Shams Rashid1, Jiaxin Shao1, and Peng Hu1,2
1Radiological Sciences, University of California, Los Angeles, Los Angeles, CA, United States, 2Biomedical Physics Inter-Departmental Graduate Program, University of California, Los Angeles, Los Angeles, CA, United States
Synopsis
We
present a wideband inversion pulse of 8kHz bandwidth designed with a
hyperbolic secant 4 (HS4) adiabatic inversion pulse, for use in wideband late
gadolinium enhancement (LGE) MRI in patients with implantable cardioverter-defibrillators
(ICDs). This HS4 wideband pulse has more than twice the bandwidth of the
previously reported wideband hyperbolic secant (HS) adiabatic inversion pulse
for wideband LGE. We demonstrate that the wideband HS4 pulse is superior to the
wideband HS pulse in eliminating hyperintensity artifacts resulting from
off-resonance induced by an ICD in the myocardium.Purpose
Wideband
late gadolinium enhancement (LGE) MRI was recently proposed for scar imaging in
patients with implantable cardioverter defibrillators (ICDs)1-3.
Prior to wideband LGE, scar imaging in ICD patients was difficult due to the
appearance of hyperintensity artifacts resulting from severe off-resonance
induced by the ICD, which can obscure scar1,2. Wideband LGE uses a
wideband inversion pulse with a spectral bandwidth (BW) of 3.8kHz to replace
the conventional inversion pulse (typical BW≈1kHz), and successfully eliminates
hyperintensity artifacts in the majority of ICD patients1-3.
However,
the wideband inversion pulse can only invert spins in the range of ±1.9kHz,
whereas the off-resonance in ICD patients’ myocardium could be as large as 6kHz1.
To counteract off-resonance larger than 1.9kHz, the center of the frequency
sweep of the wideband inversion pulse was shifted by ±1500Hz (depending on the off-resonance
polarity). However, this requires repeat scans. In some patients, the
off-resonance varies across the myocardium, being <1.9kHz in some slices and
>1.9kHz in other slices. Furthermore, we have observed a case where the
wideband inversion pulse with +1500Hz frequency shift did not completely
eliminate the hyperintensity artifact (Fig1).
These problems may be solved by increasing the
BW of the wideband inversion pulse beyond 3.8kHz; however, B1 amplitude of this
pulse would then be too high to implement. We present a different wideband adiabatic
inversion pulse with 8kHz BW which can resolve the above mentioned issues.
Methods
The
previously reported wideband inversion pulse of BW 3.8kHz was designed using an
adiabatic hyperbolic secant (HS) inversion pulse1,3. This pulse
required a peak B1 amplitude of 19μT. To achieve a BW of 6 kHz with the HS
pulse would require an optimal B1 amplitude of 28μT, which is beyond hardware capability.
We designed our 8kHz BW pulse using a hyperbolic secant 4 (HS4) adiabatic
inversion pulse4,5. The HS4 pulse requires a lower peak B1 amplitude
than an HS pulse of similar BW. In our pulse design, we scaled the frequency
modulation function4,5 to achieve a frequency sweep of 8kHz. We used
a Bloch simulation to study the inversion profile and determine the peak B1
amplitude required.
The
HS4 wideband inversion pulse was implemented in an inversion-recovery spoiled
gradient-echo LGE sequence with TR/TE=3.9/1.5ms, flip angle=25°, readout BW=500Hz/pixel,
matrix: 144x256, resolution=1.4x1.9mm, slice thickness=8mm. The HS4 wideband
LGE sequence was tested in a phantom, 2 non-contrast healthy volunteers and 2
patients with ICDs who had been referred to cardiac MRI prior to ventricular
tachycardia ablation. All ICD patient scans were carried out under safety guidelines published in the literature for MRI of ICD patients
1-3. As a comparison, images in the same slice positions were obtained using the original HS wideband LGE sequence with frequency offsets of 0Hz and ±1500Hz. In the
phantom scans, an ICD was placed at 4cm from the phantom. The LGE sequence was
used to suppress the signal from the phantom contents. In the healthy
volunteers, an ICD was attached to the body coil near the left shoulder of the
volunteer to reproduce the myocardial off-resonance of ICD patients. All scans
were done on a 1.5T Siemens (Erlangen, Germany) scanner. Images were studied to
determine the extent of the hyperintensity artifacts.
Results & Discussion
Bloch
simulation results of the wideband HS4 pulse are shown in Fig2. The HS4 pulse
required a peak B1 amplitude of 19µT to produce inversion efficiency of 93%
(Fig2C). Fig. 2D shows the adiabatic behavior of the HS4 pulse. A minimum B1
amplitude of 13.2µT is required to produce 80% inversion efficiency.
Phantom
results are shown in Fig3. The HS4 wideband pulse reduces artifacts more
effectively than the HS pulse. Fig4 shows images from a healthy volunteer. Fig4C shows that the HS4 wideband pulse is more successful in eliminating
artifacts than the HS pulse (with or without frequency offset).
Fig5 shows images from an ICD patient. The HS pulse with no frequency offset
produced hyperintensity artifacts, which resembled scar, in apical slices but
not in basal slices (Fig5A&B). Using the HS4 wideband pulse, the
hyperintensity artifact was removed, and no other artifact was generated.
Conclusion
We
present a wideband inversion pulse designed with an HS4 adiabatic pulse with 8kHz
BW. This pulse leads to further reduction of hyperintensity artifacts in LGE
over the HS wideband pulse and will reduce the necessity of repeated scans with
frequency offsets. The HS4 pulse may also reduce false positives in scar
identification in ICD patients. The HS4 wideband pulse should have important
applications in ICD patients for 2D LGE as well as 3D LGE for ventricular and
atrial scar imaging and T1 mapping.
Acknowledgements
This work was supported in part by the American Heart
Association (AHA) (15POST22700041) and
the National Institutes of Health (NIH) (R21HL118533).References
1. Rashid S, Rapacchi S, Vaseghi M,
Tung R, Shivkumar K, Finn JP, Hu P. Improved late gadolinium enhancement MR
imaging for patients with implanted cardiac devices. Radiology.
2014;270:269-274
2. Stevens SM, Tung R, Rashid S, Gima
J, Cote S, Pavez G, Khan S, Ennis DB, Finn JP, Boyle N, Shivkumar K, Hu P.
Device artifact reduction for magnetic resonance imaging of patients with
implantable cardioverter-defibrillators and ventricular tachycardia: Late
gadolinium enhancement correlation with electroanatomic mapping. Heart
rhythm : the official journal of the Heart Rhythm Society. 2014;11:289-298
3. Rashid S, Rapacchi S, Shivkumar K,
Plotnik A, Finn JP, Hu P. Modified wideband three-dimensional late gadolinium
enhancement MRI for patients with implantable cardiac devices. Magnetic
resonance in medicine : official journal of the Society of Magnetic Resonance
in Medicine / Society of Magnetic Resonance in Medicine. 2015
4. Tannus A, Garwood M. Improved performance of
frequency-swept pulses using offset-independent adiabaticity. Journal of
Magnetic Resonance, Series A. 1996; 120:133-137
5. Tannus A, Garwood M. Adiabatic
pulses. NMR in biomedicine. 1997;10:423-434