Curtis Wiens1, Nathan S. Artz1,2, Hyungseok Jang1, Alan McMillan1, Kevin Koch3, and Scott B. Reeder1,4,5,6,7
1Radiology, University of Wisconsin, Madison, WI, United States, 2Diagnostic Imaging, St. Jude Children's Research Hospital, Memphis, TN, United States, 3Biophysics and Radiology, Medical College of Wisconsin, Milwaukee, WI, United States, 4Medical Physics, University of Wisconsin, Madison, WI, United States, 5Biomedical Engineering, University of Wisconsin, Madison, WI, United States, 6Medicine, University of Wisconsin, Madison, WI, United States, 7Emergency Medicine, University of Wisconsin, Madison, WI, United States
Synopsis
Large magnetic susceptibility differences between metallic implants and
tissue generate severe B0 inhomogeneities that present several
challenges for MR: excitation of the entire off-resonance spectrum, distortion
in the frequency encoding direction, and severe intra-voxel dephasing. In this work we propose an ultra-short echo time acquisition with
broadband excitation. Single point encoding was use to avoid in-plane
distortions while 3D undersampling facilitated clinically feasible acquisition
times. The effects of RF pulse duration
on signal loss near the implant and T1 weighting were evaluated in a
total hip replacement phantom and a volunteer with a total knee replacement.Introduction
The large magnetic susceptibility differences that
occur between metallic implants and tissue generates severe B
0 inhomogeneities. This leads to several challenges
including excitation of the entire off-resonance spectrum, distortion in the frequency
encoding direction, and severe intra-voxel dephasing. 3D-Multispectral Imaging (3D-MSI) techniques such
as MAVRIC (1) and SEMAC (2) have substantially reduced metal related
artifacts. However, these techniques are
limited by distortions related to frequency encoding (3,4) that manifest as
signal loss or signal pile-up artifacts.
The
purpose of this work was to develop a clinically feasible, fully phase encoded
method to image near metallic implants.
Theory
In
this work we propose an ultra-short echo time (UTE) acquisition with broadband
excitation of the entire off-resonance spectrum in a single acquisition, by
using low flip angles and non-selective RF pulses. Large in-plane distortions
caused by the combination of frequency encoding and broadband excitation were
avoided by using single point imaging (SPI) while clinically feasible
acquisition times were obtained through 3D undersampling schemes. Figure
1 describes the proposed UTE-SPI pulse sequence where excitation occurs with
the gradients on at a maximum gradient amplitude (“G
RF”) to minimize
unwanted slice selectivity and signal excitation loss near the implant
(5). Immediately after excitation, gradients
are ramped.
Methods
All
acquisitions were performed at 3T (MR 750, GE
Healthcare, Waukesha. WI) using a 16-channel
wrap coil (NeoCoil, Pewaukee, WO). All UTE-SPI
datasets were reconstructed using a 3D GRAPPA (6) reconstruction. UTE-SPI acquisitions of a total hip replacement phantom composed of a Co/Cr/Mo alloy head and titanium stem (Alliance X-Series,
Biomet Orthopedics, Inc., Warsaw, IN) were acquired using three different RF
pulses: 8μs (GRF=30%, TE=270μs), 24μs (GRF=10%, TE=310μs) and 100μs (GRF=0%, TE=336μs).
The
following imaging parameters were held constant over all acquisitions: R=2x2x1, matrix=224x148x80, TR=1.5ms, FA=2°, Field of
View=224x170x80mm, BW=±250kHz, Gmax=45mT/m, acquisition time
=9.5min.
UTE-SPI (8μs and 24μs RF pulses), T1-weighted MAVRIC, and
2D-FSE acquisitions were acquired of a
volunteer with a total knee replacement.
UTE-SPI
imaging parameters were: R=2x2x1, TR=1.5ms, TE=366μs, matrix=224x170x40, GRF=0%, Gmax=45mT/m, BW=±250kHz, Field of View=224x170x180mm,
FA=2°/6° (8μs/24μs), acquisition
time=5:32min. T1-weighted
MAVRIC imaging parameters were: R=2x2, TR=1245ms, TE=6.7ms, FA=90°, ETL=24, matrix=224x170x40, Field of
View=224x170x180mm, acquisition time=5:22min. 2D-FSE imaging parameters were:
TR=4000ms, TE=8.8ms, FA=111°, ETL=24, matrix=256x256x30, Field of View=224x224mm,
slice thickness=4mm, slice spacing=1mm, NEX=2.
Results and Discussion
Figure
2 demonstrates that short RF pulses allow for improved spectral coverage and
smaller minimum TEs (through the use of higher G
RF) allowing imaging
of signal close to the implant. Improved
T
1 weighting can be achieved using longer pulses with larger maximum
attainable flip angles (Figure 3). A
comparison of the 24μs (FA=6°) UTE-SPI to T
1-weighted
MAVRIC with matching resolution showed both acquisitions had similar contrast
and acquisition efficiency. Blue arrows highlight
signal pile-up and ripple artifacts seen in the MAVRIC image. UTE-SPI techniques also open the potential to
image short T
2* signal such as the plastic liners of implants.
Future work will examine the feasibility of imaging these liners.
Conclusions
3D undersampled UTE-SPI acquisitions with
broadband excitation can generate distortion free images near metallic implants
in clinically feasible acquisition times.
The trade-offs of shorter and longer RF pulses were demonstrated in a
total hip replacement phantom and a volunteer with a total knee replacement. Shorter RF pulses captured signal closer to
the implant by improving spectral coverage and allowing shorter TEs while longer
pulses improve T
1 weighting through the larger maximum attainable
flip angles.
Acknowledgements
The authors acknowledge the support of NSERC, NIH
(UL1TR00427) and GE Healthcare.References
1. Koch KM, Lorbiecki JE, Hinks RS, King KF. A multispectral three-dimensional acquisition technique for imaging near metal implants. Magn. Reson. Med. 2009;61:381–390.
2. Lu W, Pauly KB, Gold GE,et al. SEMAC: Slice encoding for metal artifact correction in MRI. Magn. Reson. Med. 2009;62:66–76.
3. Koch KM, King KF, Carl M, Hargreaves BA. Imaging near metal: The impact of extreme static local field gradients on frequency encoding processes. Magn. Reson. Med. 2013:71: 2024–2034.
4. Smith MR, Artz NS, Wiens C, et al. Characterizing the limits of MRI near metallic prostheses. Magn. Reson. Med. 2014: doi: 10.1002/mrm.25540.
5. Jang H, Wiens CN, McMillan AB. Ramped hybrid encoding for improved ultrashort echo time imaging. Magn. Reson. Med. 2015: doi: 10.1002/mrm.25977.
6. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn. Reson. Med. 2002;47:1202–1210.