Kosuke Ito1, Atsushi Kuratani1, Yukio Kaneko2, and Masahiro Takizawa1
1Healthcare Company, Hitachi Ltd., Chiba, Japan, 2Research and Development Group, Hitachi Ltd., Tokyo, Japan
Synopsis
A technique of
ICA-selective 3D TOF imaging using asymmetrically RF-shimmed Pre-saturation pulse
is presented. By using 4-channel RF transmit coil, spatially asymmetric RF
transmission is achieved. Then apply RF shimming parameter to Pre-saturation
pulse, the saturation effect becomes spatially asymmetric, and achieved selectively
visualize blood flow from each ICA. This technique provides high contrast
between blood and brain parenchyma because it does not make TR longer.Purpose
The purpose of
this study is imaging blood flow from each internal carotid artery (ICA) by using
RF shimming with 4-channel RF transmit coil at 3T.
Introduction
RF shimming
1 is routinely
used in high field MRI. Recently, to improve B
1 homogeneity in
specific region or to reduce SAR, spatially asymmetric RF transmission using RF
shimming with 4-channel RF transmit coil has been reported
2-4. In
this study, we applied asymmetrically RF-shimmed pre-saturation (PreSat) pulse
in 3D TOF imaging to selectively image blood flow from left or right ICA. ICA-selective
3D TOF imaging with applying 2D excitation to PreSat pulse has already been
presented
5. The 2D excitation can select blood flow from each ICA, but
2D excitation also causes long duration of PreSat pulse leading to long TR and long
scan time. Especially, the long TR sometimes lowers image contrast between
blood and brain. The presented asymmetrically RF-shimmed PreSat pulse is
expected to visualize blood flow from each ICA without using longer TR.
Materials and
Methods
Experiments
Experiments were
conducted using 3T whole body MRI system (Hitachi, Ltd.). B1 maps
were acquired using 4-channel RF transmit and receive coil6. A 16-channel
receive only head coil was used for 3D TOF scan.
Calculation
of RF shimming parameters for PreSat pulse
Figure 1 shows
the slice of PreSat pulse with blue rectangle. The slice of PreSat pulse is
located at just below the FOV. B1 map at the slice was acquired with
multi Td method7, 8. RF shimming parameters were calculated as follows.
1. Set two ROIs
(ROIT and ROIF) to corresponding to locations of two
ICAs. The size of ROIs was 23 mm × 23 mm (Fig. 2).
2. RF shimming
parameters were determined to minimize mean of B1 in ROIT,
subject to mean of B1 in ROIF is equal to mean of B1
with QD transmission.
As B1
in ROIF becomes small, blood flow from ROIF was expected not
to be saturated by the PreSat pulse. Thus blood flow from ROIF is
expected to be visualized.
3D TOF scan
5 healthy
volunteers were evaluated. Scan parameters were as follows; TR 20 ms, TE 3.3
ms, FA 18 deg, and matrix 352 x 320 x 118. Scan time was 4:22. 3D TOF was
performed with and without PreSat pulse. The PreSat pulse was applied to left
or right ICA.
Evaluation
B1
mean in ROIT and ROIF were evaluated. Signal intensity of
two ICAs is also compared to signal intensity of white matter.
Results and
Discussions
Figure 2 shows B
1
maps at PreSat slice. (a): ROI
T is set at left ICA, and (b): ROI
T
is set at right ICA. In both cases, spatially asymmetric B
1 distribution
is achieved. Table 1 shows means of B
1 in ROI
T and ROI
F.
Mean of B
1 in ROI
F is smaller than that in ROI
T
by 87%. Figure 3 shows MIP images of a volunteer, (a): without PreSat pulse,
(b): ROI
T is set at left ICA, and (c): ROI
T is set at
right ICA. In the image without PreSat pulse, blood flow from both left and
right ICA are imaged, and with PreSat pulse, blood flow from ROI
T is
saturated, so blood flow from left or right ICA is selectively imaged. In table
2, signal intensity of white matter and left and right ICA are summarized. Signal
intensity of white matter and ICA of ROI
F was almost same between
with and without PreSat pulse. By applying PreSat pulse, the signal intensity of
blood flow from ROI
T becomes smaller than that of white matter, so
no blood signal is seen in the MIP images. These results show that presented asymmetrically
RF shimmed PreSat pulse minimizing the B
1 in ROI
F enables
selectively imaging blood flow from only left or right ICA.
Conclusions
ICA selective
TOF imaging by using asymmetrically RF-shimmed PreSat pulse was presented. The
results demonstrates this imaging technique effectively visualize blood flow
from left or right ICA selectively. This technique provides high contrast
between blood and brain parenchyma because it does not make TR longer.
Acknowledgements
No acknowledgement found.References
[1] J. Nistler
et al. ISMRM 2007; 15: 1063. [2] K. Ito et al. ISMRM 2014; 22: 942. [3] Y.
Kaneko et al. ISMRM 2013; 21: 2756. [4] Y. Kaneko et al. ISMRM 2015; 24: 3132.
[5] T. Nishihara et al. ISMRM 2012; 20: 2497. [6] Y. Soutome et al. ISMRM 2013;
21: 2756. [7] K. Ito et al. ISMRM 2013; 21: 2598. [8] K. Ito et al. ISMRM 2013;
21: 2599.