Nadja M Meßner1,2, Sebastian Weingärtner1,3,4, Johannes Budjan5, Dirk Loßnitzer6, Theano Papavassiliu2,6, Lothar R Schad1, and Frank G Zöllner1
1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany, 2DZHK (German Centre for Cardiovascular Research) partner site Mannheim, Mannheim, Germany, 3Department of Electrical and Computer Engineering, University of Minnesota, Minneapolis, MN, United States, 4Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States, 5Institute of Clinical Radiology and Nuclear Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany, 61st Department of Medicine Cardiology, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
Synopsis
Partial volume
artifacts in myocardial T1-mapping are a major source of quantification
inaccuracy. In this study, saturation-recovery T1-mapping at 3T was adapted to
allow for systolic imaging in order to take advantage of the increased
myocardial wall thickness. Estimated T1- and ECV- values for SR T1-mapping
during systole were 1554±3ms/0.29±0.03 compared to 1581±35ms/0.31±0.04 at
diastole. In conclusion, our results show that SR T1-mapping in systole
might be an alternative to derive T1- and ECV-values with reduced effects of
partial volume.Purpose and Background
Spatially resolved quantification of the longitudinal
relaxation time T1 has recently emerged as an important
quantitative biomarker with promising potential for non-invasive tissue characterization,
especially in the presence of diffuse myocardial fibrosis [1]. The combination
with post-contrast T1-time and hematocrit enables the estimation of the extracellular
volume fraction (ECV), which was shown to indicate fibrotic remodeling [2].
However, partial-volume effects near the highly
intense blood pool signal corrupt the methods’ quantification accuracies and
impair their reproducibility. Thus, data acquisition during systole has been
proposed for inversion-recovery (IR) techniques [3,4], in order to benefit from
the increased myocardial wall thickness. Saturation-recovery (SR) T1-mapping is
known to provide more accurate T1-values and better resilience towards heart-rate
variability and sequence variations at 1.5T [5].
Hence, in this work we
sought to study the feasibility of systolic myocardial T1- and
ECV-mapping with a SR technique at 3T.
Materials and Methods
10 healthy volunteers (5m, 5f; 25±4y) underwent T1-mapping before
and 15 min after injection of a Gd based contrast agent (0.2 mmol/kg Dotarem;
Guerbet, Aulnay-sous-Bois, France) at a 3T MRI scanner (Magnetom Skyra; Siemens
Healthcare, Erlangen, Germany) with a 30 channel receiver coil array.
The T1-mapping
sequence comprised a Saturation-Pulse Prepared Heart-rate independent Inversion
Recovery (SAPPHIRE) magnetization preparation [5], adapted for systolic
acquisition as explained in Fig.1. Systolic T1-maps were compared to
conventional SAPPHIRE T1-maps acquired during diastole. A WET module [6] was
used for saturation and an adiabatic full passage tan/tanh pulse [7] for
inversion, followed by a single-shot ECG-triggered bSSFP readout: TR/TE/α=2.6ms/1.0ms/35°, in-plane resolution=1.7×1.7mm2,
slice-thickness=6mm, field-of-view=440×375mm2, bandwidth=1085Hz/px, #k-space lines=139, linear
profile ordering, startup-pulses=5 Kaiser-Bessel, GRAPPA-factor=2.
T1-weighted images were motion
corrected with MoCo (Advanced Retrospective Technique; Siemens Healthcare,
Erlangen, Germany). The thickness of the myocardium was evaluated as the
area between the LV endo- end epicardial borders. T1- and ECV values between
systole and diastole, were statistically compared using a paired student's
t-test at a significance level of p<0.05.
Results
Fig. 2 depicts representative systolic T1-maps. As illustrated, good visual quality, with a homogenous myocardium and
sharp delineation towards the blood pool were obtained throughout the study.
Fig. 3 visualizes the effects of partial-voluming in a cross-section plot in
the septal region of the left ventricle. In the diastolic T1-maps, major T1-time elevation at the myocardial blood interface can be observed, leaving only
a small plateau in the center of the myocardium unaffected of partial-voluming. The increased myocardial thickness in
the systolic images shows a larger plateau for evaluation of myocardial T1-times.
An increase in apparent myocardial thickness in the T1-maps during systole
(apical: 255±85%; mid-ventricular: 254±63%; basal: 209±40%) compared to
diastole has been measured in average over all volunteers.
Fig. 4 shows the average pre-contrast T1-values, T1-time precision
and ECV values of all 10 volunteers in AHA-16-segment bullseye
plots. Systolic SAPPHIRE T1-times (1554±83 ms) are significantly lower than
diastolic T1-times (1581±36 ms) (p<10-7). Systolic ECV values
(0.29±0.03) are significantly lower than diastolic ECV values (0.31±0.04)
(p<10-7).
Discussion and Conclusion
Systolic T1-mapping and ECV-mapping at 3T is feasible with the saturation recovery T1-mapping technique SAPPHIRE and provides robust image
quality. Shorter T1-times during systole are most likely to be explained
by a reduction of partial-voluming, achieved by an increase in myocardial
thickness, as indicated by the comparable plateau values in Fig 3. Difference
between systolic T1- and ECV-values is significant despite the
inter-subject variability. This finding is in accordance with a ShMOLLI
(Shortened MOLLI) study comparing diastolic and systolic T1-values
[4]. The reported T1-times are in good agreement with a recent study on saturation
recovery at 3T (SASHA: 1431-1491 ms)
[8].
Saturation-recovery T1-mapping is inherently robust to changes in
the R-R interval. Therefore, the proposed sequence design could potentially be
used for HR insensitive T1-mapping in arrhythmogenic patients, with no susceptibility
to imaging artifacts induced by the major variations in the duration of the diastolic
rest-period. Future studies in this cohort are warranted.
In conclusion, our results
show that SR T1-mapping in systole might be an alternative to derive T1- and ECV-values with reduced
effects of partial volume.
Acknowledgements
No acknowledgement found.References
[1] Moon et al., JCMR 2013 [2] Kellman et al., JCMR 2012 [3] Meßner
et al., Proc. ISMRM 2015, p.2607 [4] Ferreira et al., JCMR 2015 [5]
Weingärtner et al., MRM 2014 [6] Ogg et al., J Magn Reson B. 1994 [7]
Kellman et al., MRM 2014 [8] Chow et al., JCMR 2015 [9] Roujol et al, Radiology 2014