Anh Tu Van1, Barbara Cervantes2, Ernst J Rummeny2, Axel Haase1, and Dimitrios Karampinos2
1Zentralinstitut für Medizintechnik, Technische Universität München, Garching, Germany, 2Diagnostic and Interventional Radiology, Technische Universität München, Munich, Germany
Synopsis
Diffusion-preparation
is regaining popularity recently thanks to its compatibility with any type of
readout scheme. Using Bloch simulations, the current work provides
understanding of the motion-induced magnitude and phase modulation of the
signal obtained with diffusion-prepared sequences. When motion induces a
constant phase across the voxel, dephasing signal before tipping up and
rephasing it before readout should be used to remove motion-induced magnitude
modulation. When the induced phase is not constant, smaller voxel sizes assist
towards stabilizing both magnitude and phase of the signal. In vivo data using
diffusion-prepared 3D TSE are shown supporting the observations of the Bloch
simulations. Introduction
Diffusion-preparation (DPrep) is regaining popularity thanks to its compatibility with any type of readout [1-4]. However, motion during diffusion-encoding results in spatially
varying spin phase that modulates the magnitude of
both longitudinal and transverse magnetization after the tip-up RF pulse (Fig.1). To mitigate phase-induced
magnitude modulation, previous studies added a dephase gradient before the tip-up
RF pulse and a rephase gradient in the readout [1,2]. The goal of this work is to characterize
the effectiveness of the dephase/rephase gradient in reducing the motion-induced
signal modulation and to propose additional strategies for improving the motion
insensitivity of DPrep sequences.
Theory
Letting $$$\phi_i$$$ be the spin phase after the diffusion
encoding period, $$$\theta_i$$$ be the phase induced by the dephasing gradient,
the longitudinal magnetization after the tip-up pulse is
$$m_i^z=cos(\phi_i+\theta_i)=\frac{1}{2}(e^{j(\phi_i+\theta_i)}+e^{-j(\phi_i+\theta_i)})\
(1)$$
After the excitation, refocusing pulse, and the rephase gradient,
the transverse magnetization of a spin is
$$m_i^{xy}=\frac{1}{2}(e^{j(\phi_i+\theta_i)}+e^{-j(\phi_i+\theta_i)})e^{-j\theta_i}e^{-j\pi/2}$$
$$m_i^{xy}=\frac{1}{2}(e^{j\phi_i}+e^{-j(\phi_i+2\theta_i)})e^{-j\pi/2}$$
and the measured magnetization of a voxel is
$$M^{xy}=\sum_{i \in
voxel}\frac{1}{2}(e^{j\phi_i}+e^{-j(\phi_i+2\theta_i)})e^{-j\pi/2}\ (2)$$
Equation (2) implies that the resulted voxel signal depends on
both the motion-induced spin phase $$$\phi_i$$$ and the phase added by the
dephase/rephase gradients pairs $$$\theta_i$$$. Furthermore, with arbitrary
varying $$$\phi_i $$$ it is almost impossible to design dephase/rephase
gradients that can mitigate the phase-induced signal modulation.
Consider the special case where $$$\phi_i$$$ varies slowly enough
spatially so that over a voxel with small enough size it is approximately
constant, meaning $$$\phi_i \approx \phi_{voxel}$$$ for all spins within this
voxel (call Condition 1), Equation (2) can be rewritten as
$$M^{xy} =
\frac{1}{2}e^{j(\phi_{voxel}-\pi/2)}(1–e^{-j2\phi_{voxel}}\sum_{i \in
{voxel}}e^{-j2\theta_i}) \ (3)$$
Assuming further that the dephase/rephase gradients are designed
so that $$$\sum_{i \in voxel}e^{j2\theta_i} = 0$$$ (call Condition
2), the transverse magnetization of the voxel is
$$M^{xy}=\frac{N_{spin}}{2}e^{j(\phi_{voxel}-\pi/2)}\ (4)$$
where $$$N_{spin}$$$ is the total number of voxel spins. Equation
(4) implies that while transverse magnetization has magnitude that is
independent of motion-induced phase, its phase depends on the motion-induced
phase. There are two important points that Equation (4) and its associated necessary
conditions implies on the phase sensitivity of the DPrep sequence:
1) When the voxel size decreases, Condition 1 can
be met by more $$$\phi_i$$$ patterns and therefore the resulted signal is more
robust in avoiding motion-induced magnitude modulation.
