Martyn Paley1, Steven Reynolds1, Sarah Calvert2, and Allan Pacey2
1Immunity, Infection and Cardiovascular Disease, University of Sheffield, Sheffield, United Kingdom, 2Oncology and Metabolism, University of Sheffield, Sheffield, United Kingdom
Synopsis
A novel encoding method known as Continuously Ordered B0 Readout acquisition or COBRA is described. The method uses an additional B0 coil with a unique field at every point in space to perform Volume Frequency Encoding. The field increases as a monotonic function within the coil and can be sorted to the appropriate 3D location for reconstruction. 3D data sets have been acquired and reconstructed at 9.4T.Purpose
To develop a fast imaging method using 3D volume frequency encoding (VFE).
Background
Use of auxiliary field coils
to create steps which allow simultaneous parallel slice imaging such as in B1Ac-MAMBA
(1,2) aim to increase MR throughput dramatically. A new simultaneous volume frequency
encoding (VFE) imaging method known as COBRA (Continuously Ordered B0
Readout Acquisition) is introduced which uses a B0 field varying
uniquely at each point within the 3D imaging volume. The frequency at each
point can be mapped a priori and then used in conjunction with a single
broadband pulse-acquire sequence to provide complete, rapid volume coverage in
a single FID.
Theory and Methods
The B0 fields
from a supplementary ten turn helical coil, with a variable pitch, set
proportional to the square root of the z coordinate were calculated using the
Biot-Savart law in Matlab (Mathworks, Natick, CA), figure 1, with a 29x29x29
matrix. Figure 1 shows the vector field from the simulated COBRA coil and
Figure 2 shows the calculated Bz field map from positions ascending
through the coil. Figure 3 shows the positive, sorted Bz field
values from within the coil varying with a continuous, monotonic function.
A prototype COBRA coil was wound on a 5mm, water
filled NMR tube according to the above equation using 0.2 mm diameter wire and
placed in a 10mm RF coil at the center of an Avance III scanner operating at
9.4T/400MHz (Bruker, Ettlingen, DE). A pulse-acquire sequence was collected
with a receiver bandwidth of 100ppm and a RF pulse bandwidth of 10KHz with 24389
acquisition points. A current of 0.25A
was used to generate a maximum additional COBRA field of 0.3mT as measured from
the maximum frequency shift produced. An image volume set was created by
reordering the frequency data to the known 3D locations as modeled for the
COBRA coil a priori, using the sort index.
Results
Figure
4 shows the indices of the field sorting vector used to reconstruct the images by
assigning signal to the predicted 3D field locations. Each 3D location has a unique B0
allowing Volume Frequency Encoding (VFE).
Figures 5 shows a simulated, unsmoothed 1DFT of a COBRA FID calculated
for a centrally located 14x14x29 rectangular phantom with increasing intensity
and added Gaussian noise and Figure 6 shows the reconstructed image.
Figure 7 shows the acquired 9.4T
COBRA data set for a uniform cylindrical Gd-doped water phantom filling the constructed
coil with similar characteristics to the simulation shown in Figure 5. Figure 8
shows the 3D sorted, positive field masked, 29x29x29 volumetric COBRA images reconstructed
from a single FID.
Discussion and Conclusion
As only static field
encoding is used, the acquisition is completely silent and encodes an entire
volume within a single FID. The images have a nominal resolution of 0.22x0.22x0.44
mm and thus have relatively low SNR in a single shot. Improved encoding but
with restricted physical access can be produced using conical coils. Helical
coils could also be used together with a static Z-gradient. Fiducial markers
might improve registration of data sets to field calculations.
Further development of the
COBRA 3D volume frequency encoding (VFE) method could be a useful addition for
rapid 3D MR imaging particularly in cases where peripheral nerve stimulation or acoustic noise is
a concern.
Acknowledgements
We acknowledge funding from MRC grant MR/M010473/1References
1. Feinberg DA, Setsompomp K, Ultra-fast MRI of the human brain with simultaneous multi-slice imaging.J
Magn Reson. 2013; 229: 90–100
2. Paley M et al., B1AC-MAMBA: B1 array combined with multiple-acquisition micro B0 array parallel magnetic resonance imaging. Magnetic Resonance in Medicine
2003; 249:1196–1200.