Klaus Eickel1,2,3, Lutz Lüdemann3, David Porter1, and Matthias Günther1,2
1Fraunhofer MEVIS, Bremen, Germany, 2mediri GmbH, Heidelberg, Germany, 3University Hospital Essen, Essen, Germany
Synopsis
The
application of simultaneous multi-slice imaging to a segmented EPI allows the acquisition of multiple
contrasts while retaining sufficient temporal resolution and spatial coverage.
Contrast-enhanced perfusion measurements were performed on a pig’s hip/leg
resulting in mainly muscle-perfusion images with separated S0 and
T2* maps. The separation of the different signal contributions potentially allows for a quantitative analysis in contrast-enhanced dynamic imaging.Introduction
A
number of acceleration strategies have been proposed for MRI, especially for
dynamic imaging, where there is a trade-off between temporal and spatial
resolution. To acquire multiple contrasts within one dataset for simultaneous T1
and T2* measurements1,2,3 additional
acceleration is required. Simultaneous multi-slice imaging
(SMS) is a complimentary technique4,5 which can be
combined with standard k-space sampling strategies, such as parallel imaging
and keyhole methods. CAIPIRINHA-like6
signal shifting for more efficient use of coil-sensitivity profiles improves SMS
reconstruction7. A multi-shot segmented EPI yields sufficient
signal also for longer echo-times and allows a flexible acquisition scheme with
respect to the desired echo times. Additionally it enables integrated
controlled slice-aliasing to reduce g-factor penalty5. To
satisfy these aspects we introduce a novel incorporation of SMS to a multi-contrast,
segmented EPI sequence8 (Fig.1) which allows extended spatial coverage in contrast-enhanced dynamic imaging to fully cover larger parts of the body (e.g.
leg’s muscles) without exceeding temporal constrains.
Subjects and Methods
One
sedated pig was studied with an 1.5T MR-scanner (Aera, Siemens
Healthcare/Germany). After bolus-injection of a contrast-agent (CA) (dose of 0.1
mmol/kg at 5ml/s injection-rate of Dotarem, Guebert/France) dynamic images of
the hip and right leg were acquired for 11 minutes according to 500
measurements at a temporal resolution of 1.3 s. For a FOV=400x200 mm
2
at a spatial resolution of 3.1 x 3.1mm
2 with 24 slices of 5 mm thickness
(20% gap, interleaved acquisition scheme to reduce inter-slice crosstalk) could
be achieved due to the 4 times (R=MB) accelerated imaging speed. Despite the
sampling rate and a reasonable FOV three contrasts (TE1=9 ms, TE2=1.5 ms,
TE3=34 ms) with a TR=261 ms could be acquired without additional in-plane
acceleration. The 4 simultaneously excited slices (MB=4) were shifted by ¼ FOV
along phase-encoding direction (PE) with blipped CAIPI where gradient blips were
integrated into multi-shot scheme
8,9. Sagittal slice
orientation was chosen such that the aorta and femoral artery were covered for detection of the arterial input
function (AIF). For offline SMS reconstruction a 5x5 MB-kernel was trained with
MB and Single-Band (SB) data which was acquired before CA administration and with
identical imaging parameters (TR
ACS=1500 ms and TE
ACS=TE
SMS).
Slice-GRAPPA reconstruction of the SMS data potentially regularizes the SNR of
the SMS data according to the SNR of the auto-calibration-signal (ACS)
5,10
which reduces apparent T1 weighting in the SMS images. For comparison two SMS-reconstruction
approaches were tested: SB & MB (both TR=1500 ms) and SB & synthetic MB
from SB (TR=1500 ms), in both cases 2 prescans were run for steady-state
conditions. Time penalty for ACS data acquisition was 20s prior to the dynamic
scan.
Collecting multiple contrasts in one acquisition allows signal separation into
a S
0 (TE=0 ms) and T2* signal by fitting the data to the exponential
T2* decay after SMS reconstruction.
Results
The
application of SMS yields sufficient temporal resolution for AIF detection in a
large vessel while anatomical coverage allows perfusion analysis of manually
segmented muscles. As addressed by the FLEET-ACS approach
10
echo-spacing is crucial for correct contrast recovery during slice-GRAPPA
reconstruction, in addition to echo-spacing also the TR has to be kept constant
for MB-kernel training data. The quality of the reconstructed images meets the
requirements for post-processing (Fig.2) such that signal separation by fitting the
different acquired contrasts (Fig.3) can be successfully performed. By
correcting the signal for the T2* contribution a correction of the
susceptibility related changes can be achieved which becomes especially
important for the distribution phase with high CA concentrations
1.
