Yusuke Inoue1, Gou Ogasawara1, Keiji Matsunaga1, Kaoru Fujii1, Hirofumi Hata2, and Yuki Takato2
1Kitasato University School of Medicine, Sagamihara, Japan, 2Kitasato University Hospital, Sagamihara, Japan
Synopsis
We evaluated two
commercially available methods for non-uniformity correction, an image-based
method (SCIC) and a calibration-based method (PURE), in Gd-EOB-DTPA-enhanced MR imaging using a 3T
scanner. SCIC improved uniformity for the precontrast images; however, artificial
hyperintensity in the liver surface was evident especially in the
hepatobiliary-phase images. Quantitative evaluation of contrast effects were
severely distorted by SCIC. PURE improved uniformity in the precontrast and
hepatobiliary-phase images, and appeared to aid quantitative evaluation of the
signal intensity after contrast administration. PURE is indicated to be a
useful non-uniformity correction method in Gd-EOB-DTPA-enhanced MR imaging using a 3T
scanner.PURPOSE
Image non-uniformity causes
substantial problems especially in abdominal imaging using a 3T scanner.
1,2
We evaluated two commercially available methods for non-uniformity correction in Gd-EOB-DTPA-enhanced MR imaging using a 3T
scanner.
METHODS
MR imaging: Twenty
patients who underwent Gd-EOB-DTPA-enhanced
imaging on a GE 3T clinical scanner were
retrospectively analyzed. Dynamic imaging and hepatobiliary-phase imaging were performed using a liver
acquisition with volume acceleration (LAVA) sequence. For hepatobiliary-phase
imaging, the same tuning parameters (receiver gain, transmitter gain, center
frequency, and gradient shim) as dynamic imaging were entered manually to
ensure direct comparison of signal intensities.
Non-uniformity
correction: Image
non-uniformity correction was performed using two methods provided by the
manufacturer: surface coil intensity correction (SCIC) and phased-array
uniformity enhancement (PURE)3. We generated uncorrected images
first, and uniformity correction was applied retrospectively to compare three
sets of images created from the same image data. PURE utilizes
proton-density-weighted images acquired with both the
body coil and the surface coil to calibrate the coil
sensitivity. SCIC is an image-based method and requires no additional scan.
Data analysis: Superficial hyperintensity and focal hyperintensity in
the liver were assessed visually in the precontrast and hepatobiliary-phase axial images
to determine the signal uniformity in the
uncorrected, SCIC, and PURE images. The most prefereble image set was selected
for each phase of a given patient.
The histogram of
the liver signal intensity was assessed for the precontrast and
hepatobiliary-phase images. The entire liver were demarcated manually on seven
slices, and a histogram was created with 15-point smoothing. The signal range
showing frequencies of more than half of the frequency at the mode were
determined, and the width of the range were divided by the mode signal to
calculate full-width at half-maximum (FWHM) of the histogram as a marker of
uniformity.
The signal intensities
for the liver, muscle, and spleen were evaluated before and after contrast
administration. Three circular ROIs were set for the liver, avoiding
superficial hyperintensity on SCIC images, two circular ROIs were set for the
back muscle, and one circular ROI was set for the spleen. The liver-to-muscle
signal ratio (LMR) and liver-to-spleen signal ratio (LSR) were calculated at
each phase. Contrast enhancement ratio was calculated from the liver signal (CER-Liver),
LMR (CER-LMR), and LSR (CER-LSR), dividing the postcontrast value by the
precontrast value.
RESULTS
Examples of images are presented in Figure 1. Superficial hyperintensity was observed in
the uncorrected images of all patients with no correction but in no PURE
images. With SCIC, it was shown in 8 and all 20 patients for the precontrast
and hepatobiliary-phase images, respectively. Focal hyperintensity was noted
only in the PURE images (16 and 13 patients in the precontrast and hepatobiliary-phase
images, respectively), and was located in the lateral segment in all cases. The
SCIC and PURE images were judged the most preferable in 14 and 6 patients for
the precontrast images, respectively, and in 1 and 19 patient(s) in the hepatobiliary-phase
images.
On histogram analysis, FWHM were significantly
smaller for the SCIC and PURE images, indicating better uniformity, than for
the uncorrected images (Fig. 2). In the
comparison of the two correction methods, precontrast FWHM was significantly
smaller for the SCIC images, whereas hepatobiliary-phase FWHM was significantly
smaller for the PURE images.
Uniformity correction largely influenced
the estimates of LMR and LSR (Fig. 3). The PURE
images yielded larger values, especially for LMR, compared to the uncorrected
and SCIC images. LMR and LSR remained relatively constant irrespective of the
imaging phase. The CERs were identical between the uncorrected and PURE images
as predicted theoretically. The SCIC images provided quite different profiles
of signal enhancement (Fig. 4). CER-Liver in the SCIC images were definitely lower than in the uncorrected
images. CER-LMR was close to 1 irrespective of the imaging phase,
and CER-LSR also tended to converge to 1.
DISCUSSION AND CONCLUSION
SCIC allowed favorable non-uniformity
correction on the precontrast images; however, superficial hyperintensity was
evident espeicially in the hepatobiliary phase. Without correction,
underestimation of LMR is considered due to overestimation of the muscle signal
located superficially. SCIC did not remove the underestimation of LMR, and distrurbed
the assessement of the temporal changes in the quantitative parameters of
contrast enhancement. Postcontrast SCIC images should be interpreted
considering such effects and should not be used for quantitative evaluation.
PURE improved uniformity in the precontrast and hepatobiliary-phase images, increasing
LMR and preserving temporal signal changes after contrast administration. PURE
is indicated to be a useful non-uniformity correction method in Gd-EOB-DTPA-enhanced MR imaging
using a 3T scanner.
Acknowledgements
None.References
1. Chang KJ,
Kamel IR, Macura KJ, et al. 3.0-T MR imaging of the abdomen: comparison with
1.5 T. Radiographics. 2008;28(7):1983-98.
2. Lee VS, Hecht
EM, Taouli B, et al. Body and cardiovascular MR imaging at 3.0 T. Radiology.
2007;244(3):692-705.
3. Liney GP, Owen
SC, Beaumont AK, et al. Commissioning of a new wide-bore MRI scanner for
radiotherapy planning of head and neck cancer. Br J Radiol.
2013;86(1027):20130150.