Judy Alper1,2, Priti Balchandani1, Francois Fay1, and Hadrien Dyvorne1
1Translational and Molecular Imaging Institute, Icahn School of Medicine at Mount Sinai, New York, NY, United States, 2Department of Biomedical Engineering, City College of New York, New York, NY, United States
Synopsis
MRI has long been used as a detection tool for cells labeled
with superparamagnetic iron-oxide nanoparticles (SPIOs). Positive contrast
imaging of off-resonance SPIO signal provides benefits over negative contrast
methods and imaging the SPIOs at 7 Tesla (7T) allows for leveraging greater
off-resonance sensitivity for quantitative imaging of smaller cell populations.
In this study, we imaged a cell phantom containing SPIO labeled macrophages at
7T. We demonstrated the performance of frequency shift imaging (FSI), a new
acquisition technique, for characterizing the magnetic signature of SPIOs, as
compared to negative contrast methods.
INTRODUCTION
MRI has been a primary tool used for detecting and tracking
the location of cells labeled with superparamagnetic iron-oxide nanoparticles
(SPIOs). While most SPIO applications have relied on negative contrast
sequences (1),
positive contrast imaging of the off-resonance SPIO signal provides clear
benefits (2, 3), such as reduction of background tissue signal and possible quantification of
labeled cells in the targeted organ. Moreover, SPIO imaging at ultrahigh field
strengths, such as 7 Tesla (7T), makes it possible to leverage the greater
off-resonance sensitivity afforded by higher field strengths to provide
quantitative imaging of smaller cell populations. In this work we introduce frequency
shift imaging (FSI), a novel acquisition technique that combines efficient
interleaved spectrally selective excitations with fast spiral acquisition to
perform comprehensive characterization of the magnetic signature of SPIOs in a
reasonable scan time. We demonstrate the performance of the FSI sequence for
imaging macrophages labeled with SPIOs and compare the novel sequence with
standard negative-contrast acquisitions performed at 7T.
METHODS
Cell phantom:
Mouse tumor macrophages were grown in DMEM media in two T75 flasks. Both flasks
were incubated for 4 hours at 37° Celsius, one with 30 mg Fe/mL
ferumoxutol (labeled cells), the other without (control cells). Subsequently, the
cells were fixed with 4% PFA and four samples (of 1 million cells each) were
made by mixing labeled and unlabeled cells at 0%, 25%, 75% and 100% population ratio.
The cell samples were embedded in 2% agar gel for MR imaging.
Sequence Design: A
15 ms-long, 170 Hz bandwidth, self-refocused RF pulse was designed using the
Shinnar-Le Roux (SLR) algorithm (Fig. 1)
in order to provide minimum echo time and high spectral selectivity, as
described in (2).
Center-out 3D stack-of-spiral readouts were implemented to minimize the echo
time in order to reduce signal loss due to transverse relaxation. 15 different
frequency points (-1400 to 1400 Hz) were sampled at 200 Hz intervals in
interleaved fashion. Interleaving frequency sampling in time allows for fast
repetition rates (24 ms between consecutive pulses) while providing sufficient
signal recovery (TR=360 ms between consecutive excitations of the
same frequency band).
MR imaging: The
stack-of-spiral FSI sequence acquired 30 partitions at isotropic 1.0 mm
resolution, using segmented spiral readouts (140 mm field of view, 15 arms per
2D spiral, 5.5 ms per readout). Total acquisition time for interleaved FSI was
10 minutes. In addition, standard negative-contrast sequences were acquired at
1.0 mm isotropic resolution on the same phantom, including a 3D gradient echo
(TE/TR 2.4/5.3 ms) and a 2D interleaved spin echo (TE/TR 8.7/1000 ms).
RESULTS
The FSI measured multi-frequency spin echo signal displays
the dipole pattern of the magnetic field of the SPIOs . The regions along the
main magnetic field axis are excited by positive frequencies, while regions
perpendicular to the main magnetic field are excited by negative frequencies.
This difference in region excitement between positive and negative frequencies
is seen in the different orientations of the highlighted dipole patterns.
Moreover, lower frequency shifts lead to increased signal and larger excitation
patterns, while higher frequency shifts result in the opposite (Fig 2). The residual background signal
that appears at 200 Hz is a consequence of the large water linewidth at 7T.
Positive contrast is shown to have increased localization
and background suppression compared to negative-contrast sequences (Fig 3). The positive contrast technique
can provide a framework for quantitative analysis, since it allows for the
detection of pattern size, which is visually proportional to the amount of
labeled cells.
Discussion
In this study, we obtained a comprehensive characterization
of the SPIO magnetic signature in one acquisition and with fast spiral readout using
the novel FSI sequence. We have demonstrated the efficiency of FSI for positive
contrast imaging of SPIO labeled macrophages at 7T, as compared to negative
contrast methods. Positive contrast imaging suggests a path for quantifying
labeled cells in a targeted organ, with significant biomedical applications. For
future work, we plan to compare detection limits between 3T and 7T to evaluate the
advantages of 7T imaging for SPIOs. We
also plan to perform this imaging in
ex
vivo tissue samples and
in vivo animal
models. Our final goal is to develop electromagnetic
models to quantify labeled cells content
in
vivo, based on observed multi-frequency patterns.
Acknowledgements
NIH grant R00 NS070821 (Priti Balchandani) Icahn School of Medicine Capital Campaign, Translational and Molecular Imaging Institute and Department of Radiology, CCNY Department of Biomedical EngineeringReferences
1. Bulte JW, Kraitchman DL. Iron oxide MR
contrast agents for molecular and cellular imaging. NMR Biomed.
2004;17(7):484-99.
2. Balchandani P, Yamada M, Pauly J, Yang
P, Spielman D. Self-refocused spatial-spectral pulse for positive contrast
imaging of cells labeled with SPIO nanoparticles. Magn Reson Med.
2009;62(1):183-92.
3. Cunningham CH, Arai T, Yang PC,
McConnell MV, Pauly JM, Conolly SM. Positive contrast magnetic resonance
imaging of cells labeled with magnetic nanoparticles. Magn Reson Med.
2005;53(5):999-1005.