Furthermore T1 effects of CA leakage can be quantified
11.
Figure 4 shows the S
0(t) signal inside the femoral artery for the first 50
measurements, enabling AIF extraction. In contrast to regions with fast
CA-dynamics the S
0(t) time-curve in muscle-tissue represents the
extravasation process into the tissue and the wash-out after a maximum CA-concentration
has be reached (Fig.5).
Discussion and Conclusion
By using a segmented EPI acquisition
scheme echo-times can be selected flexibly and reasonable signal in late
contrasts can be achieved. TR and TE issues in SMS reconstruction have been
addressed and the resulting SMS perfusion images show expected signal dynamics.
Contrast-enhanced perfusion measurements were performed on a pig’s hip/leg
resulting in mainly muscle-perfusion images with separated S0(t) and
T2* maps which allow a more complex pharmacokinetic analysis1,2. For further acceleration and/or shorter minimal TEs SMS
segmented EPI can be combined with (in-plane) undersampling strategies5.
It has be demonstrated that the incorporation of SMS
into multi-contrast, segmented EPI can overcome existing limitations for
dynamic imaging while retaining desired acquisition parameters such as temporal and
spatial resolution even for imaging of relatively large FOVs.
Acknowledgements
Experimental setup and facility provided by University Hospital Essen and Deutsche Forschungsgesellschaft. Support for sequence development was also recieved from David Feinberg and Liyong Chen.References
[1] Heilmann, M. et al. Simultaneous dynamic T1 and T2* measurement for AIF assessment
combined with DCE MR in a mouse tumor model. Magnetic Resonance Materials in Physics, Biology and Medicine 2007;20(4):193-203.
[2] Lüdemann, L. et al.: Simultaneous Quantification
of Perfusion and Permeability in the Prostate Using Dynamic Contrast-Enhanced
Magnetic Resonance Imaging with an Inversion-Prepared Dual-Contrast Sequence.
Annals of Biomedical Engineering 2009;37(4):749-762.
[3] Sourbron, S. et al.
Bolus-tracking MRI with a Simultaneous T1- and T2*-Measurement. Magnetic
Resonance in Medicine 2009;62(3):672-681.
[4] Feinberg, D. et al. Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI
and Fast Diffusion Imaging. PloSone 2010; 5(12):e15710.
[5] Setsompop, K. et
al. Blipped-Controlled Aliasing in Parallel Imaging for Simultaneous Multislice
Echo Planar Imaging With Reduced g-Factor Penalty. Magnetic Resonance in
Medicine 2012;67(5):1210-1224.
[6] Breuer, F. et al.
Controlled Aliasing in Parallel Imaging Results in Higher Acceleration
(CAIPIRINHA) for Multi-Slice Imaging. Magnetic Resonance in Medicine
2005;53(3):684-691.
[7] Larkmann, D. et al.
Use of Multicoil Arrays for Separation of Signal from Multiple Slices
Simultaneously Excited. Journal of Magnetic Resonance Imaging 201;13(2):312-317
[8] Eickel, K. et al.
Multi-Contrast Segmented EPI Sequence using Simultaneous Multi-Slice
Acquisition with Controlled Aliasing Readout for Dynamic Imaging: First Results.
Abstract ISMRM Workshop on SMS Imaging, Pacific Grove, USA 2015.
[9] Polimeni, J. et al.
Rapid multi-shot segmented EPI using the Simultaneous Multi-Slice acquisition
method. Abstract ISMRM, Melbourne, Australia 2012.
[10]
Polimeni, J. et al. Reducing Sensitivity Losses Due to Respiration and Motion
in Accelerated Echo Planar Imaging by Reordering the Autocalibration Data
Acquisition. Magnetic Resonance in Medicine 2015; doi: 10.1002/mrm
[11]
Eichner, C. Slice Accelerated Gradient-Echo Spin-Echo Dynamic Susceptibility
Contrast Imaging with Blipped CAIPI for Increased Slice Coverage. Magnetic
Resonance in Medicine 2014;72(3):770-778.