Wideband Adiabatic Inversion with a Hyperbolic Secant 4 (HS4) Pulse for Late Gadolinium Enhancement MRI of Patients with Implanted Cardiac Devices
Shams Rashid1, Jiaxin Shao1, and Peng Hu1,2

1Radiological Sciences, University of California, Los Angeles, Los Angeles, CA, United States, 2Biomedical Physics Inter-Departmental Graduate Program, University of California, Los Angeles, Los Angeles, CA, United States

Synopsis

We present a wideband inversion pulse of 8kHz bandwidth designed with a hyperbolic secant 4 (HS4) adiabatic inversion pulse, for use in wideband late gadolinium enhancement (LGE) MRI in patients with implantable cardioverter-defibrillators (ICDs). This HS4 wideband pulse has more than twice the bandwidth of the previously reported wideband hyperbolic secant (HS) adiabatic inversion pulse for wideband LGE. We demonstrate that the wideband HS4 pulse is superior to the wideband HS pulse in eliminating hyperintensity artifacts resulting from off-resonance induced by an ICD in the myocardium.

Purpose

Wideband late gadolinium enhancement (LGE) MRI was recently proposed for scar imaging in patients with implantable cardioverter defibrillators (ICDs)1-3. Prior to wideband LGE, scar imaging in ICD patients was difficult due to the appearance of hyperintensity artifacts resulting from severe off-resonance induced by the ICD, which can obscure scar1,2. Wideband LGE uses a wideband inversion pulse with a spectral bandwidth (BW) of 3.8kHz to replace the conventional inversion pulse (typical BW≈1kHz), and successfully eliminates hyperintensity artifacts in the majority of ICD patients1-3.

However, the wideband inversion pulse can only invert spins in the range of ±1.9kHz, whereas the off-resonance in ICD patients’ myocardium could be as large as 6kHz1. To counteract off-resonance larger than 1.9kHz, the center of the frequency sweep of the wideband inversion pulse was shifted by ±1500Hz (depending on the off-resonance polarity). However, this requires repeat scans. In some patients, the off-resonance varies across the myocardium, being <1.9kHz in some slices and >1.9kHz in other slices. Furthermore, we have observed a case where the wideband inversion pulse with +1500Hz frequency shift did not completely eliminate the hyperintensity artifact (Fig1).

These problems may be solved by increasing the BW of the wideband inversion pulse beyond 3.8kHz; however, B1 amplitude of this pulse would then be too high to implement. We present a different wideband adiabatic inversion pulse with 8kHz BW which can resolve the above mentioned issues.

Methods

The previously reported wideband inversion pulse of BW 3.8kHz was designed using an adiabatic hyperbolic secant (HS) inversion pulse1,3. This pulse required a peak B1 amplitude of 19μT. To achieve a BW of 6 kHz with the HS pulse would require an optimal B1 amplitude of 28μT, which is beyond hardware capability. We designed our 8kHz BW pulse using a hyperbolic secant 4 (HS4) adiabatic inversion pulse4,5. The HS4 pulse requires a lower peak B1 amplitude than an HS pulse of similar BW. In our pulse design, we scaled the frequency modulation function4,5 to achieve a frequency sweep of 8kHz. We used a Bloch simulation to study the inversion profile and determine the peak B1 amplitude required.

The HS4 wideband inversion pulse was implemented in an inversion-recovery spoiled gradient-echo LGE sequence with TR/TE=3.9/1.5ms, flip angle=25°, readout BW=500Hz/pixel, matrix: 144x256, resolution=1.4x1.9mm, slice thickness=8mm. The HS4 wideband LGE sequence was tested in a phantom, 2 non-contrast healthy volunteers and 2 patients with ICDs who had been referred to cardiac MRI prior to ventricular tachycardia ablation. All ICD patient scans were carried out under safety guidelines published in the literature for MRI of ICD patients1-3. As a comparison, images in the same slice positions were obtained using the original HS wideband LGE sequence with frequency offsets of 0Hz and ±1500Hz. In the phantom scans, an ICD was placed at 4cm from the phantom. The LGE sequence was used to suppress the signal from the phantom contents. In the healthy volunteers, an ICD was attached to the body coil near the left shoulder of the volunteer to reproduce the myocardial off-resonance of ICD patients. All scans were done on a 1.5T Siemens (Erlangen, Germany) scanner. Images were studied to determine the extent of the hyperintensity artifacts.

Results & Discussion

Bloch simulation results of the wideband HS4 pulse are shown in Fig2. The HS4 pulse required a peak B1 amplitude of 19µT to produce inversion efficiency of 93% (Fig2C). Fig. 2D shows the adiabatic behavior of the HS4 pulse. A minimum B1 amplitude of 13.2µT is required to produce 80% inversion efficiency.

