Synopsis
Diffusion
weighted imaging (DWI) has been widely used to evaluate the rate of water
diffusion in long T2 tissues such as the white matter of the brain.
Tissues such as cortical bone, tendons, ligaments and menisci have short T2 and
have little or no signal with conventional clinical MR sequences. Water
diffusion in these tissues is inaccessible. Ultrashort echo time (UTE)
sequences with TEs down to 8 µs have been employed to image the short T2
tissues. The combination of UTE and stimulated echo diffusion preparation
(UTE-DWI) allows water diffusion to be evaluated in short T2 tissues. In this
study we report the use of a novel stimulated echo (STE) diffusion prepared 3D
UTE Cones sequence (3D UTE-Cones-DWI) for assessment of water diffusion in the
Achilles tendon using a clinical 3T scanner. The magic angle dependence of apparent diffusion coefficients (ADC) determined
through this sequence was also investigated.Introduction
Tissues such as cortical bone, tendons, ligaments and menisci have short T
2 and
have little or no signal with conventional clinical MR sequences. Water
diffusion in these tissues is inaccessible. Ultrashort echo time (UTE)
sequences with TEs down to 8 µs have been employed to image the short T
2
tissues. The combination of UTE and stimulated echo diffusion preparation
(UTE-DWI) allows water diffusion to be evaluated in short T2 tissues. In this
study we report the use of a novel stimulated echo (STE) diffusion prepared 3D
UTE Cones sequence
1 (3D UTE-Cones-DWI) for assessment of water diffusion in the
Achilles tendon using a clinical 3T scanner. The magic angle dependence of apparent diffusion coefficients (ADC) determined
through this sequence was also investigated.
Materials and
methods
The
remarkable feature of STE-diffusion preparation is that it allows use of long
diffusion time which can be much longer than the T2 of the tissue being studied2.
This is extremely important not only for imaging but also for the measurement
of ADCs in short T2 tissues, such as the Achilles tendon. The sequence diagram
of the 3D UTE-Cones-DWI is shown in Figure 1. The STE-diffusion
preparation pulse cluster is composed of two pairs of tip-up and tip-down RF pulses.
The composite pulse is insensitive to B1 inhomogeneity. Useful b-values can be
obtained by using a large Tmix time and a short diffusion gradient duration.
This is of benefit when imaging short T2 tissues. After the STE diffusion
preparation, a train of cones k-space spokes is employed for fast data acquisition.
Before each STE-diffusion data acquisition, a long wait time is needed for longitudinal
signal recovery.
Cadaveric
Achilles tendon samples dissected from human ankle specimens (n=3) were
harvested for this study. Data were acquired with the 3D UTE-Cones-DWI on a clinical 3T
scanner (GE Healthcare Technologies, Milwaukee, WI). The Tmix and duration time of each diffusion gradient
were fixed at 120ms and 4ms, respectively. A slice-direction diffusion gradient
with four amplitudes (i.e. 10, 20, 30, 40 mT/m) was used to generate four b-values
(i.e. 14, 57, 127, 227 s/mm2). Other imaging parameters were as
follows: flip angle=10°, TE=32 µs, 5.4 ms per-spoke, 32 spokes per-TR, TR=1500 ms,
FOV=8cm, matrix=128*128, slice thickness=4mm, slice number=10. In addition,
multiple TE data were acquired for fast measurement of T2* values using a 2D radial-UTE
sequence3 (flip angle=10°, FOV=8×8cm2, matrix=128×128, slice
thickness=4mm, 18 TEs = 10µs, 0.1, 0.2, 0.3, 0.4, 0.6, 0.8, 1, 1.5, 2, 3, 4, 5,
6, 7, 8, 9, 10 ms, TR=100ms). The above two protocols were applied to the
tendon sample five times with five different angular orientations (i.e. 0°,30°,57°,69°,90°) between the tendon fiber direction and the B0 field.
Both
the multiple b-value diffusion data and multiple TE data were processed using a
non-linear curve-fit function (i.e. lsqcurvefit) in Matlab to obtain ADC and
T2* values.
Results
Upper row in Figure 2 shows sagittal views of
the tendon images acquired with a clinical GRE sequence (TE/TR = 4/16.7 ms) at the
five angular orientations. The arrows besides the specimen indicated the orientation
of the fibers. Lower row in Figure 2 shows corresponding axial UTE-Cones-DWI images
of the same tendon, where the imaging plane is perpendicular to the fiber
orientation. High signal is observed in the Achilles tendon at different angles
with the 3D UTE-Cones-DWI sequence, demonstrating the feasibility of diffusion
weighted imaging of short T2 tissues.
Figure 3 shows the results of the T2* and ADC measured
at the five angular orientations. T2* increased more than 9 times when the
fibers were oriented from 0° to 54° relative to the signal with fibers parallel
to the B0 field. A strong but smaller magic angle effect was observed in the ADC,
with an increase roughly half compared with T2*.
Dscussion and conclusions
The study showed that it was possible to obtain
high quality diffusion weighted images of the Achilles tendon which has a short
T
2 around 2ms when the fibers are parallel to B0. A marked magic angle effect
was seen in ADC. Here, the signal equation for DWI based on the spin-echo EPI
sequence was employed for data fitting. This simplification was not be very
accurate due to T
1 recovery effects during the Cones data acquisition time.
Multiple cones spokes were acquired with each STE preparation, with each spoke
covering both the k-space center and periphery, resulting in a mixed contrast. More
complicated model may be needed to compensate for the affects due to T
1 recovery
and multi-spoke acquisition to improve the estimation of ADC in short T
2
tissues such as the Achilles tendon.
Acknowledgements
No acknowledgement found.References
1. Carl M, Bydder GM, Du J. UTE Imaging with Simultaneous Water and Fat Signal Suppression Using a Time-Efficient Multispoke Inversion Recovery Pulse Sequence. Magn Reson Med 2015; doi:10.1002/mrm.25823.
2. Guha A, Wyatt C,
Karampinos DC, Nardo L, Link TM, Majumdar S. Spatial variations in magnetic resonance-based diffusion
of articular cartilage in knee osteoarthritis. Magn Reson Imaging 2015;33:1051-1058.
3. Du J,
Chiang AJ, Chung CB, Statum S, Znamirowski R, Takahashi A, Bydder GM.
Orientational analysis of the Achilles tendon and enthesis using an ultrashort
echo time spectroscopic imaging sequence. Magn Reson Imaging 2010;28:178–184.