2) DPrep sequence is not readily applicable to multi-shot
acquisition due to the motion-dependent magnetization phase. To enable
multi-shot acquisition special diffusion-encoding schemes that reduce motion
effects such as velocity-compensated or acceleration-compensated diffusion
encoding [4-6] or phase navigation must be used.
Method
Simulation: Bloch
simulations were performed considering two types of motion-induced phase
$$$\phi_i$$$: constant phase in $$$(-\pi, \pi)$$$ and linear phase across the
voxel $$$\alpha x+\pi/4$$$ with $$$\alpha\in(-\pi/mm,\pi/mm)$$$. Different
voxel sizes were simulated to show the dependence of motion robustness on voxel
size.
In vivo
experiment: Sagittal diffusion-weighted
imaging of the upper knee of a healthy volunteer was conducted using a
16-channel knee coil on a 3 T system (Ingenia, Philips Healthcare) with a
diffusion-prepared 3D TSE sequence, with sequence parameters: FOV=160×127×100
mm3, voxel=1.7×1.7×1.7mm3, TR/TE=1700/19, TSE factor =60, Nex=2, total scan
duration = 3min40s, using both traditional and velocity-compensated diffusion
preparation at b=400s/mm2.
Results
Simulation: Without dephase/rephase gradients, magnitude of
the simulated voxel at echo oscillates even when $$$\phi_i$$$ is constant over
the voxel (Figure 2a). When $$$\phi_i \in (-\frac{\pi}{2}, \frac{\pi}{2})$$$,
spins are tipped to +z, otherwise spins are tipped to -z, which results in the
toggling of the signal phase between $$$-\pi/2$$$ and $$$\pi/2$$$ states. When
dephase/rephase gradients are used, the signal magnitude is stable for all
values of $$$\phi_i$$$ while the phase varies and equals to $$$-\pi/2+\phi_i$$$
(Figure 2b). Figure 3 shows higher robustness of smaller voxel size in
mitigating magnitude modulation caused by linear phase across the voxel. Both
magnitude and phase of the signal exhibit smaller change with respect to the
slope of the linear phase as the voxel size decreases.
In-vivo: Diffusion-weighted images of the human knee are
shown in Figure 4. Since the readout module is multishot, dephase/rephase
gradients alone cannot help with the motion problem and velocity-compensated
diffusion preparation is needed to recover the signal loss. When
dephase/rephase gradients are not used, motion-induced signal loss occurs in
localized area. Since the rephase gradient also has the capability to crush
T1-recovered signal, using dephase/rephase gradients gives better fat
suppression.
Conclusion
We have shown that in Dprep sequences,
motion-induced magnitude modulation can be avoided when employing
dephase/rephase gradients. However, the signal phase remains motion-dependent,
which makes multishot acquisition possible only with motion-compensated Dprep. When the induced phase is
not constant, smaller voxel size is more robust in stabilizing both magnitude
and phase of the signal.
Acknowledgements
We would like to thank Philips
Healthcare for research support.References
[1] Jeong, Magn Reson Med 56:1173, 2006; [2] Alsop,
Magn Reson Med 38:527, 1997; [3] Cervantes, Proc. ISMRM 2015, p. 101; [4]
Nguyen, Magn Reson Med 72:1257, 2014; [5] Welsh, IEEE TMI 34: 1843, 2015; [6] Stoeck, Magn Reson Med, doi: 10.1002/mrm.25028.