Phantom results are shown in Fig3. The HS4 wideband pulse reduces artifacts more effectively than the HS pulse. Fig4 shows images from a healthy volunteer. Fig4C shows that the HS4 wideband pulse is more successful in eliminating artifacts than the HS pulse (with or without frequency offset).

Fig5 shows images from an ICD patient. The HS pulse with no frequency offset produced hyperintensity artifacts, which resembled scar, in apical slices but not in basal slices (Fig5A&B). Using the HS4 wideband pulse, the hyperintensity artifact was removed, and no other artifact was generated.

Conclusion

We present a wideband inversion pulse designed with an HS4 adiabatic pulse with 8kHz BW. This pulse leads to further reduction of hyperintensity artifacts in LGE over the HS wideband pulse and will reduce the necessity of repeated scans with frequency offsets. The HS4 pulse may also reduce false positives in scar identification in ICD patients. The HS4 wideband pulse should have important applications in ICD patients for 2D LGE as well as 3D LGE for ventricular and atrial scar imaging and T1 mapping.

Acknowledgements

This work was supported in part by the American Heart Association (AHA) (15POST22700041) and the National Institutes of Health (NIH) (R21HL118533).

References

1. Rashid S, Rapacchi S, Vaseghi M, Tung R, Shivkumar K, Finn JP, Hu P. Improved late gadolinium enhancement MR imaging for patients with implanted cardiac devices. Radiology. 2014;270:269-274

2. Stevens SM, Tung R, Rashid S, Gima J, Cote S, Pavez G, Khan S, Ennis DB, Finn JP, Boyle N, Shivkumar K, Hu P. Device artifact reduction for magnetic resonance imaging of patients with implantable cardioverter-defibrillators and ventricular tachycardia: Late gadolinium enhancement correlation with electroanatomic mapping. Heart rhythm : the official journal of the Heart Rhythm Society. 2014;11:289-298

3. Rashid S, Rapacchi S, Shivkumar K, Plotnik A, Finn JP, Hu P. Modified wideband three-dimensional late gadolinium enhancement MRI for patients with implantable cardiac devices. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2015

4. Tannus A, Garwood M. Improved performance of frequency-swept pulses using offset-independent adiabaticity. Journal of Magnetic Resonance, Series A. 1996; 120:133-137

5. Tannus A, Garwood M. Adiabatic pulses. NMR in biomedicine. 1997;10:423-434

Figures

Fig 1: Wideband LGE image (with +1500Hz frequency shift) in an ICD patient where hyperintensity artifact (red arrow) could not be completely eliminated.

Fig 2: Bloch simulation results of the HS4 wideband inversion pulse. A: amplitude modulation function, with peak B1 amplitude of 19μT. B: Frequency modulation function with a frequency sweep of 8kHz. The frequency sweep corresponds to the spectral BW. C: Inversion profile (simulated at T1=400ms, T2=40ms) for the pulse given by A & B, showing a BW of 8kHz. D: Adiabatic behavior shown by a 3D plot of inversion profile vs. B1 amplitude and RF frequency.

Fig 3: Phantom results. Artifacts resulting from positive and negative off-resonance are marked by +ve and −ve. A: Phantom setup. B: With the HS pulse (no frequency offset), some artifacts remain near the ICD. C: With −1500Hz offset HS pulse, the artifact caused by -ve off-resonance is reduced, but the artifact caused by +ve off-resonance is increased. D: Vice-versa with the +1500Hz offset HS pulse. E: With the HS4 pulse, artifacts caused by +ve and -ve off-resonance are both reduced.

Fig 4: Volunteer results. A: With the wideband HS pulse with no frequency offset, a hyperintensity artifact was produced in the anterior wall of the left ventricle, which resembles scar (red arrow). B: The artifact was eliminated using the +1500 Hz offset HS pulse; however, this produced signal loss in the blood pool. This also produced hyperintensity artifacts in other regions (red*). C: With the HS4 pulse, the hyperintensity artifact was eliminated and no residual artifacts were produced.

Fig 5: ICD Patient Images. A & B: In the basal regions, no hyperintensity artifacts were produced using the HS pulse with no frequency offset (A); however, in an apical slice, a hyperintensity artifact was produced, which resembled scar (B, red arrow). C: The HS pulse with +1500 Hz offset removed the artifact, but produced hyperintensity artifact in another region (red*). D: Using the HS4 pulse, the hyperintensity artifact was removed, and no other artifact was generated.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3169