High-resolution imaging of muscular fat fraction - comparison of chemical shift-encoded imaging and T2-based imaging
Lena Trinh1, Emelie Lind1, Pernilla Peterson1, and Sven Månsson1

1Dept. of Translational Medicine, Medical Radiation Physics, Skåne University Hospital, Lund University, Malmö, Sweden

Synopsis

Chemical shift-encoded imaging is a quantitative method commonly used to estimate fat fraction (FF) in various body parts. However, for a reliable assessment, this technique requires short inter echo spacing which can be challenging if high spatial resolution is desirable. An alternative quantitative method, based on the difference in T2-relaxation time between fat and water, was examined and compared to the chemical shift-encoded imaging method. This T2-based technique successfully estimated FF in phantoms at high resolution and large matrix size, when the chemical shift-encoded method failed.

Purpose

In quantitative studies of fat infiltration in skeletal muscle, such as studies of e.g. diabetes1, chemical shift-encoded imaging is commonly used. However, this technique requires short echo spacing to be reliable, especially at 3T, which is in contradiction to the high spatial resolution and high matrix size needed to resolve the detailed fat patterns in the muscle. The purpose of this study is to use a fat quantification technique based on T2-relaxation2, and evaluate its accuracy and precision at both low and high spatial resolution compared to chemical shift-encoded imaging. Fat quantification based on the different T2 relaxation times of water and fat, rather than the difference in resonance frequency, may be a way to obtain very high resolution images of fat fraction (FF), in turn improving the differentiation between intermuscular and intramuscular fat.

Methods

A phantom was constructed using 6 vials filled with different ratios of water and Intralipid fat emulsion (20%, Fresenius Kabi) to achieve a range of FFs (0, 5, 10 and 20 %). In vivo data was acquired from a calf of a heathy volunteer. Two sequence types were used for phantom and in vivo data on a 3T scanner (Trio, Siemens Healthcare): multi spin echo (MSE) for the T2-based method and multi gradient echo (MGE) for the chemical shift-based method. Images were acquired at two matrix sizes (128x128 and 512x512). The following settings were used for the MSE-acquisitions: TE1/ΔTE/TR (phantom) = 11.7/11.7/2000 ms, TE1/ΔTE/TR (in vivo) = 9.20/9.20/2000 ms and BW128/BW512 = 425/391 Hz/px. The settings for the MGE-acquisitions were: TE1128/ΔTE128/TR128 = 1.34/1.55/500 ms, TE1512/ΔTE512/TR512 = 2.57/3.92/500 ms, BW128/ BW512 = 1776/651 Hz/px and flip angle = 12. Slice thickness and FOV were set to 5 mm and 280x280 mm2 for all acquisitions.

FF was calculated from MSE-data by fitting the linear parameters W (water) and F (fat) in the signal equation $$$S = We^{-t/T2_{W}}+Fe^{-t/T2_{F}}$$$, using fixed values of T2F and T2W2. The T2-values were obtained by drawing regions of interest (ROIs) in a T2-map, calculated from a monoexponential fit in each voxel. In the phantom measurement, the T2 values of water and Intralipid were used. Therefore, the FF map of the phantom represented the Intralipid fraction and were divided by five to get the actual FF values, knowing the true FF of undiluted Intralipid. In the in vivo case, ROIs were drawn in subcutaneous fat and in muscle tissue. An iterative least squares procedure, correcting for T2* and off resonance and including a multipeak fat model, was used to separate the fat and water signals from MGE data3,4. To study the effect of deviations between assumed and actual T2-values on the estimation of FF, simulations were carried out.

Results

The resulting FF maps of the phantom and calf using both methods are presented in Figure 1 and Figure 2, respectively.

In Figure 3, the mean estimated FFs are plotted against the known FFs in the phantom. Chemical shift-encoded imaging underestimates the FF at higher resolution while the T2-based method estimates the FF accurately. At the smaller matrix size and lower resolution, both methods perform well. The FF variation within the vials were markedly larger for chemical shift-encoded imaging than for the T2-method at the higher matrix size.

The estimated FFs in three muscle ROIs (tibialis anterior, soleus and gastrocnemius) are presented in Figure 4. Compared to the FFs obtained using chemical shift-encoded imaging at low resolution, the other approaches either under- or overestimates the FF.

The simulated errors in FF due to inaccurate T2Fat and T2Muscle are presented in Figure 5. These results suggest that it is more important to estimate T2Muscle accurately compared to T2Fat at the relevant FF (muscle FF ≈ 5%).

Discussion

At higher resolution, chemical shift-encoded imaging underestimates the FF in phantom and in vivo whereas the T2-based method accurately estimates the FF in phantom but overestimates it in vivo. It is important to accurately estimate T2, especially of the muscles, to get a correct FF estimation. This, however, is not always easy to achieve with the presence of other not visible components within the ROI, e.g. fat. This may in part explain why the T2-method overestimates FF in vivo at high resolution, and why the T2-method at low resolution performs poorly in some muscle ROI. At low resolution, the FF calculation is also more sensitive to the effect of Gibbs artifacts which results in inferior performance.

Conclusion

The phantom results show that the T2-based method can achieve an accurate estimation of FF at both high and low resolution, compared to chemical shift-encoded imaging.

Acknowledgements

This study was supported by grants from The Magnus Bergvall Foundation and Director Albert Påhlsson's Trust for Researchers and Charity.

References

1. Goodpaster BH, Wolf D. Skeletal muscle lipid accumulation in obesity, insulin resistance, and type 2 diabetes. Pediatric Diabetes 2004:5:219-226.

2. Kan HE, Scheenen TWJ, Wohlgemuth M, et al. Quantitative MR imaging of individual muscle involvement in fascioscapulohumeral muscular dystrophy. Neromuscular Disorders 2009:19:357-362.

3. Yu H, Shimakawa A, McKenzie CA, et al. Multiecho Water-Fat Separation and Simultaneous R2* Estimation Wth Multifrequency Fat Spectrum Modeling. Magnetic Resonance in Medicine. 2008:60:1122-1134.

4. Ren J, Dimitrov I, Sherry AD, and Malloy CR. Composition of adipose tissue and marrow fat in humans by 1H NMR at 7 Tesla. J Lipid Res. 2008:49(9):2055-62.

Figures

Estimation of FF in phantom using chemical shift-encoded imaging at matrix size A) 128x128 B) 512x512, and T2-based method at C) 128x128 D) 512x512. The known FFs are 5%, 10%, 20%, 0%, 10% and 20% in clockwise direction. T2Water=1820 ms and T2Intralipid=291 ms were used in the T2-based method.

Estimation of FF in the left calf of a healthy volunteer using chemical shift-encoded imaging at matrix size A) 128x128 B) 512x512, and T2-based method at C) 128x128 D) 512x512. T2Muscle=38.2 ms and T2Fat=152 ms were used in the T2-based method.

The estimated FFs and the standard deviations in phantom using chemical shift-encoded imaging (CSEI) and the T2-based method, at matrix sizes 128x128 and 512x512, are shown. The standard deviation of the FF measured in the water (FF=0%) using the T2-based method at 512x512 matrix was ±34% (not fully shown).

The estimated FFs and the standard deviations in vivo using chemical shift-encoded imaging (CSEI) and the T2-based method, at matrix sizes 128x128 and 512x512, are shown. The standard deviations not fully shown for T2-based 128x128 are ±6.5 % (tibialis anterior) and ±8.9 % (gastrocnemius).

The simulated estimation of FF when using inaccurate T2-value, A) True T2Fat = 152 ms and B) True T2Muscle = 38 ms, in the T2-based method.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3281