and
repeatability of 2D MRF scans in the presence of slice profile imperfections. Slice
profile imperfection causes deviation between the actual flip angles and
nominal flip angles, which affects the accuracy of measured T1 and T2 values.
This error can be corrected by simulating the RF excitation pulse in the
dictionary. No extra scan time or post-processing time is needed once the new
dictionary is simulated. The accuracy of both T1 and T2 is improved after slice
profile correction. MRF also demonstrates good repeatability, with the
coefficient of variance (CV) of 1.17% for T1 and 3.08% for T2.purpose
The
goal of this study is to characterize and improve the accuracy and
repeatability of 2D MRF scans in the presence of slice profile imperfections. [1].
MRF is an acquisition and processing framework that utilizes pseudorandomized
acquisition scheme to enable simultaneous quantification of multiple tissue
parameters, such as T1 and T2. In its
original implementation, the randomized acquisition scheme were simulated into
a dictionary using the nominal scan parameters such as TR and flip angle. The
best fit between acquired signal and dictionary provides the underlying tissue
parameters. However, because flip angle is an input variable of the dictionary
simulation and affects the degree of both T1 and T2 effects on a signal
timecourse, the agreement between nominal flip angle and actual flip angle is
critical for obtaining accurate T1 and T2 results. Both slice profile
imperfections and the local B1+ field may cause flip angles to be
inaccurate. The B1+ effect varies
based on location and can be corrected by pre-measuring the B1 map, for
example, using the Bloch Seigert method[2].The slice profile imperfection is
not location specific in 2D and thus can be simulated in the dictionary without
increasing the dimensionality of the dictionary. This study simulated the slice
profile of different RF pulses and validated the accuracy and repeatability of
the results.
Methods
The
dictionary with slice profile correction required additional simulation of RF
pulse and slice selection gradient. Based on the RF duration, time-bandwidth
product (TBW) and slice thickness, the sinc pulse and the corresponding slice
selection (SS) gradient were simulated. At the beginning of each TR, the RF
pulse and slice selection gradient were first divided into segments with a 10 ms step
size. The RF excitation, dephasing from the SS gradient and off-resonance, as
well as relaxation were simulated continuously from one segment to the next.
After all the segments were simulated, the dephasing and relaxation from the
rest of the TE/TR time were simulated. Finally, the slice profile was simulated
using 50 isochromats across a distance four times wider than the nominal slice
thickness to account for out-slice excitation and then summed together to
become the signal evolution from one pixel. A total of 287709 signal time
courses, each with 3000 time points, with different sets of T1, T2 and
off-resonance parameters were simulated. The total simulation time was 4.7 hours.
Acquisition
10 cylindrical
gel-phantoms with T1 and T2 ranging from 200 to 1700 ms and 20 to 110 ms,
respectively, were scanned in a 3T scanner (Skyra, Siemens). Because RF
duration and TBW have an impact on the MT effect [3] and slice profile, sequences
with four different RF durations and TBWs were applied in separate scans to
estimate the potential MT effects of the gel phantom and slice profile effect
on the resulting T1 and T2 values: sinc pulse with duration of 800 us and TBW
of 2, duration of 2000 us and TBW of 8, duration of 4800 us and TBW of 2, and duration
of 4800 us and TBW of 8. The total acquisition time for these four sequences
were 30, 40, 50 and 50 seconds, respectively. The T1 and T2 results were
compared to the results from gold standard, spin-echo based measurements. The
common acquisition parameters of these four sequences are: FOV of 300x300mm2,
matrix size of 256x256, image resolution of 1.2x1.2 mm2, and slice
thickness of 5 mm. 3000 TRs were acquired from each acquisition. For a repeatability
study, both MRF scan with RF duration of 2000 us and TBW of 8 and standard
spin-echo based measurements were scanned 5 consecutive days, each day with 3
repetitions. The mean and standard deviation of T1 and T2 from both methods
were calculated.
Results
Figure
1 compares the T1 and T2 values from MRF, without and with slice profile
simulation, to those from standard measurements. Both T1 and T2 accuracy is improved,
despite the change in RF durations. Figure 2 shows the results from repeatability
study, with bi-directional error bars representing the standard deviation of
all measurements. MRF results are in agreement with the results from gold standard
measurements, with a concordance correlation coefficient (CCC) of 0.9874 for T1
and 0.9905 for T2. In addition, both MRF and standard measurements demonstrate
high repeatability. The coefficient of variance (CV) is 1.17% for T1 and 3.08%
for T2 from MRF scans.
Discussion
This study demonstrates a slice profile
correction method for 2D MRF scans. No extra scan time or processing time is
needed once the new dictionary is simulated. Both T1 and T2 accuracy are
improved after slice profile simulation. This work also demonstrates high
repeatability of the MRF scans.
Acknowledgements
The
authors would like to acknowledge funding from Siemens Healthcare and NIH
grants NIH 1R01EB016728-01A1 and NIH 5R01EB017219-02References
1. Ma D., Gulani V., Seiberlich N. et al, Magnetic resonance fingerprinting. Nature, 2013, (495) 187-193
2. Chen Y., Jiang, Y., Ma D., et al, Magnetic resonance fingerprinting (MRF) for rapid quantification abdominal imaging. ISMRM 2014, 0561
3. Bieri O., Scheffler K. Optimized Balanced Steady-state Free Precession Magnetization Transfer Imaging. MRM, 2007, 511-518.