Jeong Hee Yoon1, Jeong Min Lee1, Robert Grimm2, Son Yohan3, Kiefer Berthold2, Tobias Kai Block4, Hearsh Chandarana4, and Joon Koo Han1
1Radiology, Seoul National University Hospital, Seoul, Korea, Republic of, 2Siemens Healthcare, Erlangen, Germany, 3Siemens Healthcare Korea, Seoul, Korea, Republic of, 4Radiology, NYU School of medicine, New York, NY, United States
Synopsis
Acquisition of optimal late arterial phase is
essential for focal liver lesion diagnosis and detection at gadoxetic
acid-enhanced liver magnetic resonance imaging (MRI). However, it is often challenging
to obtain late arterial phase due to short arterial window and frequent
involuntary motion during arterial phase. We attempted to characterize the
frequency, onset, and duration of the involuntary motion after gadoxetic acid
injection and whether acceptable quality of late arterial phase can be obtained
using incoherent undersampling technique.Purpose
To investigate clinical
feasibility of free-breathing gadoxetic acid-enhanced liver MRI using incoherent
undersampling in patients with transient dyspnea.
Methods
In this IRB-approved prospective ongoing study, 17
patients (M:F =10:7, mean age 56.2 ± 13.1 years, range 25-73) have
been enrolled and informed consent was obtained from all patients. Gadoxetic
acid-enhanced liver MRI using golden-angle radial sparse parallel (GRASP)
sequence was performed at 3T (1). A total of 10cc of saline was injected
intravenously to check intravenous line patency at a rate of 1.5mL, then, a standard
dose of gadoxetic acid (0.025mmol/kg) was intravenously injected at a rate of
1.5mL/sec, followed by 20mL of saline chaser. Precontrast T1-weighted image
(T1WI), T1WI of saline injection phase (for 1-1.5 minutes, each) and
post-contrast T1WI were obtained (for 3min 30sec). On post-contrast scan,
consecutive 1204 spokes were continuously acquired. Imaging parameters were as
follows: TR/TE = 3.4/1.5, flip angle =12º; FOV 450x450 mm2,
imaging matrix = 256x256; slice thickness was 3-4mm with interpolation. Later,
hepatobiliary phase was obtained 20 minutes after contrast media injection. For
T1WI, Patients’ vital sign and respiratory motion were carefully monitored, and
their breathing capacity was checked before contrast media injection.
Postcontrast images were reconstructed by grouping 89 consecutive spokes and
the time resolution was 13 seconds with or without respiratory gating (window,
0.5). The recorded respiratory pattern during post-contrast phase was compared
with that of saline injection phase to identify patients with transient
dyspnea. The images were analyzed by two radiologists in consensus regarding a)
motion artifact; b) aliasing artifact; c) overall image quality of the arterial,
portal venous and transitional phases on five-point scale, in all patients.
Regarding with motion and aliasing artifacts, higher score indicates a significant
artifact, and higher score of image quality indicates a better exam.
Results and Discussion
In 17 patients, the mean breathing capacity was 35.6 ±
17.1 seconds
(range, 0, 60) and mean breathing cycle (inspiration to next inspiration) was
4.16 ± 1.56 seconds (range, 2.38, 8.61). On
precontrast phase, two patients showed an irregular breathing pattern and the
others showed a regular breathing pattern. During saline injection period, none
of the patients showed transient irregular motion (Fig 1). However, transient motion
was observed in 88.2% (15/17) of the patients on postcontrast phase (Fig 2). In fifteen patients with transient irregular breathing motion, the
transient motion was observed right after contrast media injection, and the
mean duration of the transient motion after gadoxetic acid injection was 33.8 ±
5.3 (range, 26.6, 43.7). Motion artifact substantially decreased after gating
on arterial, portal and transitional phases (P=0.001) whereas aliasing artifact tended to increase after gating
(Table 1). Overall image quality of gated GRASP reconstruction showed significantly
better on late arterial and portal venous phases (Table 1, Fig 3). However, there was no
significant difference of image quality on early arterial and transitional
phases (P=0.22, 0.50, respectively). Our study results showed that the onset of transient
dyspnea and its duration in patients who underwent gadoxetic acid-enhanced
liver MRI. Given that common fixed delay of 15 seconds at gadoxetic
acid-enhanced liver MRI (2), there might be high possibility of unwanted motion
contamination on late arterial phase. Thus, a strategy might be needed to
acquire an arterial phase with acceptable image quality timely and consistently.
Conclusion
Our study showed that
incoherent undersampling using GRASP technique may play a role for obtaining
arterial phase at gadoxetic acid-enhanced liver MRI consistently in all
patients. Furthermore, late arterial phase and following portal venous phase
image quality could be improved on gated GRASP reconstruction images.
Acknowledgements
This study was funded by Bayer-Schering Pharmaceuticals.References
1) Chandarana et al, Invest Radiol 2013;48(1):10-6; 2)
Pietryga et al, Radiology 2014;271(2):426-34