Zhe Zhang1, Xiaodong Ma1, Bing Zhang2, Ming Li2, Chun Yuan1,3, and Hua Guo1
1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China, People's Republic of, 2Department of Radiology, The Affiliated Drum Tower Hospital of Nanjing University Medical School, Nanjing, China, People's Republic of, 3Vascular Imaging Laboratory, Department of Radiology, University of Washington, Seattle, WA, United States
Synopsis
Single-shot EPI DWI
suffers from geometric distortion which can mask the pathology. Single-shot TSE
DWI can help settle the distortion problems, but the spatial resolution is
limited. In this work, a multi-shot Cartesian TSE DWI method with improved
in-plane resolution is proposed. The ghost artifacts from shot-to-shot phase variations
are corrected using an image domain correction method. The results show that
the proposed method provides diffusion images with improved resolution and insensitivity
to susceptibility induced distortion artifacts.Purpose
Single-shot
echo planar imaging (ss-EPI) is widely used for clinical diffusion weighted imaging (DWI), for
its fast acquisition and motion insensitivity. However, ss-EPI suffers from geometric distortion caused by magnetic susceptibility differences, which has
the potential to mask the pathology, especially when imaging slices around
skull base or temporal lobes
1. Alsop has introduced a single-shot turbo
spin echo (ss-TSE) sequence for DWI which is much
less sensitive to the distortion and chemical shift artifacts in EPI
2.
However, DWI using ss-TSE has the limitation in spatial resolution due to the image
blurring caused by long echo train duration, and the requirement of single shot
to avoid artifacts caused by motion-induced phase errors. In this work, a
multi-shot Cartesian TSE DWI method with improved in-plane resolution is
proposed, and SF-MUSE method
3,4 is applied to correct the ghost
artifacts induced by shot-to-shot phase variations.
Methods
Pulse
Sequence The non-CPMG Cartesian TSE sequence proposed by Alsop2 is implemented
as in Fig. 1. The multi-shot TSE (ms-TSE) phase encoding profile order is set
to linear to reduce image blurring, and partial Fourier acquisition is used for
shortening TE.
Phase Correction Since physiological
motion encoded with diffusion gradients can introduce phase variations in different shots, phase correction is indispensable for multi-shot DWI scans. In this
work, as no additional navigators are acquired, the 2D nonlinear phase
information is calculated inherently and used for correction in image domain. A
MUSE-based method, SF-MUSE3,4 is implemented for shot-to-shot phase
correction.
Experiments All scans were performed on a Philips 3.0T MRI scanner
(Philips Healthcare, Best, The Netherlands) using an 8-channel head coil. Data
were acquired from 3 healthy volunteers. All human studies were performed under
IRB approval from our institution. For the ms-TSE DWI, FOV = 230 × 192 mm2,
in-plane resolution = 1.5 × 1.5 mm2,
slice thickness = 5 mm, acquisition matrix = 152 × 128, Phase encoding
direction = right/left, 4 shots with 19 echoes per shot, partial Fourier factor
= 0.6, echo train duration = 114 ms, TE/TR = 80/2300 ms, number of slices = 24,
scan time = 4.5 min, diffusion preparation was applied in 3 orthogonal
directions with b = 1000 s/mm2, NSA = 4. For comparison, ss-TSE DWI
(echo train duration = 402 ms), ss-EPI DWI with GRAPPA = 2 (effective echo
spacing = 0.416 ms), and 6-shot EPI DWI (effective echo spacing = 0.143 ms) with in-plane
resolution = 1.5 × 1.5 mm2 were also scanned for comparison.
The ms-TSE DWI images reconstructed with no phase
correction and with 2D nonlinear SF-MUSE phase correction were compared. The
ss-EPI DWI reconstructed with GRAPPA, and 6-shot EPI DWI reconstructed with
SF-MUSE images were also compared for evaluation of the distortion reduction.
Results and Discussion
Fig. 2
shows the isotropic DWI of two representative slices using ssTSE (a), ms-TSE
with no correction (b), ms-TSE with 2D phase correction using SF-MUSE (c),
along with the ss-EPI (d) and 6s-EPI images (e). The phase error causes
severe ghost artifacts if no correction is performed. The proposed 2D
correction method can correct these ghost artifacts, and provide less image
blurring compared with ss-TSE. ms-TSE
with SF-MUSE correction provides similar contrast but distortion-free diffusion
images compared with ss-EPI, and even 6-shot EPI (much shorter echo spacing
than ss-EPI) scans (yellow arrow heads).
The scan time for ms-TSE
DWI is longer compared with ss-EPI (1 min). Phase modulation and other methods1,5,6
can also be used for higher SNR efficiency non-CPMG TSE DWI at the cost of
higher SAR. The imaging time and echo train duration can be further reduced by
replacing the TSE readout with GRASE. Compared with PROPELLER DWI methods1,5,
our proposed method is more efficient but limited in ultra-high resolution,
which is the limitation of self-navigated MUSE-based methods. Additionally, ms-TSE
DWI can also be a good candidate for imaging around metal implants.
Conclusion
The proposed multi-shot Cartesian TSE DWI
method with SF-MUSE phase correction can provide diffusion images free from
ghost artifacts and insensitive to susceptibility induced distortions.
Acknowledgements
This work was supported by National Natural Science
Foundation of China (61271132, 61571258) and Beijing Natural Science Foundation (7142091).References
[1] Pipe
JG, et al. Magn Reson Med 2002;52:42-52.
[2] Alsop
DC. Magn Reson Med 1997;38:527-33.
[3] Chen
N-K et al. Neuroimage 2013;72.
[4] Zhang
Z et al. Neuroimage 2015;105.
[5] Pipe
JG, et al. ISMRM 2009;164.
[6] Bastin
ME, et al. Magn Reson Med 2002;48:6-14.