Shelley HuaLei Zhang1, Jr-Yuan George Chiou1, Robert Vincent Mulkern1, Pelin Aksit Ciris2, Stephan Maier1, and Lawrence Panych1
1Radiology, Brigham and Women's Hospital, Boston, MA, United States, 2Department of Biomedical Engineering, Akdeniz University, Antalya, Turkey
Synopsis
Magnetic resonance spectroscopic imaging (MRSI) technique
is often plagued by signal contamination (e.g. from fat) from outside the
tissue of interest, particularly near organ boundaries. The goal is to
demonstrate and validate in phantoms and healthy volunteers a spatial
localization technique that features localized signal excitation and therefore eliminates
the need for post-processing filtering. Here we demonstrated superior PSF’s at prostate-relevant
metabolic peaks for a 2D Gaussian-based PSF-Choice encoding scheme. The
reduction of contamination from signals in surrounding tissue was further
demonstrated in an in-vitro phantom. The feasibility of the method was demonstrated
in a small group of prostate MRSI studies with normal volunteers.Purpose
Magnetic resonance spectroscopic imaging (MRSI) provides unique
characterization of metabolites within cancerous or benign tissue.
Unfortunately, the technique is often plagued by signal contamination (e.g.
from fat) from outside the tissue of interest, particularly in voxels near organ
boundaries. Post-processing filters are typically applied, which have the
undesirable side effect of further lowering spatial resolution. [1] The purpose of this
study is to demonstrate and validate in phantoms and healthy volunteers a
spatial localization technique that features spatially localized signal
excitation and therefore eliminates the need for post-processing filtering.
Methods
Pulse Sequence:
The sequence was customized based on 3D-PROSE; the
excitation pulse was replaced by a series of four hard pulses weighted with a Point
Spread Function (PSF)–Choice [2] encoding scheme along two spatial directions. A
512-pulse echo planar train read out with alternating polarity was added along
the third direction during an FID acquisition. Each blip collects 32 points
during the ramp and plateau.
Phantom Validation:
The point-spread-function obtained with this technique
was verified at 3T, on a GE prostate phantom (metabolic solution enclosed in a 10-cm
diameter sphere). A PRESS-localized spectroscopy box with the volume set to one
eighth of an imaging voxel was prescribed. Sixteen acquisitions were collected
with 0, ¼,
½ and ¾-voxel shifts in each of the two PSF-encoding directions. The
localization quality of the technique was also evaluated using an in-vitro phantom consisting of a lime
immersed in extra-virgin olive oil. The absence of contamination from neighboring
tissue was demonstrated under the condition of poor spatial resolution, which
is often the case in spectroscopic imaging to balance SNR. The imaging volume was
8×8×8 cm3, and acquisition matrix was 8×8×8 along EPI/PSF/PSF2
encoding directions, resulting in a voxel size of 1 ml.
In vivo Study:
Healthy
volunteer scans were conducted in this IRB-approved study. Imaging was
performed at 3T using an 8-element cardiac phase array coil with the following
parameters: TE/TR = 85/1200 ms, EPSI readout along the R/L direction, imaging
volume = 8×8×8 cm3, acquisition matrix = 8×8×8 along EPI/PSF/PSF2
encoding directions, voxel size = 1ml. The customized PROSE sequence was
prescribed on T2 weighted oblique axial slices, and a spectroscopy press box
was selected tightly within the prostate, and no saturation bands were
prescribed.
Results
PSF Mapping: As shown
in Fig. 1, the mapped PSF from the customized PSF-choice encoding scheme is
relatively free of ripples surrounding the main peak, compared to the standard
Fourier-based encoding scheme. The PSF is demonstrated for choline (Cho),
creatine (Cr) and the citrate (Cit) metabolic peaks.
Lipid Contamination:
By selecting the spectroscopy press box across
compartment of both lime and oil, the spectra in Fig 2a demonstrates
predominantly citrate signal within the lime voxels, and predominantly lipid
signal within the oil voxels. Note the lipid signal was attenuated due to
spectral filtering inherent to the PROSE sequence. The signal is constrained
within the press box in the sagittal plane where PSF-Choice encodings were
applied in both directions (Fig. 2b left), while there is evident ringing
artifact along the EPI read out direction due to the Fourier-based encoding
(Fig. 2b right).
Volunteer Study: Major metabolic
peaks within prostate have been identified, including Cho/Cr/Cit located at around
3.2/3/2.6 ppm respectively, and a representative spectrum within a single voxel is displayed in Fig. 3b.
Discussion
While the spatial localization was achieved along both PSF-encoding directions,
further improvement is needed in constraining the localization along EPI read
out direction to avoid fat signal contamination into selected press box.
Conclusion
We have demonstrated the features of PSF maps at metabolic peaks in a 2D
Gaussian-based PSF encoding scheme and also the comparisons of signal
contamination from surrounding tissue in in-vitro phantom. We further investigated
its feasibility with in-vivo studies for prostate MRSI
Acknowledgements
NIH R01CA160902, R01EB010195, P41EB015898 and BRI Microgrant.References
[1] Posse S., et al, MR spectroscopic imaging:
principles and recent advances. JMRI 2013;37(6):1301-25
[2] Panych LP, et al, Investigation of the PSF-choice
method for reduced lipid contamination in prostate MR spectroscopic imaging. MRM
2012 2012;68(5):1376-82.