Comparison of a novel, motion-robust MPRAGE imaging sequence with conventional MPRAGE imaging
Manojkumar Saranathan1, Puneet Sharma1, Unni Udayasankar1, and Rihan Khan1

1Dept. of Medical Imaging, University of Arizona, Tucson, AZ, United States

Synopsis

Conventional MPRAGE imaging is an integral part of most neuroimaging protocols. We investigated a new radial fan beam ordered MPRAGE sequence for its motion robustness and compared it to conventional MPRAGE imaging on 29 patients. The radial fan beam ordering had significantly better image quality than conventional centric ordered MPRAGE when rated for motion artifacts and overall diagnostic quality by 2 trained neuroradiologists.

Purpose

The MPRAGE sequence [1] is an integral component of clinical neuroimaging. It is also widely used for volumetry and as an anatomical reference, typically for fMRI studies. A typical 1mm3 isotropic resolution scan takes on the order of 4-5 min, making it susceptible to subject motion. We investigated a novel 3D MPRAGE sequence that employs a radial fan beam segmentation of ky-kz space and compared it with conventional MPRAGE in 29 patients for motion robustness and diagnostic preference.

Materials and Methods

Radial fan beam (RFB) segmentation [2] is an efficient way to traverse k-space, whilst achieving a true centric acquisition for improved contrast. 2D (ky-kz) k-space is segmented into M radial fan beams (Figure 1) and each fan beam is acquired following an inversion module. Within each fan beam, the N k-space points are ordered in increasing radial distance kr to maximize image contrast. This decouples Nz, the number of slices, from N. The points are still on a Cartesian grid, enabling fast online reconstruction.

We hypothesized that RFB MPRAGE would exhibit motion robustness similar to that of 2D radial imaging, especially on pediatric and geriatric patients compared to conventional MPRAGE.

After informed consent, 29 patients were scanned on a 1.5T scanner (HDx 15M4, GE Healthcare) using an 8-channel receive array head coil. Scan parameters: 24 cm FOV, 224x224x160 acquisition matrix, 1.2 mm thick sagittal sections, 10° flip, BW +12 kHz, TR 8.5-10ms, 2X parallel imaging, 3.3 min scan time. The TS/TI of the two sequences were adjusted to maintain comparable GM-WM contrast and scan time. The two MPRAGE sequences were acquired following injection of Gadolinum contrast. In half the patients, the order of the MPRAGE sequences were swapped to remove any image quality differences arising from delayed acquisition or motion.

For image analysis, the two sequences were rated independently by two experienced neuroradiologists in a blinded fashion. Motion artifacts were ranked on a scale of 1-4: 1 – Severe, could miss lesions; 2 – Moderate, unlikely to miss lesions; 3 – Mild, does not affect image quality or lesion detection; 4 – No artifacts. Overall image preference was graded on a 0 (no preference) to 2 (strongly preferred) scale, relative to one another. These ratings were subsequently adjusted by sign to reflect preference for RFB (+) or conventional (-). Inter-observer variability was assessed using Cohen's Kappa statistic and a Wilcoxon signed rank test was used to determine statistical significance.

Results

For motion artifacts, the mean image scores for RFB and conventional MP-RAGE were 3.82+0.39 and 2.97+0.82 for reader 1 and 3.69+0.54 and 2.86+0.88 for reader 2. Motion artifacts were significantly reduced in RFB compared to conventional MPRAGE (p < 0.05). Both readers overwhelmingly preferred RFB MPRAGE over conventional MPRAGE (mean values of 1 and 1.24 for readers 1 and 2). For both readers, RFB was preferred in 26/29 cases, conventional in 1/29 and no preference for 2/29 cases (κ = 1).

Figure 2 shows representative images from a 19-year old subject evaluated for concussion showing the effect of mild motion. RFB MPRAGE (right) is clearly superior to conventional MPRAGE (left) in minimizing artifacts and preserving image quality. Figure 3 shows representative images from a 71-year old subject who was unable to hold still. The streak artifacts in RFB MPRAGE (right) are more benign compared to the severe ghosting artifacts in conventional MPRAGE (left). Most other series in the examination were also degraded by motion artifacts. An 81-year old patient referred for glioblastoma is shown in Figure 4, again showing reduced artifacts with RFB MPRAGE (right) compared to conventional MPRAGE (left).

Conclusion

The radial fan beam segmentation, originally proposed for flexible choice of N (α pulses per inversion pulse) and 2D parallel imaging, was used for improving the motion robustness of MPRAGE imaging in this study. We demonstrated significantly reduced motion artifacts and improved diagnostic image quality (as measured by strong radiologist preference and scores) for the RFB MPRAGE sequence compared to conventional MPRAGE in a largely pediatric/geriatric patient population. This was achieved with no increase in scan time or advanced image or navigator-based motion correction schemes.

Acknowledgements

No acknowledgement found.

References

1. Mugler JP et al. JMRI 1:561-7 (1991)

2. Saranathan M et al. MRM 73:1786-94 (2015)

Figures

Figure 1. The radial fan beam segmentation scheme: within each fanbeam, the N points are ordered in increasing kr and each fanbeam is acquired per inversion pulse. Note that the points still lie on a Cartesian grid, enabling fast reconstruction.

Figure 2. Comparison of conventional MPRAGE (left) with the new RFB MPRAGE (right) obtained from a 19-year old patient imaged for concussion. Note the almost absence of motion artifacts in the RFB MPRAGE.

Figure 3. Comparison of conventional MPRAGE (left) with the new RFB MPRAGE (right) obtained from a 71-year old patient who was unable to hold still. Note the more benign streak artifacts in the RFB MPRAGE (right) compared to the significant ghosting in conventional MPRAGE (left). Most other series in this examination had motion artifacts.

Figure 4. An 81-year patient referred for glioblastoma. The RFB MPRAGE (right) shows reduced artifacts compared to conventional MPRAGE (left).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3296