Nadja M Meßner1,2, Sebastian Weingärtner1,3,4, Johannes Budjan5, Dirk Loßnitzer6, Uwe Mattler5, Theano Papavassiliu2,6, Lothar R Schad1, and Frank G Zöllner1
1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany, 2DZHK (German Centre for Cardiovascular Research), partner site Heidelberg/Mannheim, Germany, 3Department of Electrical and Computer Engineering, University of Minnesota, Minneapolis, MN, United States, 4Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States, 5Institute of Clinical Radiology and Nuclear Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany, 61st Department of Medicine Cardiology, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
Synopsis
Myocardial
T1- and ECV-mapping for the detection of fibrosis is commonly performed at 1.5T with inversion-recovery (IR) techniques such as MOLLI.
As an
alternative, we studied the robustness and precision of the saturation-recovery
(SR) T1-mapping techniques SAPPHIRE and SASHA at 3T in 20 healthy volunteers.
The resulting T1- and ECV- reference values for SR T1-mapping
were 1578±42ms/0.30±0.03 (SAPPHIRE) and 1523±46ms/0.31±0.03
(SASHA), revealing the underestimation of T1-times by MOLLI to be approximately
20-29%.
Therefore, we suggest SR T1-mapping with its high accuracy, low
precision-loss, and good inter-subject variability as a valuable alternative to
IR T1-mapping at 3T.
Background
Quantitative mapping of myocardial T1 and the extracellular volume fraction (ECV) shows promising
diagnostic value in cardiomyopathies such as diffuse fibrosis [1]. The most commonly used method, the
modified Look-Locker Inversion-recovery (MOLLI) sequence [2], is known to underestimate myocardial T1 by ~20% at 1.5T [3] and is prone to several factors confounding quantification accuracy.
To overcome these problems, saturation-recovery
(SR) based myocardial T1-mapping methods like the SAturation-recovery single-SHot Acquisition (SASHA) [4] and a combined SR and IR method, the Saturation Pulse Prepared Heart-rate independent Inversion-REcovery (SAPPHIRE) magnetization preparation, have been introduced [5].
Recently, several studies took advantage of increased imaging SNR at 3T to achieve increased T1-map quality and quantification precision compared to 1.5T.
Purpose
In this study we sought to investigate the feasibility and robustness of saturation-recovery (SR) myocardial T
1-mapping at 3T and to establish accurate reference values for native T
1-times
and ECV of the healthy myocardium.
Methods
SAPPHIRE and SASHA T1-mapping were compared with the MOLLI sequence on a 3T MRI scanner (Magnetom
Skyra; Siemens Healthcare, Erlangen, Germany) with a 30 channel receiver coil
array. In all three sequences, imaging readout was performed with an ECG-triggered single-shot bSSFP readout and the following parameters: TR/TE/α=2.6ms/1.0ms/35°, in-plane
resolution=1.7×1.7mm2, slice-thickness=6mm, field-of-view=440×375mm2,
bandwidth=1085Hz/px, #k-space lines=139, linear profile ordering, startup-pulses=5
Kaiser-Bessel, GRAPPA-factor=2. Magnetization saturation was achieved using a composite “Water suppression Enhanced through T1-effects” (WET) [6] saturation module, magnetization inversion with an adiabatic full passage tan/tanh pulse [7]. The 5(3)3 MOLLI scheme was employed for native
and the 4(1)3(1)2(1) scheme for post-contrast T1-mapping. Pulse-efficacy, accuracy and precision of the methods were studied in phantom.
20 healthy subjects (27±5 years,10 M) were scanned for native and post-contrast T1-times, and regional ECV
values were calculated by involving blood hematocrit. The in-vivo scan protocol is depicted in Fig.1.
T1-times and ECV values were
compared between the sequences using ANOVA. Subjective image quality and susceptibility artifact
rating, as assessed by two blinded readers, and in-vivo precision were
statistically analyzed with Kruskal-Wallis- and Mann-Whitney U-tests.
Inter-subject variability was compared using Bartlett- and F-tests.
Results
WET saturation tests in phantom resulted in an average saturation
efficacy >99% across a broad T1-range. The SR methods
showed excellent accuracy in phantom (<3.9% deviation from the spin-echo reference). However, intra-compartment variability of T1-time was 29% and 50% lower using MOLLI compared with SAPPHIRE and SASHA,
respectively.
Example in-vivo T1-maps of two subjects are shown in Fig. 2. Fig. 3 shows the segment-wise analysis of T1 and ECV. Averaged over all volunteers and segments, SAPPHIRE and SASHA yielded
significantly higher T1-times (SAPPHIRE: 1578±42 ms, SASHA: 1523±46 ms), ECV
values (SAPPHIRE: 0.30±0.03, SASHA: 0.31±0.03) and T1-time variation
(SAPPHIRE: 60.1±8.7 ms, SASHA: 70.0±9.3 ms) compared with MOLLI (T1: 1181±47 ms,
ECV: 0.27±0.03, precision: 53.7±8.1 ms). No significant difference was found in
the inter-subject variability of T1-times or ECV values between the three
methods (T1: p=0.90, ECV: p=0.78). The average quality and artifact scores of the T1-mapping methods were: MOLLI: 3.4/3.6, SAPPHIRE: 3.1/3.4, SASHA: 2.9/3.2; 1: poor - 4: excellent, as displayed in Fig. 4.
Discussion
For all methods, native T1-maps revealed a robust image quality throughout the study. The SR image quality rating was slightly lower and artifact incidence higher compared with MOLLI. However, SR showed significantly better quantification accuracy. SR results in 15-30% lower in-vivo precision compared with MOLLI. Previous studies report differences up to 150% at 1.5T [3], rendering SR at 3T a more competitive alternative to IR. Accordingly, no significant difference was
found in the inter-subject variability among the three methods.
ECV-values were increased compared with reported values at 1.5T [3], potentially due to inaccurate assessment of blood pool T
1-values, as caused by inflowing blood and imperfect saturation. Imaging at 3T using bSSFP yielded high SNR, but led to off-resonance artifacts and SAR limitations. Frequency scouts were successfully used to minimize off-resonance effects. For further improved image quality, SAR reduction is warranted by excitation pulse optimization.
Conclusion
Given
the high accuracy, the low level of precision loss, and the good inter-subject
variability, SR T
1-mapping can be a considered a valuable
alternative to MOLLI T
1-mapping at 3T.
Acknowledgements
No acknowledgement found.References
[1] Burt et al., Radiographics 2014 [2] Messroghli et al., MRM 2004 [3] Roujol et al., Radiology 2014 [4] Chow et al., MRM 2013 [5] Weingärtner et
al., MRM 2014 [6] Ogg et al., JMRB 1994 [7] Kellman et al., MRM 2014