Hongda Shao1, Soorena Azam ZAnganeh1, Rong Luo1, Jun Chen1, Graeme Bydder1, and Jiang Du1
1Radiology, University of California, San Diego, San Diego, CA, United States
Synopsis
Directly assessing the
integrity of myelin in white matter is important for diagnosis and assessment
of prognosis in multiple sclerosis (MS). However, the protons in myelin have
extremely short T2s and cannot be directly imaged with conventional
clinical MRI sequences. Adiabatic
inversion recovery prepared ultrashort echo time (IR-UTE) sequences can detect
signal from myelin protons and efficiently suppress the signal from water. In
this study we aimed to further validate the IR-UTE technique in sheep brain
using a D2O exchange model.Introduction
Directly
assessing the integrity of myelin in white matter may be important for
diagnosis and assessment of prognosis in multiple sclerosis (MS), a disabling
disease of the central nervous system
1. However, the protons in myelin have
extremely short T2s (less than 1 ms) and cannot be
directly imaged with conventional clinical MRI sequences which have TEs of
several milliseconds or longer
2-4. As a result, conventional clinical sequences
only provide an indirect assessment of myelin. It would be a major achievement to directly image myelin and
quantitatively evaluate its MR properties such as T2*, T1 and proton density, and
potentially so provide a more specific and sensitive evaluation of the damage
to this tissue in MS. Adiabatic inversion recovery prepared ultrashort echo
time (IR-UTE) sequences can detect signal from myelin protons and efficiently
suppress the signal from water
5. In this study we aimed to further validate the
IR-UTE technique in sheep brain using a D2O exchange model.
Methods
The IR-UTE contrast mechanism is shown in Figure
1. The longitudinal magnetization of long T2 white matter is inverted by the
adiabatic inversion pulse (duration = 8.64 ms) while the short T2 myelin signal
is largely saturated. UTE data acquisition starts after a delay of time to
inversion (TI) when the inverted long T2 magnetization approaches the nulling
point in order to detect myelin signal recovered during TI without water
contamination. Six sheep brain specimens were purchased from a local slaughter
house. The brainstem of each sheep brain was dissected into ~5 mm slab and
imaged using a GE 3T Signa TwinSpeed MR scanner (GE Healthcare Technologies,
Milwaukee, MI) and a 1-inch solenoid coil for signal excitation and reception.
The MR protocol included four sequences, including a proton-density weighted
fast spin echo (PD-FSE) sequence (TR = 6000 ms) to measure PD of long T2 white
matter, an IR-FSE sequence (TR = 2000 ms, TI = 50, 75, 100, 150, 200, 300, 400,
500, 700, 1000, 1500 ms) to measure T1 of long T2 white matter, an IR-UTE
sequence (TR=1000ms, TI~240 ms) to measure PD of myelin, and IR-UTE imaging
with a series of TEs (TE = 0.01, 0.1, 0.2, 0.4, 1.2, 2.0 ms) to measure T2* of
myelin in white matter of the brain stem. Other imaging parameters included a
flip angle of 60o, a bandwidth of 62.5 kHz, a FOV of 4 cm, a slice
thickness of 2 mm, reconstruction matrix of 192×192. The same MR protocol was
applied to each brain stem sample before and after D2O exchange for 11 hours
and one week. During the D2O exchange, each brainstem was immersed in 30 ml D2O
solution (99.8% purity). Changes in PD of long T2 white matter and myelin were
measured as the signal change in PD-FSE and IR-UTE images as a function of
exchange time, respectively. T2* was quantified using a mono-exponential decay
model. Myelin T2* was also plotted against D2O exchange time.
Results
Figure 2 shows PD-FSE images of
a brainstem sample imaged at different stages of D2O exchange. There is a
significant signal drop with increased D2O exchange time, consistent with the
replacement of H2O with D2O which has no MR signal.
Figure 3 shows an example of IR-UTE imaging of the brainstem at
different TEs, and T2* analysis for different stages of D2O exchanges. A short
consistent T2* was demonstrated for the brainstem at different stages of D2O
exchange, consistent with myelin being detected and quantified.
Discussion and Conclusion
The IR-UTE sequence allow
uniform inversion and nulling of the long T2 components in white
matter of the brainstem, and can provide selective imaging of myelin. The IR-UTE
sequence can also be used to measure the MR properties such as T2*
and tissue properties such as the PD of myelin. These approaches have the a
priori potential to detect pathological changes in myelin, and thus provide a
new opportunity to characterize MS in a more specific and potentially more
sensitive way than has been possible with conventional clinical sequences which
are indirect
6.
Acknowledgements
The authors acknowledge grant support from the NIH (1R01 NS092650).References
1. Noseworthy JH,
Lucchinetti C, Rodriguez M, Weinshenker BG. Multiple sclerosis. N Engl J Med
2006; 2006; 354:938-952.
2. Waldman A, Rees JH, Brock CS, Robson MD,
Gatehouse PD, Bydder GM. MRI of
the brain with ultra-short echo time pulse sequences. Neuroradiology 2003; 45:
887-892.
3. Horch RA, Gore JC, Does MD. Origins of the
ultrashort T2 1H NMR signals in myelinated nerve: a direct measure of myelin
content? Magn Reson Med 2011; 66: 24-31.
4. Wilhelm MJ, Ong HH, Wehrli SL, Li C, Tsai PH,
Hackney DB, et al. Direct magnetic resonance detection of myelin and prospects
for quantitative imaging of myelin density. Proc Natl Acad Sci USA 2012; 109:
9605-9610.
5. Du J, Ma G, Li S, Carl M, Szeverenyi NM,
VandenBerg S, et al. Ultrashort TE echo time (UTE) magnetic resonance imaging
of the short T2 components in white matter of the brain using a clinical 3T
scanner. NeuroImage 2013; 87: 32-41.
6. Filippi M, Rocca MA. MR imaging of multiple sclerosis. Radiology 2011;
259:659-681.