Motion correction for abdominal diffusion weighted images by using fitting accuracy guided free-form deformation.
Yihao Guo1, Zhentai Lu1, Yingjie Mei2, Jing Zhang3, and Yanqiu Feng1,4,5

1School of Biomedical Engineering and Guangdong Provincial Key Laboratory of Medical Image Processing, Southern Medical University, GuangZhou, China, People's Republic of, 2Philips Healthcare, GuangZhou, China, People's Republic of, 3Department of Medical Imaging Center, Nanfang Hospital, Southern Medical University, GuangZhou, China, People's Republic of, 4Laboratory of Biomedical Imaging and Signal Processing, The University of Hong Kong, Hong Kong SAR, China, People's Republic of, 5Department of Electrical and Electronic Engineering, The University of Hong Kong, Hong Kong SAR, China, People's Republic of

Synopsis

Free-form deformation registration has been widely used but would lead to unwanted bias in the registration of different b-value images to b0 image, especially high-b-value images, due to signal attenuation dependent on b values. We use fitting accuracy to guid free-form deformation registration. The results of our proposed method can well realign the different b-value images in liver edges and improve the fitting accuracy.

Purpose

Many voxel-based models for quantifying analysis of diffusion-weighted images have been proposed, which require the diffusion-weighted images across various b-value images aligned. Any motion of the patient during scan such as normal respiratory and cardiac motion leads to images misaligned. Respiratory-triggered acquisition method can alleviate the effect of motion in the cost of more acquisition time, but may fail in irregular breathing patterns and result in displacement between different b-value images. Thus motion correction is a precondition for precise parameter estimation. For the deformation, the rigid and affine transformation registration [1-2] could not describe perfectly the characteristics of motion in abdomen. Free-form deformation (FFD) [3] registration has been widely used but would lead to unwanted bias in the registration of different b-value images to b0 image, especially high-b-value images, due to signal attenuation dependent on b values. In this study, we propose a method that utilizes a fitting accuracy to guide the step in FFD for finding the optimal transformation based on the similarity measures of mutual information [4]. The step changed at different positions depends on the fitting accuracy, which allows small displacement at high fitting accuracy and large displacement at low fitting accuracy.

Methods

The object function was $$$E(T)=-E_{MI}(A,T(B))+\lambda *E_{smooth}(T)$$$ where $$$E_{MI}$$$ is mutual information similarity between reference image $$$A$$$ and the transformed image $$$T(B)$$$, $$$E_{smooth}(T)$$$ is the smoothness of the transformation and $$$λ$$$ is trade-off parameter which was set to 0.01. The gradient vector of the object function was calculated as $$$\triangledown E=\frac{\partial E(T^l)}{\partial E(T^l)}$$$ with the non-rigid transformation parameter $$$T$$$ after the $$$l$$$-th iteration. Based on the exponential weighted FFD, the transformation $$$T$$$ can be updated by $$$T^{l+1}=T^l+\mu *\omega(R^2)*\frac{\triangledown E}{|\triangledown E|}$$$ where $$$T^l$$$ is the $$$l$$$-th transformation, $$$T^{l+1}$$$ is the updated transformation, $$$μ$$$ is the step size and $$$\omega(R^2)$$$ is a weighted matrix. $$$SS_e$$$ is the sum of squares of the distances between observed data and fitting data and $$$SS_T$$$ is the average of the observed data. Fitting accuracy is expressed as the following equation in accordance with the former defined variables: $$$R^2=1-\frac{SS_e}{SS_T}$$$ . $$$R^2$$$ exponential function weighted equation is $$$\omega (R^2)=e^{-h*R^2}$$$ with the coefficient $$$h$$$ which was set to 2.0, amplifying the difference between weighted factors and the fitting accuracy which has been calculated. Data Acquisition: Free breathing diffusion datasets were acquired on a 3.0T Philips scanner using a single-shot spin-echo echo-planar imaging (EPI) sequence with TR/TE 1600/62 ms, matrix 256×256,in-plane resolution 1.46×1.46 mm2 slice gap 0 mm,32 slices and b-values = 0, 10, 30, 60, 100, 150, 200, 400, 600 and 1000 s/mm2.

Results

Figure.1 shows the registration results of the FFD and the proposed $$$R^2$$$-FFD. The red lines represent the edges of b0 image. Both of methods align the different b-value images to b0 image in the edge of liver. The mis-registration pointed by white arrows at high-b-value images (g, j) appear with FFD, while it does not happen at $$$R^2$$$-FFD. Figure.2 shows the comparison of the registration and deformation results acquired from FFD and $$$R^2$$$-FFD of b400 and b1000 images. There are mis-registration (b, d) pointed by red arrows and some folding effect (f, h) in deformed mesh with the FFD, while the registration results are without mis-registration (c, e) and the deformed meshes are smooth (g, i) with the $$$R^2$$$-FFD. Figure.3 shows the comparison of the mapping before and after registration with our proposed method. After registration, fitting accuracy $$$R^2$$$ is raised that pointed by the white arrows. And the mapping D is homogeneous and its periphery is clear in liver parenchyma.

Discussion and Conclusion

FFD registration method can describe non-grid deformation but the same step in different positions would fail in finding the optimal transformation. Our proposed method introduces the fitting accuracy to guide the step for finding the accurate transformation, which allows small displacement at high fitting accuracy and large displacement at low fitting accuracy. We can obtain accuracy parameters form the IVIM model by using our proposed registration method. Further evaluation of the proposed method on more subjects is warranted in a future study.

Acknowledgements

No acknowledgement found.

References

[1] Mazaheri, Y. Academic radiology 2012; 12:1573-80 [2] Buerger, C. Medical physics 2015; 1:69-80 [3] Graeme P. Penney TMI 1998; 4:586-595 [4] Hassan Rivaz MIA 2012; 2:343-358

Figures

Fig.1 The illustration of registration results with FFD and $$$R^2$$$-FFD. Reference image (a) and its yellow region (b). The red lines represent the edges of b0 image. Registration results with FFD (d, g, j) and $$$R^2$$$-FFD (e, h, k) of different b-value images (c, f, i) to b0 image.

Fig.2 Comparison the registration results and deformations of FFD and $$$R^2$$$-FFD. Reference image (a). Registration results (b, d) corresponding deformations (f, h) with FFD; Registration results (c, e) corresponding deformations (g, i) with $$$R^2$$$-FFD.

Fig.3 Comparison the parametric maps of IVIM model before and after registration of $$$R^2$$$-FFD. The upper row is $$$R^2$$$ before and after registration. The middle row is the mapping of IVIM fitting parameter D before and after registration. The last row corresponding the ROI in the middle row.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
3068