Yihao Guo1, Zhentai Lu1, Yingjie Mei2, Jing Zhang3, and Yanqiu Feng1,4,5
1School of Biomedical Engineering and Guangdong Provincial Key Laboratory of Medical Image Processing, Southern Medical University, GuangZhou, China, People's Republic of, 2Philips Healthcare, GuangZhou, China, People's Republic of, 3Department of Medical Imaging Center, Nanfang Hospital, Southern Medical University, GuangZhou, China, People's Republic of, 4Laboratory of Biomedical Imaging and Signal Processing, The University of Hong Kong, Hong Kong SAR, China, People's Republic of, 5Department of Electrical and Electronic Engineering, The University of Hong Kong, Hong Kong SAR, China, People's Republic of
Synopsis
Free-form deformation registration has been widely used but would
lead to unwanted bias in the registration of different b-value images to b0
image, especially high-b-value images,
due to signal attenuation dependent on b
values. We use fitting accuracy to guid free-form deformation registration. The results of our proposed
method can well realign the different b-value images in liver edges and improve the fitting accuracy.Purpose
Many voxel-based models for quantifying analysis of diffusion-weighted
images have been proposed, which require the diffusion-weighted images across
various b-value images aligned. Any motion of the patient during scan such as normal
respiratory and cardiac motion leads to images misaligned. Respiratory-triggered
acquisition method can alleviate the effect of motion in the cost of more
acquisition time, but may fail in irregular breathing patterns and result in
displacement between different b-value
images. Thus motion correction is a precondition for precise parameter
estimation. For the deformation, the rigid and affine transformation
registration [1-2] could not describe perfectly the characteristics of motion
in abdomen. Free-form deformation (FFD) [3] registration has been widely used
but would lead to unwanted bias in the registration of different b-value images to b
0 image, especially high-b-value images, due to signal attenuation dependent on b values. In this study, we propose a
method that utilizes a fitting accuracy to guide the step in FFD for finding the
optimal transformation based on the similarity measures of mutual information
[4]. The step changed at
different positions depends on the fitting accuracy, which allows small
displacement at high fitting accuracy and large displacement at low fitting accuracy.
Methods
The
object function was $$$E(T)=-E_{MI}(A,T(B))+\lambda *E_{smooth}(T)$$$ where $$$E_{MI}$$$ is mutual information similarity between reference image $$$A$$$ and the transformed image $$$T(B)$$$, $$$E_{smooth}(T)$$$ is the smoothness of the
transformation and $$$λ$$$ is trade-off parameter which was set to 0.01. The gradient vector of
the object function was calculated as $$$\triangledown E=\frac{\partial E(T^l)}{\partial E(T^l)}$$$ with the non-rigid
transformation parameter $$$T$$$ after the $$$l$$$-th iteration. Based on the exponential weighted FFD, the transformation $$$T$$$ can be updated by $$$T^{l+1}=T^l+\mu *\omega(R^2)*\frac{\triangledown E}{|\triangledown E|}$$$ where $$$T^l$$$ is the $$$l$$$-th transformation, $$$T^{l+1}$$$ is
the updated transformation, $$$μ$$$ is the step size and $$$\omega(R^2)$$$ is a weighted matrix. $$$SS_e$$$ is the sum of squares of the distances
between observed data and fitting data and $$$SS_T$$$ is the average of the
observed data. Fitting accuracy is expressed as the following equation in
accordance with the former defined variables: $$$R^2=1-\frac{SS_e}{SS_T}$$$ . $$$R^2$$$ exponential
function weighted equation is $$$\omega (R^2)=e^{-h*R^2}$$$ with the coefficient $$$h$$$ which was set to 2.0, amplifying the difference between
weighted factors and the fitting accuracy
which has been calculated.
Data
Acquisition: Free breathing diffusion datasets were acquired on a 3.0T Philips
scanner using a single-shot spin-echo echo-planar imaging (EPI) sequence with TR/TE 1600/62 ms, matrix 256×256,in-plane
resolution 1.46×1.46 mm
2 slice gap 0 mm,32
slices
and b-values = 0, 10, 30, 60, 100, 150, 200, 400, 600 and 1000 s/mm
2.Results
Figure.1
shows the registration results of the FFD and the proposed $$$R^2$$$-FFD.
The red lines represent the edges of b
0 image. Both of methods align the different b-value
images to b
0 image in the
edge of liver. The mis-registration pointed by white arrows at high-b-value images (g, j) appear with FFD,
while it does not happen at $$$R^2$$$-FFD.
Figure.2 shows the comparison of the registration and deformation results
acquired from FFD and $$$R^2$$$-FFD of b
400 and b
1000 images. There
are mis-registration (b, d) pointed by red arrows and some folding effect (f,
h) in deformed mesh with the FFD, while the registration results are without
mis-registration (c, e) and the deformed meshes are smooth (g, i) with the $$$R^2$$$-FFD. Figure.3 shows the
comparison of the mapping before and after registration with our proposed
method. After registration, fitting accuracy $$$R^2$$$ is raised that pointed by the white arrows. And the
mapping D is homogeneous and its periphery is clear in liver parenchyma.
Discussion and Conclusion
FFD registration method can describe non-grid
deformation but the same step in different positions would fail in finding the
optimal transformation. Our
proposed method introduces the fitting accuracy to guide the step for finding
the accurate transformation, which allows small displacement at high
fitting accuracy and large displacement at low fitting accuracy. We can obtain accuracy parameters
form the IVIM model by using our proposed registration method. Further evaluation of the proposed method on more subjects is warranted
in a future study.
Acknowledgements
No acknowledgement found.References
[1] Mazaheri, Y. Academic radiology 2012; 12:1573-80 [2] Buerger, C. Medical physics 2015; 1:69-80 [3] Graeme P. Penney TMI 1998; 4:586-595 [4] Hassan Rivaz MIA 2012; 2:343-358