Petros Martirosian1, Rolf Pohmann2, Christina Schraml3, Holger Schmidt3, Nina F Schwenzer3, Martin Schwartz1, Klaus Scheffler2, Konstantin Nikolaou3, and Fritz Schick1
1Section on Experimental Radiology, University of Tübingen, Tübingen, Germany, 2Max Planck Institute for Biological Cybernetics, Tübingen, Germany, 3Department of Diagnostic and Interventional Radiology, University Hospital of Tübingen, Tübingen, Germany
Synopsis
The
separation of portal-venous and hepatic arterial blood supply is important for the
evaluation of chronic liver diseases and the characterization of liver lesions.
The aim of this study was to investigate the capability of a pseudo-continuous
arterial spin labeling sequence, to separate arterial and portal venous
perfusion of the liver using a background suppression technique and different
tagging plane orientations. It was demonstrated that the presented method provides
high quality perfusion images of the liver without application of intravenous
contrast agent and offers promising approaches for the separation of arterial
and portal-venous perfusion fractions.Introduction
The
separation of portal-venous and hepatic arterial blood supply is important for the
evaluation of various liver diseases, especially for the diagnosis of
hepatocellular carcinoma.
1 Arterial spin labeling (ASL) has been
shown to be a promising non-invasive approach for perfusion measurement in
various organs such as the brain or the kidney.
2,3 A small number of
studies have reported the application of ASL for measurement of
liver perfusion in humans.
4,5,6,7 However, liver perfusion imaging with
ASL has been limited due to respiratory and cardiac motion artifacts. It was
demonstrated that background suppression (BS) reduced the sensitivity of ASL measurements to physiological motion.
8 Furthermore, the separation of
arterial and portal-venous perfusion fractions remains difficult, which is
caused by the complex geometry and physiology of blood supply to the liver. The
aim of this study was to investigate the capability of pseudo-continuous ASL
(pCASL) with BS to separate arterial and portal-venous perfusion of the liver
by varying the orientations of the tagging plane.
Materials and Methods
Three healthy volunteers
were examined on a 3T MR scanner (MAGNETOM Prisma, Siemens Healthcare) with body
and spine array coils. The perfusion of the liver was measured using a pCASL
echo planar imaging sequence.9 Four to six sagittal slices were
acquired with parameters: TR/TE, 8000/15 ms; slice thickness/gap, 8/4 mm; in-plane
resolution, 3×3 mm2; acquisition matrix, 86×106; readout bandwidth,
2360 Hz/Pixel. Tagging
flip angles (FA) of 30°-35° and a gradient strength of 7 mT/m were used. Post-labeling
delay (PLD) and tag duration (TD) were set to 2000 ms. 36 label-control image
pairs were acquired within approx. 5 min. BS for liver tissue (T1=800 ms) was utilized
using a double inversion approach. Images were acquired by employing a timed
breathing protocol.
One approach
to separate portal-venous and arterial perfusion is by directly tagging the
hepatic artery (HA) and portal vein (PV) blood supply in two separate
measurements. For the labeling of HA the tagging plane was placed in an axial facial,
cranial to the liver (Figure 2, plane 4), and PV was tagged using a plane below
HA (plane 3). Another approach is based on labeling of the total liver and PV
blood flow, extracting the arterial perfusion by subtraction. For labeling
of total liver perfusion, the tagging plane was placed perpendicular to the PV
or parallel to the aorta (planes 1, 2). An additional measurement with a tagging
plane outside the body (plane 5) was performed as control.
Any images
which showed strong motion artefacts were discarded from subsequent analysis
using Matlab. Data from averaged difference images were used to create a
mask of liver parenchyma, by segmenting those voxels whose signal was 2 times
larger than the median value (corresponding to PV, its branches and hepatic
veins). For an estimation of quantitative perfusion values, the mean perfusion
over the liver parenchyma mask was calculated using a kinetic model.10
Results
Figure 1 shows
an example of perfusion-weighted images measured with and without BS. Images
with BS show clearly higher image quality with reduced motion artifacts. In
Figure 2, mean difference images of total liver perfusion as well as
portal-venous and arterial perfusion can be seen. Perfusion maps of total liver and
PV blood flow are shown in Figure 3. In Figure 4, perfusion maps
for total liver and HA blood flow are depicted.
