Stefan Ruschke1, Holger Eggers2, Hendrik Kooijman3, Houchun H. Hu4, Ernst J. Rummeny1, Axel Haase5, Thomas Baum1, and Dimitrios C. Karampinos1
1Department of Diagnostic and Interventional Radiology, Technische Universität München, Munich, Germany, 2Philips Research, Hamburg, Germany, 3Philips Healthcare, Hamburg, Germany, 4Radiology, Phoenix Children’s Hospital, Phoenix, AZ, United States, 5Zentralinstitut für Medizintechnik, Technische Universität München, Garching, Germany
Synopsis
As the spatial resolution of a multi-echo
gradient-echo imaging sequence increases, the echo time step increases
resulting in an increased echo time step and an echo time selection that
degrades the noise performance of water-fat separation. Time-interleaved gradient
echo imaging combined with bipolar (flyback) gradients can achieve reasonable
echo time steps at high resolution without dramatically increasing scan time.
However, bipolar gradients are associated with known phase errors problems,
which can lead to fat quantification errors. The present work develops a
methodology for acquiring bipolar time interleaved multi-echo gradient echo
data and for correcting the relevant phase errors.Purpose
An important consideration in the design of multi-echo gradient-echo
pulse sequences employed for water-fat imaging is the trade-off between spatial
resolution and echo time step. As the spatial resolution increases, the echo
time step also increases and can yield to an echo time selection that degrades
the noise performance of water-fat separation [1]. Time-interleaved gradient
echo imaging sequences using mono-polar (non-flyback) gradients can reduce the
echo time step with the disadvantage of requiring more shots and increasing
thus scan time. Time-interleaved gradient echo imaging can be alternatively
combined with bipolar (flyback) gradients in order to achieve reasonable echo
time steps at high resolution without increasing scan time [2-5]. However,
bipolar gradients are associated with known phase errors problems, which can
lead to fat quantification errors [4-5]. The present work develops a
methodology for acquiring bipolar time interleaved multi-echo gradient echo and
for correcting the relevant phase errors.
Methods
Sequence & Phase Correction:
A bipolar time-interleaved multi-echo gradient echo sequence was implemented (Fig. 1). For phase correction purposed, a reference scan was performed before the actual measurement. It acquired the center k-spacpe line twice for echo time with both polarities is played out before the actual sequence. A 1D phase correction was then performed based on the reference scan data. [6] Furthermore, the concomitant gradient coefficients [7] are calculated based on the gradient waveforms and then corrected during the reconstruction process.
Signal model:
The signal model as described by Peterson et al. [8] was used, fitting for compelx water and fat, a common T2*, fieldmap and a complex error map term:
$$ S\left(t_n\right)=\left(\rho_w+\rho_f\sum_{m=1}^{M}\alpha_me^{i2\pi{}\Delta{}f_mt_n}\right)e^{i2\pi\psi{}t_n}e^{\left(-1\right)^ni\theta}$$
Measurements:
All studies were performed on a
3T scanner (Ingenia, Philips Healthcare).
Agar based water-fat phantoms [9] with nominal
fat fractions of 0, 5, 10, 15 and 100 % have been used in the phantom experiment.
Used parameters in vivo pancreas experiment: The Bi-polar TIMGRE used 2x3 echoes, voxel size = 0.9 x 0.9 x 5mm3, TR = 8 ms, TE1 = 2, dTE = 0.8;
Used parameters in vivo spine experiment: The Bi-polar TIMGRE used 2x3 echoes, voxel size = 0.9 x 0.9 x 10 mm3, TR = 8 ms, TE1 = 2, dTE = 0.8; mono-polar TIMGRE used 2x3 echoes, voxel size = 2.2 x 2.2 x 10 mm3, TR = 7.9 ms, TE1 = 2, dTE = 0.8.
Results
A good agreement was achieved in estimating both fat fraction and T2*
values in the water-fat phantoms between the monopolar and bipolar sequences
(Figure 2). Figures 3 and 4 compare the fat quantification results between the
two sequences in the calf muscle and the lumbar spine respectively, showing a
good visual agreement between the two methods. Figure 5 shows the fat
quantification results in the abdomen using the bipolar sequence. Despite the
relatively low SNR of the abdominal scan, the pancreas geometry can be well
delineated with the bipolar sequence.
Discussion & Conclusion
A novel methodology was developed incorporating a bipolar time-interleaved multi-echo gradient echo acquisition with a reconstruction process correcting for phase error effects. The fat quantification accuracy of the developed methodology was validated in phantoms and the feasibility of the method was shown in vivo in three different body regions. The developed bipolar sequence was compared with a standard monopolar sequence, showing comparable water-fat separation and fat quantification results, but at higher spatial resolution than the monopolar sequence. The proposed methodology therefore enables reliable fat fraction mapping at high spatial resolution and within reasonable acquisition times and might be useful in measuring fat fraction changes in smaller organs (e.g. the pancreas).
Acknowledgements
The present work was supported by Philips
Healthcare.References
[1] Pineda, Magn Reson Med 54:625, 2005,
[2] Lu, Magn Reson Med 60:198, 2008,
[3] Yu, J Magn Reson Imaging 31:1264, 2010,
[4] Petterson, Magn Reson Med 71:219, 2014,
[5] Sollman, Magn Reson Med, 2015, doi: 10.1002/mrm.25807,
[6] Ruscke, Proc. ISMRM 2015, p. 3657,
[7] Bernstein, Magn Reson Med 38:300, 1998,
[8] Peterson, Magn Reson
Med. 2014 Jan;71(1):219-29,
[9] Hines, Magn Reson Med. 2009;30:1215–1222.