Xin Tian1,2, Hua Li1, Xiaoyu Jiang1, Jingping Xie1, John C Gore1, and Junzhong Xu1
1Radiology and Radiological Sciences, Vanderbilt University, Nashville, TN, United States, 2Radiology, The Second Hospital of Hebei Medical University, Shijiazhuang, China, People's Republic of
Synopsis
Two diffusion-based
method, the CG (constant gradient) and FEXI (filtered exchange imaging) methods, have been developed to provide a flexible and safer means to
measure transcytolemmal water exchange rate $$$k_{in}$$$
non-invasively in vivo. However, neither methods have been fully validated up
to date. In the present work, computer simulations and in vitro experiments
with well-controlled cultured cells with different sizes and permeabilities
were performed to evaluate the accuracy of the CG and FEXI methods. The results
suggest that $$$k_{in}$$$ can
be accurately estimated when $$$k_{in}$$$ < 10 Hz. Although the FEXI method provides less accurate results,
the linear dependence of AXR on $$$k_{in}$$$ suggesting it is still a reliable method.PURPOSE
The measurement of transcytolemmal
water exchange rate $$$k_{in}$$$ may provide insights for more specific
diagnosis of pathophysiological status of biological tissues. Two
diffusion-based methods, the CG
1 (constant gradient) and FEXI
2 (filtered exchange imaging) methods, have been developed to provide a
flexible and safer means to measure $$$k_{in}$$$ non-invasively in vivo. However, neither
methods have been fully validated up to date. In the present work, computer
simulations and in vitro experiments with well-controlled cultured cells with
different sizes and permeabilities were performed to evaluate the accuracy of
the CG and FEXI methods.
METHODS
Theory: $$$k_{in}$$$ can be
directly fitted from the CG method by $$$k_{in}=-\frac{\partial S(\tilde{q}^2,t_D)}{\partial t_D}$$$ where $$$\tilde{q}$$$ is
constant, $$$t_{D}$$$ is
diffusion time, and $$$b$$$ value is sufficiently large. The FEXI method uses a
diffusion filter block before ADC detection, and then provides apparent
exchange rate AXR, which is $$$k_{in}$$$
modulated with extracellular water fraction $$$f_{ex}$$$, i.e. $$$ \text{AXR}=k_{in}\times f_{ex}$$$.
In vitro study: Human myelogenous leukemia K562 cells were
cultured, fixed and then divided into three groups (six samples in each) treated
with vehicle (control), low (0.025% w/v) and high (0.05%) concentrations of
saponin. Saponin is a natural detergent that selectively removes membrane
cholesterols and increases cell membrane permeability without altering other
cell properties3. All MR experiments were performed on an Agilent/Varian
4.7T MRI system. The CG experiments employed 30 gradients separation $$$\Delta$$$ up to 832 ms. The FEXI experiments used diffusion
weighting $$$b_{f}$$$ = 3000
s/mm2 in the filter block and $$$b$$$ = 0 and 1500 s/mm2
in the detection block, and 19 mixing times ranging from 7 to 3000 ms.
Computer Simulation: Water molecules were allowed to diffuse in
intra- and extracellular spaces with a cell membrane permeability separating
the compartments. $$$D_{in}$$$
= 1 μm2/ms, $$$D_{ex}$$$ = 2 μm2/ms, and three different cell diameters 5, 10, and 20
μm were simulated while the intracellular volume fraction was fixed as 61.8%.
18 ground truth $$$k_{in}$$$
values (0.02 – 30 Hz) were simulated. All other parameters were the same as
those used in the in vitro experiments.
RESULTS
Figure 1 shows the
simulated CG method provides accurate estimations of $$$k_{in}$$$ especially
when it is smaller than 15 Hz, which is in the typical physiological range of
many biological tissues. Figure 2 shows the FEXI method constantly
overestimates $$$k_{in}$$$ even with corrections of extracellular water
fraction. However, AXR and fitted $$$k_{in}$$$ showed approximately linear
dependence on the ground truth $$$k_{in}$$$. The cell diameter had minor
influences on both methods when $$$k_{in}$$$ < 10 Hz,
in which the CG and FEXI methods agree well with each other. Note that this $$$k_{in}$$$ range
corresponds to intracellular water lifetime $$$\tau_{in}$$$ > 100 ms, which
is typical for biological tissues
4. Figure 3 and 4 show the
experimental results, and the correlation between of the two methods are
provided in Figure 5. Both methods have the sensitivity to probe the variations
of cell membrane permeability. The linear dependence of AXR from FEXI and $$$k_{in}$$$
from CG methods is consistent with simulations, suggesting AXR is also a
reliable indicator of $$$k_{in}$$$.
DISCUSSION AND CONCLUSION
The CG method has been long used to characterize
transcytolemmal water exchange rate $$$k_{in}$$$. However, due to demands of
high b values, its application in practice is limited. Although the FEXI method
provides less accurate estimations of $$$k_{in}$$$, the approximately linear
dependence of AXR on $$$k_{in}$$$ suggests AXR is a promising means to measure $$$k_{in}$$$. Considering
the feasibility of spatially mapping $$$k_{in}$$$ in humans in vivo, the FEXI
method may have a greater clinical potential compared with the non-imaging CG
method, although its accuracy is compromised.
Acknowledgements
NIH K25CA168936;
R01CA109106; R01CA173593.References
1. Meier C, Dreher W, Leibfritz D. Diffusion in compartmental systems. I. A comparison of an analytical model with simulations. Magn Reson Med. 2003;50(3):500–9. doi:10.1002/mrm.10557.
2. Lasic S, Nilsson M, Lätt J, Ståhlberg F, Topgaard D. Apparent exchange rate mapping with diffusion MRI. Magn Reson Med. 2011;66(2):356–365. doi:10.1002/mrm.22782.
3. Li H, Jiang X, Xie J, McIntyre JO, Gore JC, Xu J. Time-dependent influence of cell membrane permeability on MR diffusion measurements. Magn Reson Med. 2015. doi:10.1002/mrm.25724.
4. Quirk JD, Bretthorst GL, Duong TQ, et al. Equilibrium water exchange between the intra- and extracellular spaces of mammalian brain. Magn Reson Med. 2003;50(3):493–9. doi:10.1002/mrm.10565.