Kathryn E Keenan1, Karl F Stupic1, Michael A Boss1, Stephen E Russek1, Tom L Chenevert2, Pottumarthi V Prasad3, Wilburn E Reddick4, Kim M Cecil5, Jie Zheng6, Peng Hu7, Edward F Jackson8, and Ad Hoc Committee for Standards in Quantitative MR9
1Physical Measurement Laboratory, National Institute of Standards and Technology, Boulder, CO, United States, 2University of Michigan, Ann Arbor, MI, United States, 3NorthShore University Health System, Evanston, IL, United States, 4St. Jude Children's Research Hospital, Memphis, TN, United States, 5Cincinnati Children's Hospital Medical Center, Cincinnati, OH, United States, 6Washington University in St. Louis, St. Louis, MO, United States, 7University of California Los Angeles, Los Angeles, CA, United States, 8University of Wisconsin, Madison, WI, United States, 9ISMRM, Berkeley, CA, United States
Synopsis
We
used the ISMRM/NIST system phantom to assess variations of T1 measurements
across MRI systems at 1.5 T and 3 T, to determine the repeatability and
reproducibility of the T1 measurements. This
study demonstrates that T1 variations from NMR-measured value are correlated
site-to-site within a vendor and by position within the head coil. The deviation from the
NMR-measured values is greater at 3 T than at 1.5 T. The VFA data has a larger
variation than IR; B1 inhomogeneity could contribute to the larger systematic
error in VFA measurements. The
ISMRM/NIST system phantom is an excellent tool for evaluating multi-site MRI
acquisition protocols.Purpose
Accurate
T1 relaxation time measurement is critical for many quantitative MRI
applications. An MRI system phantom (Figure 1 A) was developed through
collaboration between the ISMRM Ad Hoc Committee on Standards for Quantitative
MR and the National Institute of Standards and Technology (NIST).
1
The system phantom has SI-traceable components and is being monitored by NIST
for stability and accuracy. We used the system phantom to assess variations of T1
measurements across MRI systems at 1.5 T and 3 T. The purpose of this study was
to determine the accuracy, precision, repeatability and reproducibility of the
T1 measurements.
Methods
Two
system phantoms were imaged at multiple sites for three vendors across two
field strengths using head coils with 8 to 32 channels (Table 1 in Figure 1 C).
For each vendor, one location performed repeatability analysis (n=3
measurements) at 3 T; repeatability was assessed with the coefficient of
variation.
2 The study included T1 measurement by inversion recovery
(IR) using 2D FSE-IR and variable flip angle using 3D FSPGR. The T1 data was
fit using custom software (Phantom Viewer, developed at NIST). Details
regarding the phantom and data acquisition protocol are available online: https://collaborate.nist.gov/mriphantoms/bin/view/MriPhantoms/RecommendedImagingProtocols.
NiCl
2 was used to dope deionized water and create a range of T1
values from approximately 20 to 2000 ms. Across all measurements, reported
temperature ranged from 17.1 °C to 23.3 °C; temperature was
reported either of the MRI room or of the bulk water in the phantom.
Results
The deviations from the reference NMR T1
measurements at 20 °C for each of the vendors and both methods are shown for 1.5 T in
Figure 2 and 3 T in Figure 3. The repeatability study coefficients of variation
are presented in Figure 4. The CV for Vendor B, VFA is three times the CV of
the other vendors and methods; Vendor B, VFA also has the greatest deviation
from the NMR-measured values (Figure 3 B). T1 VFA deviation from reference
values for representative data sets vary by position within the head coil
(Figure 5).
Discussion
Systematic
differences in T1 relaxation time measurement can be identified as a function
of measurement method (IR or VFA), sample position and scanner type.
As
measured in our lab, the temperature variation of the NiCl
2 T1
relaxation time is non-linear and increases by 7.2% over the range of bore
temperatures (17.1 °C to 23.3 °C). Thus, the deviation from reference
T1 cannot be attributed only to temperature variation.
The
deviation from NMR-measured value is greater at 3 T than at 1.5 T.
The VFA data has a larger variation than IR; B1 inhomogeneity could contribute
to the larger systematic error in VFA measurements. Unlike 3 T, the 1.5 T data
do not exhibit an obvious pattern of deviation from the NMR-measured values,
with the exception of Vendor B, which consistently underestimates T1 with VFA.
However, additional data is needed for vendors A and B.
This study demonstrates that T1 variations
from NMR-measured value are correlated site-to-site within a vendor and by
position within the head coil. The ISMRM/NIST system phantom is an excellent
tool for evaluating multi-site MRI acquisition protocols.
Acknowledgements
We appreciate the efforts of all those who
completed scans of the phantom, without them this study would not be possible.
The SQMR committee members are listed here: http://www.ismrm.org/members-only/committee-directory/#SQMR.References
1. Russek SE, Boss MA, Jackson EF, et al.
Characterization of NIST/ISMRM MRI System Phantom. Proceedings of the 20th
meeting of the ISMRM, Melbourne, Australia, 2012. 2. Sullivan DC, Obuchowski
NA, Kessler LG, et al. Metrology Standards for Quantitative Imaging Biomarkers.
Radiology. 2015.