Fetal Cardiac MRI with self-gated iGRASP
Kostas Haris1,2, Erik Hedstrom2,3, Sebastian Bidhult2, Frederik Testud4, George Kantasis1,2, Henrik Engblom2, Marcus Carlsson2, Nicos Maglaveras1, Einar Heiberg2, Stefan R Hansson5, Hakan Arheden2, and Anthony H Aletras1,2

1Laboratory of Medical Informatics, Department of Medicine, Aristotle University of Thessaloniki, Thessaloniki, Greece, 2Lund Cardiac MR Group, Department of Clinical Physiology and Nuclear Medicine, Skåne University Hospital, Lund University, Lund, Sweden, 3Department of Diagnostic Radiology, Skåne University Hospital, Lund University, Lund, Sweden, 4Siemens Healthcare AB, Malmoe, Sweden, 5Department of Obstetrics and Gynecology, Skåne University Hospital, Lund University, Lund, Sweden

Synopsis

The aim of this study was to demonstrate the feasibility of CINE fetal MRI within a breath-hold with self-gated retrospective binning applied with continuous golden angle radial sampling and iGRASP acceleration. A bSSFP radial acquisition scheme was applied with 0.7x0.7mm2 in-plane resolution and a small golden angle of 23.1o. A total of 3800 radial spokes were acquired within a breath hold of 15s. Cardiac triggering was obtained from the centers of the readouts via Principal Component Analysis. The final images were obtained by using iGRASP. Good quality CINE images from the fetus in the third trimester were acquired.

Introduction

Fetal Cardiac Magnetic Resonance Imaging (MRI) has recently shown promise as a viable tool for evaluating the fetal circulation clinically [1]. For evaluating cardiac function, one of the main issues when performing CINE cardiac fetal imaging is gating the acquisition to the fetal ECG, which has low voltages and is contaminated by the maternal ECG. Solutions for retrospectively sorting the acquired data into the proper cardiac phases have been proposed using routine clinical pulse sequences [2]. For non-fetal MRI, self-gated approaches have also been proposed [3]. The aim of this study was to demonstrate the feasibility of CINE fetal MRI within a breath-hold with self-gated retrospective binning applied to continuous golden angle radial sampled data and reconstructed with iGRASP (iterative Golden-angle RAdial Sparse Parallel) [4].

Methods

Institutional Review Board approval and written informed consent from the study participants were obtained. Exams were performed on a 1.5T Magnetom Aera system with two 18-channel phased-array coils (Siemens Healthcare GmbH, Erlangen, Germany). Short-axis fetal cardiac data were acquired using a radial trajectory with a constant azimuthal increment of small golden angle of order 7 (ψ7 in [5]) corresponding to a rotational angle of approximately 23.1o. This radial angle increment does not cause strong eddy currents, which are a known source of reconstruction artifacts. Thus, no gradient delay correction was performed. Balanced Steady State Free Precession images were acquired with the following parameters: TE/TR=1.88/3.9 ms, flip angle 60ο, pixel size 0.7 x 0.7 mm2, slice thickness 5 mm. A total of 3800 radial spokes were acquired within a breath-hold of just under 15s. The cardiac triggering signal required for the binning stage was estimated using the cardiac motion as it was captured by the magnitude of the center of each k-space spoke. The cardiac triggering signal extraction was based on Principal Component Analysis (PCA) [3,6]. Through PCA the most common signal variation mode among all coils was determined. The principal component representing fetal cardiac motion was then selected as the one with the highest peak in the frequency range of 2–3 Hz, considering normal fetal heart rates being 110-160 beats per minute. The number of cardiac phases depended on the number of available cardiac cycles, with typical values ranging between 18-24 phases. The final images were obtained by using iGRASP reconstruction, formulated as the optimization problem:

d* = arg min { || F S d - m ||22 + λ || T d||1} (1)

where d was the fetal cardiac image series to be reconstructed in x-y-t space, T was the temporal total-variation operator, m = [m1,…,mc] were the acquired multi-coil radial k-space data with c coils, F was the NUFFT gridding operator defined on the radial acquisition pattern, S = [S1,…,Sc] were the coil sensitivity maps and λ was the regularization parameter. Typical λ values were in the range of 0.01 to 0.02. The coil sensitivity maps S were estimated by means of sum of squares. The optimization problem in Eq. (1) was solved by the nonlinear conjugate gradient implemented in MATLAB (The Mathworks, MA, USA) [4]. The reconstruction time was approximately 20 minutes per slice on a medium performance laptop.

Results

Fetal cardiac images acquired from a normal volunteer during the third trimester are shown in the figures. Figure 1 shows a mid-ventricular short-axis view of the heart at end diastole. In this image, the entire fetus and the placenta are visualized with good image quality. In Figure 2 the twenty reconstructed cardiac phases are shown cropped and magnified in order to better appreciate image quality for potential delineations.

Conclusions

In this study, we demonstrated that self-gated fetal cardiac MRI is possible with good image quality. The method was based on iGRASP, which combines sparse sampling with parallel imaging. Eddy currents that are generated due to gradient switching associated with the golden angle were reduced by using the recently proposed small angle increments. The cardiac trigger signal was extracted through multi-coil PCA. In the future these methods can be expanded and refined to allow for free breathing and sensitization to flow thus promoting the use of fetal imaging in the clinic. For clinical fetal cardiac MRI to become generally feasible, image reconstruction speed needs to be increased by an optimized computational implementation.

Acknowledgements

Funding was provided by the Greek General Secretariat for Research and Development via an Excellence grant, by the Swedish Heart and Lung Foundation, the Region of Skåne and by the Medical Faculty at Lund University. We acknowledge Siemens Healthcare for providing access to product source code. The authors also thank Prof Leon Axel for the initial presentation of iGRASP at Lund University.

References

[1] Prsa M et al Circ CVI 7:663 (2014)

[2] Jansz MS et al MRM 64:1304 (2010)

[3] Larson AC et al MRM 51:93 (2004)

[4] Feng L et al MRM 72:707 (2014)

[5] Wundrak S et al IEEE TMI 34:1262 (2015)

[6] Pang J et al MRM 72:1208 (2014)

Figures

Figure 1

Full Field Of View slice of the fetus. The yellow arrow points to the left ventricle (short-axis view). The red arrow points to the placenta.


Figure 2

All twenty reconstructed cardiac phases from Figure 1 shown cropped and magnified. The time span of these images corresponds to an RR interval of 415ms (145 heartbeats per minute).




Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3104