Amol Pednekar1, Jiming Zhang2, Debra Dees3, Benjamin Y Cheong3, and Raja Muthupillai3
1Phillips Healthcare, Cleveland, OH, United States, 2Diagnostic and interventional Radiology, CHI St Luke's Health, Houston, TX, United States, 3Diagnostic and Interventional Radiology, CHI St Luke's Health, Houston, TX, United States
Synopsis
Diastolic functional indices based on trans-mitral blood flow velocities are
pre-load dependent and early diastolic filling can be diminished by activities
such as inspiration or Valsalva maneuver. Cardiac cine MR images are
typically acquired during suspended respiration and thus could induce systemic
bias. In this study, we evaluate the impact of respiratory suspension on the
computation of volume-based diastolic indices using peak velocity-based Doppler
echo measurements as the reference. The volume based diastolic indices derived
from high temporal resolution cine MR correlated well with velocity based E/A
ratio from echo while indicating the direct impact of respiratory suspension.
Introduction
Diastolic functional indices based on
trans-mitral blood flow velocities are pre-load dependent and early diastolic
filling can be diminished by activities such as inspiration or Valsalva
maneuver1. Cardiac cine MR
images are typically acquired during suspended respiration and therefore could
induce systemic bias in the estimation of left ventricular diastolic function.
In this study, we evaluate the impact of respiratory suspension on the
computation of volume-based diastolic indices using peak velocity-based Doppler
echo measured ratio of early peak velocity (E) to peak velocity during atrial
contraction (A) measured at the tip of the mitral leaflets as the reference.
Methods
All imaging for this IRB approved prospective
study was performed on a 1.5T commercial MR scanner (Achieva, Philips
Healthcare) in 11 volunteers (3m/8f; age 42(20-60)yrs). MRI: Identical imaging parameters were used for
breath held (BH), and free breathing (FB) cine SSFP sequences (TR/TE/flip
angle: 3/1.5/60°); acqd voxel size: 2.25x2.25x8 mm3; SENSE:2, temp
res: 10-15ms; acq time: 18 RR intervals/slice; covering the LV in short-axis
orientation. FB pulse sequence is described in2. Echocardiography:
Subjects were transported to ultrasound (Philips Healthcare, IE 33) on the same
scanner bed to minimize physiologic variation and E/A ratio was obtained. Data Analysis: CMR expert drawn
endocardial contour at end diastole was propagated across the cardiac phases by
a semi-automated algorithm. Resultant LV contours were manually adjusted by CMR
expert if needed. From these contours time-LV volume curve was further analyzed
using a custom-written software in MATLAB™. The raw LV volume curve was
upsampled by a factor of 4, and the derivative of the time-volume curve was
estimated using the method described in [3]. Following parameters were calculated
from these curves: peak volume change rates during the early filling phase (REFP)
and late filling phase (RLFP), LV volume changes during early
filling phase (VEFP) and late filling phase (VLFP). We defined
REFP/RLFP and VEFP/VLFP ) as MR determined
LV volume based surrogates of velocity based echo index of E over A ratio.
Linear regression and Bland-Altman (BA) analysis was performed on the results
obtained with MR and echo to obtain slope (m), coefficient of determination (r2),
bias (mean of difference), and limits of agreement(LA, 1.96* stdev of diff).
Results
High frame rate cine SSFP sequence during free
breathing provides cine MR images with adequate temporal resolution to estimate
MR based indices (REFP/RLFP, VEFP/VLFP)
of diastolic function. Doppler based E/A ratios (mean 1.23, range 0.67–1.68) were
in good agreement with REFP/RLFP for FB (m= 2.0, r2=
0.5, bias= -0.4, LA = 1.0) and BH (m= 1.69, r2= 0.4, bias= -0.2, LA
= 0.9). Doppler based E/A ratios were in good agreement with VEFP/VLFP
for FB (m= 2.42, r2= 0.5, bias= -1.2, LA = 1.4) and BH (m= 1.59, r2=
0.3, bias= -0.8, LA = 1.0). However, the bias for both diastolic indices between
FB and BH acquisitions was more than 10% (Figure 2). The peak volume
change rate during systolic ejection phase (RSEP) was unchanged while
REFP (19.7%), RLFP (7.2%) and stroke volume (9.6%) were
higher in FB compared to BH (Table 1). The VEFP was higher by 15% while
VLFP (-7.0%) was lower in FB compared to BH (Table 1).
Conclusion
The volume based REFP/RLFP and
VEFP/VLFP ratios derived from high temporal
resolution cine MR correlated well with velocity based E/A ratio from echo. The
complex interactions between respiratory and cardiovascular systems have direct
impact on the measurement of volume-based diastolic indices. This indicates
that volume based diastolic indices, if used for longitudinal follow up, need
to be acquired preferably under free breathing condition to avoid inconsistent
level of respiratory suspension.
Acknowledgements
No acknowledgement found.References
[1]
Lung 159 (175-186), 1981, [2] ISMRM P3938, 2012, [3] JMRI, 31:4
(872-880), 2010.