Mean perfusion values for total liver blood flow were
approx. 83 and 91 ml/min/100g for tagging planes 1 and 2, respectively. Mean perfusion
values for PV and HA blood flow were approx. 54 and 13 ml/100g/min, respectively. Perfusion images of total liver blood flow
revealed a more homogenous distribution than PV blood flow (Figures 2, 3). The
measurement with the tagging plane outside the body resulted in approx. 8
ml/min/100g.
Discussion
In the present work, significant benefits of the background
suppression technique for ASL imaging of the liver could be demonstrated. The
combination of the pCASL sequence with BS provides high quality perfusion
images of the liver and allows for separation of portal-venous and arterial blood
supply. Our results show that direct measurement of relatively low arterial
perfusion signal could be confined by background noise. Nevertheless, this approach
can be advantageous in patients with highly vascularized liver lesions. The
extraction of the arterial perfusion fraction from measurements of total liver
and PV perfusion appears to be a more robust procedure, however, this approach
requires two separate measurements. Both approaches are worth being further
refined to find optimal labeling strategies for measurement of liver
perfusion.
Acknowledgements
No acknowledgement found.References
1. Pandharipandle PV, Krinsky GA, Rusinek H, Lee VS. Perfusionimaging of the liver: current challenges and future goals. Radiology 2005;234:661–673.
2. Luh WM, Wong EC, Bandettini PA, Hyde JS. QUIPPS II withthis-slice TI1 periodic saturation: a method for improving accuracyof quantitative perfusion imaging using pulsed arterial spinlabeling. Magn Reson Med 1999;41:1246–54.
3. Martirosian P, Boss A, Schraml C, Schwenzer NF, Graf H,Claussen CD, et al. Magnetic resonance perfusion imagingwithout contrast media. Eur J Nucl Med Mol Imaging
2010;37:52–64.
4. Hoad C, Costigan C, Marciani L, Kaye P, Spiller R, Gowkand P, et al. Quantifying blood flow and perfusion in liver tissue usingphase contrast angiography and arterial spin labeling. In:Proceedings of the 19th Annual Meeting of ISMRM. Montreal,Canada: International Society of Magnetic Resonance in Medicine;2011. p. 794.
5. Katada Y, Shukuya T, Kawashima M, Nozaki M, Imai H, Natori T, Tamano M. A comparative study between arterial spin labelingand CT perfusion methods on hepatic portal venous flow. Jpn J Radiol 2012;30:863–869.
6. Pan X, Qian T, Fernandez-Seara MA, Smith RX, Li K, Ying K, et al. Quantification of liver perfusion using multidelay pseudocontinuous arterial spin labeling. J Magn Reson Imag 2015,
doi:10.1002/jmri.25070.
7. Schalkx HJ, Petersen ET, Peters NH, Veldhuis WB, van Leeuwen MS, et al. Arterial and portal venous liver perfusion using selective spinlabelling MRI. Eur Radiol 2015;25:1529–1540
8. Robson PM, Madhuranthakam AJ, Dai W, Pedrosa I, Rofsky NM, Alsop DC. Strategies for reducing respiratory motion artifacts inrenal perfusion imaging with arterial spin labeling. Magn
Reson Med 2009;616:1374–1387.
9. Pohmann R, Budde J, Auerbach EJ, Adriany G, Kamil Ugurbil. Theoretical and Experimental Evaluation of ContinuousArterial Spin Labeling Techniques. Magn Reson Med 2010;63:438–446.
10.Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, Edelman RR.A general kinetic model for quantitative perfusion imaging with arterialspin labeling. Magn Reson Med 1998;40:383–396.