Can myelin protons be directly imaged with the adiabatic inversion recovery prepared ultrashort echo time (IR-UTE) sequence - a validation study based on D2O exchange in sheep brain specimens
Hongda Shao1, Soorena Azam ZAnganeh1, Rong Luo1, Jun Chen1, Graeme Bydder1, and Jiang Du1

1Radiology, University of California, San Diego, San Diego, CA, United States

Synopsis

Directly assessing the integrity of myelin in white matter is important for diagnosis and assessment of prognosis in multiple sclerosis (MS). However, the protons in myelin have extremely short T2s and cannot be directly imaged with conventional clinical MRI sequences. Adiabatic inversion recovery prepared ultrashort echo time (IR-UTE) sequences can detect signal from myelin protons and efficiently suppress the signal from water. In this study we aimed to further validate the IR-UTE technique in sheep brain using a D2O exchange model.

Introduction

Directly assessing the integrity of myelin in white matter may be important for diagnosis and assessment of prognosis in multiple sclerosis (MS), a disabling disease of the central nervous system 1. However, the protons in myelin have extremely short T2s (less than 1 ms) and cannot be directly imaged with conventional clinical MRI sequences which have TEs of several milliseconds or longer 2-4. As a result, conventional clinical sequences only provide an indirect assessment of myelin. It would be a major achievement to directly image myelin and quantitatively evaluate its MR properties such as T2*, T1 and proton density, and potentially so provide a more specific and sensitive evaluation of the damage to this tissue in MS. Adiabatic inversion recovery prepared ultrashort echo time (IR-UTE) sequences can detect signal from myelin protons and efficiently suppress the signal from water 5. In this study we aimed to further validate the IR-UTE technique in sheep brain using a D2O exchange model.

Methods

The IR-UTE contrast mechanism is shown in Figure 1. The longitudinal magnetization of long T2 white matter is inverted by the adiabatic inversion pulse (duration = 8.64 ms) while the short T2 myelin signal is largely saturated. UTE data acquisition starts after a delay of time to inversion (TI) when the inverted long T2 magnetization approaches the nulling point in order to detect myelin signal recovered during TI without water contamination. Six sheep brain specimens were purchased from a local slaughter house. The brainstem of each sheep brain was dissected into ~5 mm slab and imaged using a GE 3T Signa TwinSpeed MR scanner (GE Healthcare Technologies, Milwaukee, MI) and a 1-inch solenoid coil for signal excitation and reception. The MR protocol included four sequences, including a proton-density weighted fast spin echo (PD-FSE) sequence (TR = 6000 ms) to measure PD of long T2 white matter, an IR-FSE sequence (TR = 2000 ms, TI = 50, 75, 100, 150, 200, 300, 400, 500, 700, 1000, 1500 ms) to measure T1 of long T2 white matter, an IR-UTE sequence (TR=1000ms, TI~240 ms) to measure PD of myelin, and IR-UTE imaging with a series of TEs (TE = 0.01, 0.1, 0.2, 0.4, 1.2, 2.0 ms) to measure T2* of myelin in white matter of the brain stem. Other imaging parameters included a flip angle of 60o, a bandwidth of 62.5 kHz, a FOV of 4 cm, a slice thickness of 2 mm, reconstruction matrix of 192×192. The same MR protocol was applied to each brain stem sample before and after D2O exchange for 11 hours and one week. During the D2O exchange, each brainstem was immersed in 30 ml D2O solution (99.8% purity). Changes in PD of long T2 white matter and myelin were measured as the signal change in PD-FSE and IR-UTE images as a function of exchange time, respectively. T2* was quantified using a mono-exponential decay model. Myelin T2* was also plotted against D2O exchange time.

Results

Figure 2 shows PD-FSE images of a brainstem sample imaged at different stages of D2O exchange. There is a significant signal drop with increased D2O exchange time, consistent with the replacement of H2O with D2O which has no MR signal.

Figure 3 shows an example of IR-UTE imaging of the brainstem at different TEs, and T2* analysis for different stages of D2O exchanges. A short consistent T2* was demonstrated for the brainstem at different stages of D2O exchange, consistent with myelin being detected and quantified.

Discussion and Conclusion

The IR-UTE sequence allow uniform inversion and nulling of the long T2 components in white matter of the brainstem, and can provide selective imaging of myelin. The IR-UTE sequence can also be used to measure the MR properties such as T2* and tissue properties such as the PD of myelin. These approaches have the a priori potential to detect pathological changes in myelin, and thus provide a new opportunity to characterize MS in a more specific and potentially more sensitive way than has been possible with conventional clinical sequences which are indirect 6.

Acknowledgements

The authors acknowledge grant support from the NIH (1R01 NS092650).

References

1. Noseworthy JH, Lucchinetti C, Rodriguez M, Weinshenker BG. Multiple sclerosis. N Engl J Med 2006; 2006; 354:938-952.

2. Waldman A, Rees JH, Brock CS, Robson MD, Gatehouse PD, Bydder GM. MRI of the brain with ultra-short echo time pulse sequences. Neuroradiology 2003; 45: 887-892.

3. Horch RA, Gore JC, Does MD. Origins of the ultrashort T2 1H NMR signals in myelinated nerve: a direct measure of myelin content? Magn Reson Med 2011; 66: 24-31.

4. Wilhelm MJ, Ong HH, Wehrli SL, Li C, Tsai PH, Hackney DB, et al. Direct magnetic resonance detection of myelin and prospects for quantitative imaging of myelin density. Proc Natl Acad Sci USA 2012; 109: 9605-9610.

5. Du J, Ma G, Li S, Carl M, Szeverenyi NM, VandenBerg S, et al. Ultrashort TE echo time (UTE) magnetic resonance imaging of the short T2 components in white matter of the brain using a clinical 3T scanner. NeuroImage 2013; 87: 32-41.

6. Filippi M, Rocca MA. MR imaging of multiple sclerosis. Radiology 2011; 259:659-681.

Figures

Figure 1. The IR-UTE sequence (A) and contrast mechanism (B, C). An adiabatic IR pulse is used to invert the long T2 WM (WML). Myelin with an ultrashort T2 is largely saturated and detected by dual echo UTE acquisition and subtraction (B). Inaccurate TI leads to WML signal contamination (C).

Figure 2. PD-FSE imaging of a sheep brainstem before (A), 11 hours (B) and one week (C) after D2O exchange. Signal intensity of long WML reduced by over 30% after11 hours D2O exchange, and more than 95% after one week D2O exchange (D), consistent with H2O being replaced by D2O.

Figure 3. IR-UTE imaging of the brainstem at TEs of 8µs (A), 0.1 (B), 0.2 (C), 0.4 (D), 1.2 (E), 2.0 ms (F), T2* decay curves for myelin before (G), 11 hours (H) and one week (I) after D2O exchange. A relatively constant T2* of 219-282 µs was observed, suggesting myelin being detected.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2986