Marius Menza1, Daniela Föll2, Jürgen Hennig1, and Bernd Jung3
1University Medical Center Freiburg, Dept. of Radiology - Medical Physics, Freiburg, Germany, 2University-Heart Center Freiburg, Cardiology und Angiology I, Freiburg, Germany, 3University Hospital Bern, Institute of Diagnostic, Interventional and Pediatric Radiology, Bern, Switzerland
Synopsis
MR
Tissue Phase Mapping (TPM) is a powerful approach to assess left ventricular (LV)
function. Conventional Cartesian acquisition-strategies with k-t-based parallel
imaging acceleration allow the acquisition of a single slice within a
breath-hold, but suffer from low spatial resolution. In this work a comparison
with undersampled high-resolution spiral SPIRIT TPM for different trajectory
designs within one breath-hold and free breathing Cartesian k-t-accelerated PEAK
TPM is presented. High image quality, comparable peak velocity values and time
to peaks of spiral SPIRIT TPM for high resolution within a breath-hold might
enhance myocardial functional analysis.Introduction
MR Tissue Phase Mapping (TPM) is a powerful
approach to assess left ventricular (LV) function. Conventional Cartesian
acquisition-strategies with k-t-based parallel imaging acceleration1
allow the acquisition of a single slice within a breath-hold period with high
temporal resolution of about 20 ms. However, despite a k-t-undersampling factor
of R=5 the spatial resolution is limited to about 2.2 to 3.4 mm and therefore
only acceptable for the evaluation of the LV. Introducing navigator respiration
control enables a dramatic increase of spatial resolution allowing the
assessment of the right ventricular motion; however, at the cost of much longer
scan times2.
The aim of this study was to use undersampled spiral
imaging in combination with SPIRIT3-reconstruction to enable the
acquisition of three-directional velocity encoded TPM in a single slice during
breath-hold, while maintaining high temporal and spatial resolution.
Methods
Spiral
trajectories with 8 Interleaves were designed and optimized4 to
support a field of view (FOV) of 32cm (fully sampled calibration area) in a
radius of 1/8 for reconstruction purposes for SPIRIT. Outside the calibration
area FOV was reduced to 16 cm and 10 cm for dual density (DD) spirals (2-fold, DD
2; 3-fold, DD 3). For variable density (VD) an iterative method was used to
linearly decrease the FOV outside the calibration area proportional to the
radius in k-space until the readout duration reaches half (VD 2) and one third
(VD 3) of a fully sampled readout length (Fig. 1).
Measurements in 10
healthy volunteers (age 31±5 years) were performed on a 3T-Prisma-system
(Siemens).
For spiral TPM measurements a basal slice was acquired during 16
heartbeats using three-directional velocity-encoded black-blood prepared
off-centre spiral gradient echo sequence with prospective ECG gating and 1-1-binomial water excitation in breath-hold (Table 1). For comparison a
Cartesian gradient echo sequence was used with navigator respiration control,
1-1-binomial water excitation and k-t-accelerated
PEAK-GRAPPA with R=51,3. Spiral images were iteratively
reconstructed using conjugate gradients SPIRIT with a kernel size of 7x7. The
stopping criterion for the algorithm was a residual of less than 10-³.
Data post-processing (Matlab) included semi-automatic segmentation of the LV,
eddy current correction and a transformation of the measured in-plane
velocities (Vx,Vy) into velocity components perpendicular (Vr) and tangential
(Vφ) to the inner heart wall. For segmental analysis the basal LV was divided
according to the AHA 16-segment model5. Global (averaged over the
entire slice) and segmental systolic and diastolic peak velocities and the
corresponding time to peak (TTP) values were derived for Vr and Vz. Statistical
analysis was performed using an Anova test and multiple comparison with
Dunn-Sidak-correction between Cartesian and spiral datasets
(*p<0.05;**p<0.01).
Results
Spiral
images for all trajectory designs exhibit excellent image quality (Fig. 2). Despite
longer spiral readout durations no significant signal voids caused by
susceptibility artefacts can be observed. Undersampling artefacts are removed
by SPIRIT reconstruction.
Global
velocity time courses of Vz and Vr agree well for all spirals and PEAK (Fig. 3)
with no significant differences for global and segmental as well as systolic
and diastolic Vz peak velocities (Table 2). However, systolic global and
lateral segmental Vr peak velocities (DD 2: 2, 3, 6 of 6 segs; DD 3, VD 2, VD 3:
2, 3 of 6 segs) show significant differences between PEAK and all spirals,
whereas no significant differences could be observed in diastole. All global
and segmental TTP in systole and diastole for both velocities correlate well.
Discussion
Using
spiral SPIRIT TPM with undersampling factors of 2 and 3 for DD and VD provide
excellent image quality and agreement of peak velocities and TTP for Vz and Vr
with k-t-accelerated Cartesian acquisition.
Spirals
show more pronounced systolic and diastolic peak velocities for Vr, which might
be caused by temporal blurring due to k-t-acceleration in Cartesian data. With
a much smaller acquisition matrix (and thus lower spatial resolution) the
recently introduced study with the same acceleration factor of R=5 did not show
temporal blurring effects
1. Simpson et al.
6 reported also
an increase in systolic and diastolic of peak velocities (Vz,Vr,Vφ) for
undersampled 8 interleaved spiral acquisition, reconstructed with nonlinear
SENSE. In comparison to Simpson delineation and structural details of the LV
and RV seem to be clearly improved due to a somewhat higher spatial resolution, SPIRIT reconstruction and reduced artefacts due to shorter spiral readouts in
our study. Assessment of peak velocities and TTP revealed no significant
differences between different spirals acquisition strategies, so that for
further studies VD 3 can be used to achieve high temporal resolution.
Future
work will comprise a comprehensive biventricular analysis of myocardial motion.
Acknowledgements
No acknowledgement found.References
1.
Bauer et al. K-t
GRAPPA Accelerated Phase Contrast MRI: Improved Assessment of Blood Flow and 3-Directional
Myocardial Motion During Breath-hold, J
Magn Reson Imaging. 2013 Nov; 38(5):1054-62.
2.
Menza et al. Segmental Tissue Phase Mapping analysis of biventricular heart
function, Proceedings ISMRM 2013, No. 4483
3.
Lustig et al. SPIRiT: Iterative Self-consistent Parallel Imaging Reconstruction
From Arbitrary k-Space, Magn Reson Med. 2010 Aug; 64(2):457-71.
4.
Hargreaves et al. Time-optimal multidimensional gradient waveform design for
rapid imaging, Magn
Reson Med. 2004 Jan; 51(1):81-92.
5.
Cerqueira et al. Standardized myocardial segmentation and nomenclature for
tomographic imaging of the heart. A statement for healthcare professionals from
the Cardiac Imaging Committee of the Council on Clinical Cardiology of the
American Heart Association, Int J Cardiovasc Imaging. 2002 Feb;18(1):539-42
6.
Simpson et al. Spiral Tissue Phase Velocity Mapping in a Breath-hold with
Non-Cartesian SENSE, Magn
Reson Med. 2014 Sep; 72(3):659-68